Failure of an automated blood pressure cuff to deflate when a patient is under general anesthesia can lead to catastrophic consequences if unnoticed for more than three hours [1]. We present this as a hearsay case in which an automated blood pressure cuff of the Spacelabs Ultraview Clinical Workstation monitor (model No. 90385) applied pressure for about five hours resulting in limb thrombosis. In order to analyze this catastrophe, simulation scenarios were tested to elucidate the possible errors and malfunctions that may have led to this injury. We present the analysis of the advantages and validity of the hearsay case report. We also include our proposed criteria that should be required when a hearsay case is considered for publication.
The malfunction of an automated non-invasive blood pressure monitor (NIBP) which caused the cuff to remain inflated for greater than five hours caused the loss of an arm below the level of the cuff. While compartment syndrome is a recognized complication of NIBP monitoring [
We have heard through a clear second hand report that the events described herein actually occurred in 2011 although we do not know the exact cause of the failure to deflate. We also did not have access to the actual records. Of note, Spacelabs Healthcare terminated guaranteed service for this monitor in 2010 since it had not been manufactured for seven years [
The monitor involved in the actual case was the Spacelabs Monitor model 90385. In order to determine how these events may have occurred, with IRB approval, we have tested various conditions on ourselves with the Phillips Intellivue MP70 monitor to determine what clinical factors placed the patient at risk or obscured the threat to the clinician. This model was used was chosen for the simulation as it was available to us. We emphasize that the Phillips product was not involved in the actual case. The case was discussed with both Philips and Spacelabs technical support. The following scenario was developed so that a blood pressure cuff remaining inflated throughout a procedure under general anesthesia would go unnoticed by the clinician.
Internet searches were performed with two goals. First, the question of whether limb loss has been previously reported due to NIBP malfunction was addressed. Secondly, whether a hearsay report had been used to publish a medical complication prior to this report was also explored. The internet search engines used were Google, Pub Med, Bing, Yahoo search, AOL Search, Web Crawler, and Web MD with the keywords: Monitor, malfunction, hearsay, Compartment syndrome, and Non-invasive blood pressure. The depth of the search was five internet pages.
A patient was brought to the operating room for an operation under the care of an experienced anesthesiologist with major fluid shifts and the potential for hemodynamic instability. General anesthesia was induced without complication. The blood pressure cuff which had been applied correctly was then inflated on venipuncture mode to 60 mmHg in order to facilitate the placement of an additional intravenous catheter. Once the intravenous line was secured, as the fluid bag was about 84 cm higher than the NIBP cuff, the flow was brisk and the cuff never deflated. Subsequently, an arterial catheter was placed in the contralateral arm which also had been used for the pulse oximeter probe. Lastly an upper body warming blanket was secured, covering both arms. The blood pressure was recorded from the arterial catheter pressure reading. The MAP of “(60)” displayed under the NIBP reading was disregarded as an old NIBP reading.
After five hours, upon removing the surgical drapes, the arm with the NIBP cuff was inspected and noted to be dark, cold and pulseless, distal to the inflated blood pressure cuff. An emergent angiogram revealed extensive thrombosis of the limb’s arteries and veins. The vascular interventions to the limb proved to be futile; the arm was unable to be reperfused, dry gangrene set in and the limb was eventually amputated.
