Background: It is unclear whether post-operative errors after toric intraocular lens implantation would be more amenable to pre-operative correction with a fixed adjustment or a correction ratio that scales with the magnitude of pre-operative astigmatism. Purpose: To investigate the effect of pre-operative anterior corneal astigmatism orientation on outcomes of toric intraocular lens implantation in a large population. Methods: A retrospective cohort study of 625 patients undergoing refractive lens exchange through a superior clear corneal incision with Oculentis M-Plus toric intraocular lens implantation at an Optical Express, Inc. located in the United Kingdom and Ireland. Patients were stratified by axis of astigmatism on automated keratometry as with-the-rule, against-the-rule, or oblique. Analysis of visual acuity and refractive outcomes was performed using American National Standards Institute (ANSI) guidelines on astigmatic corrections with non-vector as well as vector analyses. Analysis was limited to one eye per patient. Results: Patients who had with-the-rule (WTR) astigmatism, compared with oblique and against-the-rule (ATR), had higher vector magnitudes of surgically induced refractive correction (2.89D, 2.55D, 2.42D; p < 0.001), and larger correction ratios (1.18, 1.02, 1.07; p < 0.001). This created a change of astigmatic axis in the with-the-rule cohort from WTR pre-operative astigmatism to an average ATR post-operative astigmatism. The overcorrection in the WTR cohort averaged 0.30D. We found that the absolute dioptric overcorrection among WTR patients did not differ significantly with magnitude of corneal astigmatism, but the correction ratio was lower among those with higher baseline cylindrical errors (p = 0.93 and 0.0002, respectively). Conclusions: Refractive lens exchange surgery using toric intraocular lenses overcorrected patients who had with-the-rule astigmatism. Degree of overcorrection did not vary with severity of pre-operative astigmatism. Incorporation of axis of astigmatism in lens selection and reduction of astigmatic correction among with-the-rule patients by an absolute value of 0.25D - 0.35D, rather than proportional adjustments, may reduce cylindrical over-correction.
Toric intraocular lenses are utilized in cataract and refractive lens exchange surgery for patients with corneal astigmatism to reduce spectacle dependence. The lens has a spherocylindrical correction and can be used to correct large powers of corneal cylinder [
This discrepancy has led to investigations into whether orientation of keratometric astigmatism may affect the extent of cylindrical correction. Earlier studies found no difference in magnitude of residual cylinder or uncorrected distance visual acuity after toric lens implantation among eyes having WTR or ATR astigmatism on the basis of axis of astigmatism [
In this study, we investigate differences in degree of astigmatic correction, residual astigmatism, and any correlation between baseline astigmatism and degree of overcorrection among a large group of patients who have toric lens implantation. In this larger population, we analyze whether any post-operative errors would be more amenable to pre-operative correction with a fixed adjustment or correction ratio that scales with the magnitude of pre-operative astigmatism.
This study was approved by the Committee on Human Research at the University of California, San Francisco and the London School of Hygiene and Tropical Medicine and adhered to the tenets of the Declaration of Helsinki.
The study group included all adult patients 70 years of age or younger who underwent primary refractive lens exchange with implantation of an Oculentis M-Plus Toric intraocular lens (Oculentis; Eerbeek, Netherlands) at Optical Express, Inc. in their United Kingdom and Ireland centers from 2010 to 2013 with at least 1.25D of pre-operative anterior corneal astigmatism. Patients were excluded if they had simultaneous additional intraocular surgeries, did not have follow-up at the final three-month outcome including manifest refraction and best-corrected visual acuity (BCVA), did not have a baseline BCVA and keratometry readings recorded in the electronic medical record, had a co-existing ophthalmic disease reducing potential visual acuity, or had greater than 3 diopters of surgically induced refractive change of the corneal surface on post-op month three keratometry, which was felt to be a marker of either data entry error or surgical complication. Patients older than 70 were excluded to reduce variation in lens density of these refractive surgeries, limiting to a population with softer lenses and more consistent surgical course. Manifest refraction was performed using a resolution-based technique in which the endpoint is the least amount of minus sphere that results in the best visual acuity (“push plus”).
The following data were routinely recorded for all patients undergoing refractive lens exchange at Optical Express, Inc., and were gathered and analyzed for this study: patient sex and age, pre- and post-operative visual acuity, manifest refraction, pre-operative automated keratometry (performed with IOL Master in all cases). Post-op month three automated keratometry was also collected on the majority of patients. Patients were classified as having WTR, ATR, or oblique based on axis of astigmatism measured by automated keratometry in positive cylinder notation (WTR axis 60 - 120 degrees, ATR 0 - 30 or 150 - 180 degrees, oblique 30 - 60 degrees or 120 - 150 degrees).
