Background: The intra-aortic balloon pumping (IABP) is the most used ventricular mechanical assist device. In recent years, the preoperative use in patients with severe ventricular dysfunction presents itself as a great benefic strategy to the postoperative recovery. This paper aim is to evaluate the IABP post-operative benefit in patients with severe ventricular dysfunction. Methods: From January 2011 to March 2016, 125 patients underwent a coronary artery bypass graft (CABG) with cardiopulmonary bypass and preoperative IABP in Teaching Hospital of the ABC Medical School and Hospital Estadual Mario Covas. The inclusion criteria were the presence of severe ischemic cardiomyopathy with left ventricular ejection fraction (LVEF) less than or equal to 40%, estimated by Doppler echocardiography using the Simpson method. The preoperative LVEF was 30.25% ± 8.53% and the diastolic diameter of the left ventricle (LVDD) 67.75 ± 16.37 mm. IABP was installed approximately 15 hours before the surgery. Results: The patients required the IABP for 2.4 ± 1.58 days, and vasoactive drugs, 4.8 ± 2.12 days. We performed 3.2 ± 1.9 grafts per patient and the total length of stay was 07 ± 5.52 days. Cardiopulmonary bypass time was 67 ± 10.95 minutes and anoxia time, 46.4 ± 10.06 minutes. Twelve patients (9.6%) had pneumonia and four (3.2%), atrial fibrillation. We observed a LVDD reduction to 63 ± 16.26 (p = 0.068) and LVEF enhancement to 36.50 ± 16.86 (p = 0.144). The data were analyzed statistically according to the Wilcoxon test. There were no deaths. Conclusion: The initial experience of the authors with the preoperative IABP in patients with severe left ventricular dysfunction suggests great benefit in post-operative recovery with im-provement of LVEF and reduction of LVDD.
Despite advances in therapy, the management of patients with severe left ventricular dysfunction is still a challenge. The intra-aortic balloon (IAB) inserted through the femoral artery was at first described in 1962, as a circulatory support [
In recent years, the device installed on myocardial revascularization preoperative in patients with severe cardiomyopathy has been shown as an important impact strategy to reduce morbid-mortality [
In a recent systematic review and meta-analysis, Poirier et al. included eleven randomized controlled trials and twenty two observational studies, analyzing a total of 46,067 patients. IABP prior to CABG was associated with a significant reduction in hospital and 30-day mortality compared to similar patients without preoperative IAPB [
The aim of this study is to evaluate the IAB post-operative benefit in patients with coronary artery disease and severe ventricular dysfunction.
From January 2011 to March 2016, a prospective study were performed at Teaching Hospital of The ABC Medical School and at Hospital Estadual Mario Covas, which analyzed patients undergoing to a coronary artery bypass graft (CABG) and IABP preoperative.
The group was composed by a hundred and twenty five patients, 108 (86%) of them were male and 17 (14%), female. Their age was 57.13 ± 8.12 years and all had severe left ventricular dysfunction, considering the left ventricular ejection fraction (LVEF), assessed in the Doppler echocardiography through the Simpson method and the diastolic diameter of the left ventricle (LVDD). None of the patients had aneurysm or left ventricular pseudoaneurysm. For ethical reasons, we chose to not form a control group without preoperative IAB.
The main comorbidities found in the study group were: smoking (96.8%); hypertension (84.8%); previous myocardial infarction (67.2%); dyslipidemia (21.6%) and alcohol consumption (18.4%). We included patients with coronary disease requiring surgical treatment and LVEF less than or equal to 40%. Patients with absolute contraindication to the intra-aortic balloon, such as severe aortic insufficiency and aortic dissection, were excluded [
All patients underwent coronary artery bypass grafting with cardiopulmonary bypass (CPB) under mild hypothermia (34˚C) and crystalloid prime exclusively with volume of 1000 ml. In 103 patients (82%) we used the AffinityR Medtronic oxygenator, and in 22 patients (18%), Medizintechnik AG MedosR. All CPB circuits had centrifugal pump. Myocardial protection was performed by intermittent aortic cross-clamping associated with mild hypothermia (34˚C). Complete revascularization was possible in 121 (96%) cases; in three patients (2.4%), one marginal branch of the circumflex artery had less than 1.25 mm diameter, forbidding the manipulation. In another patient (1.6%), the right coronary artery was totally occluded in the middle third without distally refilling. The group used as grafts to the left internal thoracic artery and saphenous vein in all cases, performing 3.2 ± 1.9 grafts per patient.
The IABP was installed about fifteen hours before the surgery, through puncture of the femoral artery using the Seldinger technique and positioning the device with the fluoroscopy assistance. The catheter was chosen according to the patient’s height, following the manufacturer's instructions. The IABP was maintained with full operation (1:1) from the time of installation and discontinued only during CPB. The weaning was begun in the first post-operative day, after the vasoactive drugs dependence reduction and according to the parameters which indicated adequate cardiac output. The IABP was removed right after shutdown, being at least 6 hours working 1:8 without causing increased demand for vasopressors or inotropics.
We chose norepinephrine and dobutamine for all patients. In 26 (20.8%), vasopressin was needed. The removal of vasoactive drugs was early: started immediately after surgery by vasopressin, noradrenaline, and then, finally, without the IABP, dobutamine.
Another echocardiogram was performed around the tenth postoperative day and the results were compared to that carried out in the run up surgery. The good prognosis of these patients was suggested by the bilirubin normality throughout this evaluated period, relevant criteria according to SAPS (Simplified Acute Physiology Score). The data was undergone to a statistical treatment according to the Wilcoxon test, being considered statistically significant when p less than 0.05.