In the literature, compartment syndrome due to NIBP malfunction has been attributed to excessive cycling due to a tremor [
As always, when monitor failure is suspected, the clinician’s first priority is to ensure the patient’s well-being. Usually, monitor failure is only a distraction. Here the monitor failure posed a direct threat to the patient. Obviously, patients who have their arms tucked in at their sides would be difficult to inspect. Likewise, chest roll migration in a patient in the prone position has been reported by Lee et al. [
The three causes for the blood pressure cuff to not deflate are either that the blood pressure module malfunctioned, debris in the system obstructed deflation or the tubing was kinked. In the simulation scenario, the clinician relied on the arterial catheter pressure readings so the tubing, BP cuff and the module were never checked for continuous inflation. The Spacelab Manual explicitly describes that in event the cuff cannot be deflated, “Remove the cuff from the limb immediately and have the module serviced by biomed or a qualified field service engineer” [
Likewise, the Philips manual also mentions the possibility that “kinked or otherwise restricted tubing can lead to a continuous cuff pressure, causing blood flow interference and potentially resulting in injury to the patient.” In that manual they recommend to “Inspect the application site regularly to ensure skin quality and inspect the extremity of the cuffed limb for normal color, warmth and sensitivity. If the skin quality changes, or if the extremity circulation is being affected, move the cuff to another site or stop the blood pressure measurements immediately” [
Very low frequency events such as this should be reported in order to increase clinical vigilance to avoid their repetition. Fortunately, malfunction of a NIBP monitor is rare. The Thai Anesthesia Incidents Study estimated the incidence of equipment malfunction/failure at 3.4 per 10,000 [
The traditional medical case report requires a motivated author unafraid of the legal consequences of publishing the case involving a disastrous clinical event. A clinician after a clinical catastrophe may be liable to a potential civil lawsuit and so is unlikely to publish the case at least until the statute of limitations has expired. This delay may result in further adverse events. A catastrophic complication is not only traumatic to the patient and family, but to the clinician involved, a consequence of medical error that is all too often not addressed. This may make it emotionally difficult for the involved clinicians to take the time and effort to publish such an error. Wu and Steckelberg describe these physicians involved in medical error as “second victims” who do not have access to professional reaffirmation and personal reassurance that is offered to “first victims” such as the patient and family members [
Another reason a catastrophic case of instrument malfunction may be unlikely to be published as a case report is that it raises concerns about the maintenance of the equipment by its owners. The product involved is the Ultraview Clinical Workstation (UCW product #90385) which was last manufactured in January 2003 [
We propose that publishing hearsay case report is a way for clinical knowledge that has been communicated verbally to inform the wider audience of the written literature when hurdles exist for its immediate publication. Underreporting of adverse events occurs with the incident reporting system [
We believe this is the first case to be presented in the hearsay form. The authors recognize the potential problem that hearsay cases present to editors and journals in that they may become inundated with repetitious and facetious submissions of third or fourth hand, poorly described case studies. Hence we agree it should only be used sparingly and as a last resort.
For hearsay cases, the authors have developed eight criteria that must be first met before publication:
1) Authors must present a case based on factual events.
2) Authors must have a source connected to the parties involved of the actual case. The report must be second hand and not simply gossip of an unknown source.
3) Authors must be able to identify the location and approximate time of event.
4) The case reported must be unique and involve a previously undescribed issue. It should only be used to warn clinicians of a harmful situation. It should never be used to advocate a technique or a clinical management guideline without an accompanying expert consensus.
5) There must be clear and present danger to patients without the publication of the report.
6) Authors must present a case that is not likely to be soon published.
7) Authors must be able to identify key features of the event.
8) Authors must have no ulterior motives or mal-intent in the publication of the report.
An additional ninth criterion for the publication is proposed of what can be referred to as a “proxy case report”. The case report may also be reported by other clinicians who receive all the records from the original source. In the event that consent can be attained from one of clinicians involved in the case, then the publication should be considered as a report by proxy and thus would be superior to being reported as a hearsay report.
We propose that the one advantage of a hearsay report over closed claims analysis is its rapidity. The strength of the laudable closed claims study analysis lies in the analysis by an impartial third party of the closed claims as well as the ability to “overcome” the hesitancy to publish cases that have poor outcomes. The closed claims analysis depends on the case entering the closed claims database. Years may pass before these critical events are examined. Some injured patients may not even file claims. A hearsay report may inform clinicians without delay of waiting for a case to be settled or statute of limitations to expire. Closed claims studies can take at least 2 years to 5 years between the actual occurrence of the critical event and the actual closure of the associated claim in order for it to be available for review [
Clinicians are inherently historians. We are taught from the beginning of our training to record clinical observations. This training emphasizes accuracy in details. Hearsay to the clinician is considered a poor substitute for a first person clinical report. While the hearsay case report is an inferior substitute to a first-hand case report, we intend it here to be an early warning for clinicians. We have defined the criteria for a hearsay case so it is not simply gossip transcribed. For the medical editor, this is a concern that the acceptance of hearsay reports will lead to a proliferation of poor, inaccurate case reports. We believe there is a role, albeit a rare one for the publication of a hearsay case.
In conclusion, five clinical events had to occur so that the inflated blood pressure cuff would go unnoticed by the experienced clinician. This is consistent with the Swiss cheese model of system accidents where multiple hazardous conditions are present when an adverse clinical event occurs [
Steven M.Shulman,YunseokNamn,StanislavLando,PatrickDiscepola, (2015) Forearm Loss Caused by Automated Non-Invasive Blood Pressure Cuff Malfunction: A Hearsay Report. Open Journal of Anesthesiology,05,227-232. doi: 10.4236/ojanes.2015.511041