Refractive lens exchange with Oculentis M-Plus toric lens implantation was performed using a 2.6 mm clear corneal incision located at 12 o’clock for both left and right eyes by a total of six different surgeons at four Optical Express centers. The native lens was removed through a superior clear corneal incision for the main wound and phaco-emulsification with either a divide and conquer or chopping technique. The M-Plus toric lens is a one-piece spherocylindrical toric acrylic plate haptic with A-constant 118.2 that is individually customized in advance for each patient with refractive correction, including cylinder, customizable to 0.01D increments. The lens also includes an inferior +3.00D near add embedded on the posterior surface. Intraocular lens power was selected using the online Oculentis toric IOL calculator (available at www.lentistoric.com) with each surgeon entering a unique value for surgically induced astigmatism ranging from 0.2 - 0.3D. A final target refraction of plano was assumed to be the goal of all surgeons given the lens allowed continuous dioptric and cylindrical correction and already included near add.
Statistical analysis was performed using standard analyses of astigmatic correction recommended for laser systems that reshape the cornea adapted to intraocular lens implantation and included both non-vector and vector analyses [
1076 eyes of 714 patients underwent refractive lens exchange with placement of toric intraocular lens from 2010 to 2013. 153 eyes of 89 patients were excluded; with the majority of eyes (91/153) being excluded for no post-op month three vision or refraction and 22 excluded for age greater than 70. This left 923 eyes of 625 patients, of which one eye was randomly selected for each patient, leaving 625 eyes of 625 patients for analysis. The large majority of these patients had with-the-rule astigmatism. There was no significant difference in baseline characteristics of age, sex, or magnitude of anterior corneal astigmatism measured by automated keratometry between the three groups (
Non-vector refractive analysis was performed. There was no significant difference in final uncorrected visual acuity between patients having WTR, ATR or oblique with respective logMar acuities of 0.093, 0.065 and 0.133 (p = 0.07, one-way ANOVA). Average magnitude of residual cylinder at the corneal plane on post-op month three manifest refraction was 0.69D, 0.62D and 0.88D in the WTR, ATR and oblique groups with no significant difference (p = 0.05, Kruskal-Wallis). There was no significant difference among cohorts in the proportion of patients within a half to one diopter of mean-spherical and cylindrical
With-the-rule | Against-the-rule | Oblique | P-value | |
---|---|---|---|---|
Number of Patients | 525 | 61 | 39 | NA |
Right Eyes (N, %) | 275 (52%) | 31 (51%) | 12 (31%) | 0.03 |
Age (±SD) | 53.5 ± 7.8 | 53.6 ± 9.1 | 53.9 ± 8.9 | 0.93 |
Sex (N, % male) | 219 (42%) | 32 (52%) | 19 (49%) | 0.11 |
Pre-Operative BCVA in LogMar (±SD) | 0.01 ± 0.11 | −0.003 ± 0.10 | 0.05 ± 0.11 | 0.02 |
Magnitude of Pre-Operative Corneal Astigmatism (±SD) | 2.49 ± 1.03 | 2.29 ± 1.34 | 2.51 ± 0.90 | 0.76 |
aAstigmatism grouping based on pre-operative automated keratometry. SD: Standard deviation. BCVA: best-corrected visual acuity.
goals at post-op month three manifest or proportion of patients achieving 20/20 or better vision. In pairwise comparison at the 20/40 level, WTR had higher proportions of patients than the oblique cohort of achieving 20/40 or better (p = 0.04), though when corrected for multiple comparisons these pairwise differences no longer achieved significance (p = 0.07) (
Vector analysis of astigmatic refractive error was performed. Magnitude of intended refractive corrections were similar among all groups. After surgical correction, the WTR cohort was over-treated with a treatment error vector of 0.30D × 89.9˚ (95% CI, 0.23D to 0.37D) leaving residual ATR astigmatism with a mean error vector of 0.34D × 7.1˚. In contrast, the ATR cohort was undertreated with a mean treatment error vector of 0.12D × 8.5˚ (95% CI, 0.02D to 0.21D) leaving a mean error vector of 0.12D × 36.7˚. Averages of vector magnitude values tended to be larger than average vectors as different orientations did not partially cancel out (
With-the-rule | Against-the-rule | Oblique | P-value | |
---|---|---|---|---|
±0.50D MSE of goal (N, %) | 369 (70%) | 47 (77%) | 29 (74%) | 0.52 |
±1.0D MSE of goal (N, %) | 478 (91%) | 55 (90%) | 35 (90%) | 0.88 |
Magnitude Residual Cylinder (±SD) | 0.69 ± 0.57 | 0.62 ± 0.53 | 0.88 ± 0.58 | 0.05 |
±0.5D of zero cylinder | 276 (52%) | 35 (57%) | 13 (33%) | 0.05 |
±1.0D of zero cylinder | 430 (82%) | 51 (84%) | 27 (69%) | 0.14 |
Mean UCVA (logMAR) | 0.093 ± 0.15 | 0.065 ± 0.14 | 0.133 ± 0.16 | 0.07 |
UCVA 20/40 or better (N, %) | 497 (95%) | 59 (97%) | 33 (85%) | 0.04a |
UCVA 20/20 or better (N, %) | 53 (10%) | 5 (8%) | 3 (8%) | 0.88 |
aThe significant difference between cohorts at the 20/40 level did not remain through pairwise analysis with multiple hypothesis adjustment. D: Diopters. UCVA: Uncorrected visual acuity. MSE: mean spherical equivalent.