Demographic characteristics from the 125 patients included are exposed on
n | sd | p | |
---|---|---|---|
Age (years) | 57.13 | 8.12 | 0.27 |
Male | 108 (86%) | - | |
Female | 17 (14%) | - | |
LVEF before surgery (%) | 30.25 | 8.53 | |
LVEF 30 days after surgery (%) | 36.50 | 16.86 | 0.144 |
LV diastolic dimension before surgery(mm) | 67.75 | 16.37 | |
LV diastolic dimension 30 days after surgery(mm) | 63 | 16.26 | 0.068 |
Vessels diseased | |||
Left main coronary artery | 13 (10.4%) | - | 0.183 |
Left anterior descending | 125 (100%) | - | 0.42 |
Right coronary artery | 113 (90.4%) | - | 0.283 |
Left circumflex artery | 108 (86.4%) | - | 0.115 |
Total bilirubin before surgery | 0.7 | 0.09 | 0.258 |
Total bilirubin after surgery | 0.99 | 0.27 | 0.331 |
Cerebrovascular disease | 20 (16%) | - | 0.136 |
Hypertension | 106 (84.8%) | - | 0.221 |
Unstable angina | 15 (12%) | - | 0.328 |
Previous myocardial infarction | 84 (67.2%) | - | 0.280 |
Hypercholesterolemia | 27 (21.6%) | - | 0.108 |
Diabetes mellitus | 98 (78%) | - | 0.245 |
History of PCI | 42 (33.6%) | - | 0.362 |
Smoking | 121 (96.8%) | - | 0.06 |
Peripheral vascular disease | 10 (8%) | - | 0.098 |
Alcohol consumption | 23 (18.4%) | - | 0.321 |
Preoperative renal insufficiency (creatinine > 1.3) | 45 (36%) | - | 0.25 |
LVEF: left ventricular ejection fraction; LV: left ventricular; PCI: percutaneous coronary intervention; SD: standard deviation
At the postoperative period, twelve patients (9.6%) had pneumonia and four (3.2%), atrial fibrillation. There were neither deaths nor mechanical complications due to the IABP catheter use.
In recent years, the development of drug therapy and percutaneous coronary insufficiency left for surgery the most severe cases. Frequently, surgery treats patients with advanced ischemic cardiomyopathy, low LVEF (<40%) and older age (>65 years), known as poor prognostic indicators. In order to improve the postoperative outcome of
these patients, the IABP preoperatively use is already gaining prominence, seeking to provide greater energy reserve for the intraoperative period of ischemia and improving postoperative cardiac function [
Nowadays, there is no agreement over which LVEF value would be benefited with the preparative IAB therapy. Some studies suggest that using preoperatively IAP has significant impact when LVEF is less than or equal to 40% [
There is no consensus on the best moment to install the IABP. While some groups do on the day before the surgery, others choose to install it right after anesthesia [
There are several methods to estimate LVEF, and the cub method, the Teicholtz and Simpson are most used. In this study, it was chosen the Simpson method that, unlike the other two, does not estimate the LV volume by a predetermined geometric shape. In shorts term, this volume is calculated by the multiple cavity fragmentation. Changes in segmental contractility often present in patients with coronary artery disease are encompassed in the calculations, making the LVEF estimated value more accurate. [
Christenson JT et al. demonstrated that patients treated with IABP in the preoperative had CPB duration lower than patients who did not [
The use of preoperative IABP in high-risk patients reduces the length of hospital stay compared to patients with IABP installed after surgery or when unused [
Cardiac performance improves immediately after myocardium revascularization. Besides, Christenson JT et al. displayed that the use of preoperative IABP increases this performance even before starting the CPB. After surgery, the gain is also more effective [
Last year, some studies observed that the mortality of the patients who received IABP in the preoperative of the myocardium revascularization surgery was 13.6%, while the mortality in patients who received the balloon in the trans or post-operative was, approximately, 35% [
Suzuki T detected that patients who received preoperative IABP had a lower incidence of ventricular arrhythmia than those who received IABP postoperatively [
Complications often encountered in the use of IABP are: lower limb ischemia, related to the insertion site, embolization of thrombi and thrombocytopenia for consumption. These complications are, mainly, related to the diameter of the balloon catheter, IABP usage time, prolonged hypotension periods and the presence of previous vasculopathy. Vascular complications arising from the IABP can reach rates of up to 20% [
In conclusion, due to the considerable evolution postoperative, as well as the improvement of LVEF and LVDD, the use of preoperative IABP is shown as an effective measure, safe and decisive in the treatment of patients with severe coronary artery disease and severe left ventricular dysfunction undergoing coronary artery bypass surgery.
The authors faced some limitations inherent in any observational study from a single center. The nonrandomized design might have procedural bias, or detection bias. The experience of the surgeon can influence the results of CABG: four experienced cardiac surgeons performed the CABG procedures in this study.
This study was conducted in the setting of a tertiary cardiovascular center in a developing country; the results might not be generalizable to other centers in developed countries. Patients requiring concomitant cardiac surgical procedures, or with mechanical complications of acute myocardial infarction, were excluded from this study and the results cannot be extended to these extreme high-risk patient populations.
Freitas, A.C.O., Gurian, D.B., Saporito, W.F., Machado, L.N., Horiuti, L. and Pires, A.C. (2016) Preoperative Intra-Aortic Balloon Counterpulsation in Coronary Artery Bypass Graft Patients with Severe Left Ventricular Dysfunction. World Journal of Cardiovascular Diseases, 6, 321- 328. http://dx.doi.org/10.4236/wjcd.2016.610036