With-the-rule | Against-the-rule | Oblique | P-value | |
---|---|---|---|---|
IRC Magnitude (±SD) | 2.49 ± 1.03 | 2.29 ± 1.33 | 2.51 ± 0.90 | 0.76 |
SIRC Magnitude (±SD) | 2.89 ± 1.13a | 2.42 ± 1.27a | 2.55 ± 1.01 | <0.001 |
Magnitude EV (±SD) | 0.69 ± 0.57 | 0.62 ± 0.52 | 0.88 ± 0.57 | 0.06 |
Correction Ratio (±SD) | 1.18 ± 0.33a,b | 1.07 ± 0.78a | 1.02 ± 0.25b | <0.001 |
Error Ratio (±SD) | 0.31 ± 0.34 | 0.38 ± 0.78 | 0.39 ± 0.29 | 0.14 |
aSignificant pairwise difference between with-the-rule and against-the-rule cohort with p-value < 0.001 after adjustments. bSignificant pairwise difference between with-the-rule and oblique cohort with p-value < 0.001 after adjustments. IRC: Intended refractive correction. SIRC: Surgically induced refractive correction. EV: Error Vector. Correction ratio is the ratio of magnitudes of SIRC divided by IRC. Error ratio is the magnitude of EV divided by IRC.
This study investigated visual and refractive outcomes three months after refractive lens exchange with toric lenses among patients having with-the-rule, against-the-rule and oblique astigmatism. We found that patients having WTR astigmatism were significantly overcorrected compared to patients having ATR and oblique astigmatism. The mean vector error for residual astigmatism in the WTR cohort converted from a WTR axis at 91.4˚ to ATR astigmatism at 7.1˚. This change in axis was due to overtreatment in the WTR axis with a mean treatment error vector of 0.30D at 89.9˚. The average correction ratio for eyes having WTR astigmatism (1.18) was significantly greater than ATR (1.07) and oblique (1.02) astigmatism (p < 0.001). We found that absolute diopters of overcorrection did not vary significantly with pre-operative cylinder, while correction ratio did.
The finding that absolute diopters of overcorrection was not found to significantly correlate with magnitude of pre-operative astigmatism was a novel finding of our study. Further, the magnitude of residual astigmatism seen here is similar to that of prior studies [
There may be several reasons for this overcorrection of WTR patients. The surgical technique in this study uses a superior clear corneal incision at 12 o’clock, a location associated with greater surgically induced astigmatism that may reduce some of the with-the-rule astigmatism (or increase against-the-rule astigmatism) [
The study has several limitations. The study was retrospective with the corresponding limitations, and prospective studies are merited. We assume that the target for residual astigmatism was zero given the M-Plus toric lens is continuously customizable on power and axis, though if residual cylinder was targeted, this would alter calculations of intended refractive correction. The population studied here was that of a refractive practice and generally younger with a high prevalence of WTR astigmatism. Orientation of astigmatism has been shown to become progressively more ATR with aging [
In conclusion, we found that there is a significant overcorrection of patients who have WTR astigmatism in refractive lens exchange with toric lens implantation using anterior corneal keratometric measurements from automated keratometry. Pre-operative consideration of axis of astigmatism and reductions in WTR correction by a fixed value of 0.25 - 0.35D rather than a correction ratio based on magnitude of pre-operative cylinder may further reduce residual refractive error and improve outcomes in surgeries using toric lenses.
The authors declare no conflicts of interest regarding the publication of this paper.
Lowry, E.A., Li, J., Kasi, S.K., Blumberg, S., Morjaria, P., Schallhorn, S. and Naseri, A. (2019) The Effect of Anterior Corneal Astigmatism Orientation on Toric Intraocular Lens Outcomes. Open Journal of Ophthalmology, 9, 84-93. https://doi.org/10.4236/ojoph.2019.92010