In the medical field, there are growing interests in applied research such as in vivo fluorescence monitoring because of excellent body transmission characteristic of the near-infrared light. However, optical noise by excitation light and illumination equipment for medical applications such as interior light, surgical light decrease efficiency of the fluorescent signal when observers such as surgeons confirm fluorescence signals in medical field. To solve these problems in medical field, we have analyzed external noise factors by effect on image realization, quantification of optical noise generation by external factors, and have suggested methods of minimize the optical noise in this paper. In case of fluorescence imaging in the operating room, it has been confirmed that fluorescent excitation light, interior light and surgical light are factors to generate optical noise. To acquire near-infrared fluorescence images and to compare fluorescence contrast under conditions of darkroom, interior light and surgical light, light emitting diodes (LEDs) sources that have peak wavelength at 740, 760 and 780 nm respectively were used as excitation light sources. In addition, short-pass filter which has transmission edge at 775 nm has been applied to minimize the optical noise in each external noise factor. By comparing contrast of each image before and after use of the short-pass filter, we confirmed that optical noise reduced 49%, 56% and 66% in external noise factors respectively.
Fluorescence imaging technique uses the emission of light by a substance that has absorbed light or other electromagnetic radiation. In most cases, the emitted light has a longer wavelength, and therefore lower energy, than the absorbed radiation. However, fluorescence imaging technique has many advantages such as high contrast, that is, signal to noise ratio (SNR), high sensitivity and simple optical instrumentation and so on [
Recently, near-infrared (NIR) fluorescence monitoring in the range of 700 - 900 nm is being utilized in the bio-medical field actively because NIR has high penetration characteristic in hemoglobin, water and lipid etc. [
In addition, Indocyanine green (ICG) which emits NIR light (center wavelength: 830 nm) has been used in target protein application since ICG was granted to use in clinical research by Food and Drug Administration (FDA) [
In particular, ICG is applied to monitor lymph nodes and blood vessels pertaining to evaluation of cerebral blood flow, liver, and circulation function because of the low toxicity and the high absorption rate of ICG in range of 600 - 900 nm [
Additionally, Beth Israel Deaconess Medical Center in USA and Hamamatsu in Japan developed FLARETM (Fluorescence-assisted resection and exploration) system and PDE (Photodynamic Eye), respectively. They have used the devices to image identification of the sentinel lymph node (SLN) and the metastasis of malignant tumors like the breast cancer and the stomach cancer [
Although preclinical/clinical studies with the NIR fluorescence imaging system have briskly proceeded, stu- dies to improve performance of the NIR fluorescence imaging such as elimination of optical noise, improvement of optical source etc have not briskly proceeded.
Especially, optical noise decreases NIR fluorescence image quality because optical noise reduces effectiveness of NIR fluorescence signals and disturbs detection of minute signals in preclinical/clinical studies. In order to minimize optical noise realized fluorescence image in a dark room. So, studies to minimize optical noise in NIR fluorescence imaging should be considered in preclinical/clinical environment necessarily.
In this paper, we suggest conditions to minimize optical noise by the fluorescent light source, the interior light and the surgical light in the operation room.
A NIR fluorescence imaging system was configured to know how optical noise factors influence during realization of NIR fluorescence imaging as shown in
A band-pass filter (#84-107, Edmund Optics) with 832 nm ± 20 nm bandwidth and a charge-coupled device (CCD) camera (Guppy Pro F-031B, Allied Vision Technology) with a small lens (#58-000, Edmund Optics) were used in the fluorescence imaging system.
In addition, an optical light source composed of 16 light emitting diodes (LEDs) with each 1 W optical power for fluorescence excitation was used in the fluorescence imaging system. Three kinds of LEDs which have the center wavelength of 740, 760 and 780 nm were prepared for comparison of fluorescence signals according to the wavelengths of the LEDs.
For the expression of the NIR fluorescence ICG (Diagnogreen, DAIICHISANKYO CO., Osaka, Japan) with the maximum absorption at 780 nm and the maximum emission at 830 nm was resuspended in 10 cc of sterile water to yield a 0.8 μl/ml stock solution.
A circular gauze coated with ICG solution (10 mm in diameter) which was put on a circular gauze without ICG solution (30 mm in diameter) was prepared as a sample for measurement of the NIR fluorescence.
The fluorescence light source, the interior light and the surgical light were selected as optical noise factors in NIR fluorescence imaging.
Most of all, spectra or the fluorescent light source, the interior light and the surgical light that were selected as optical noise factors were measured to verify influence of the factors on optical noise. The NIR fluorescence images under three optical noise factors were acquired and the acquisitions repeated after applying a short-pass filter to minimize optical noise.
Spectra of three kinds of LEDs with center wavelength of 740, 760, 780 nm, the interior light and surgical light were measured by using a spectrometer (USB 4000, Ocean Optics).
Measurement results represent as shown in
First, NIR fluorescence imaging proceeded with three kinds of LEDs in the darkroom in order to verify optical noise effects of the LEDs, excitation light sources.
Second, NIR fluorescence imaging proceeded with three kinds of LEDs under the interior light in order to verify optical noise effects of the interior light.
Finally, NIR fluorescence imaging proceeded with three kinds of LEDs under the surgical light (C100, 1700 lux, Luvis, Korea) in order to verify optical noise effects of the surgical light.
In the NIR fluorescence imaging, we controlled output powers of 740, 760 and 780 nm LEDs as 270 μW, respectively. Then, we acquired NIR fluorescence imaging with the circular gauze sample.
For comparison among the NIR fluorescence images obtained under the each condition drawing, we calculated the average value in a region of interest (ROI) of the images.
Then, we compared the contrast defined as Equation (1) by the expression of fluorescence signal (IFS, FS: fluorescence signal) intensity and background signal (IBS, BS: background signal) intensity in the NIR fluorescence image.
775 nm short-pass filter (#64-615, Edmund Optics) having the optical characteristics as shown in
used to the minimized optical noise by the optical noise factors.
First, we measured spectrum of the fluorescent light source after applying the short-pass filter to the fluorescent light source having a center wavelength of 740, 760 and 780 nm in the darkroom. Also we obtained the fluorescence image of the gauze sample by using the three light sources with the short-pass filter.
Second, we applied the short-pass filter to the interior light and the fluorescent light sources having a center wavelength of 740, 760 and 780 nm.
Then, we measured spectrum of the interior light and obtained the fluorescence image of the gauze sample by using the three light sources with the short-pass filter under the interior light with the short-pass filter.
Third, we applied the short-pass filter to the surgical light and the fluorescent light sources having a center wavelength of 740, 760 and 780 nm.
Then, we measured spectrum of the surgical light and obtained the fluorescence image of the gauze sample by using the three light sources with the short-pass filter under the surgical light with the short-pass filter.
Finally, we obtained average signals about two regions shown in
Spectral measurement results of each excitation light source showed that LED light sources with center wavelength of 760 and 780 nm except for LED light source with center wavelength of 740 nm have spectral region more than 810 nm. Therefore, it was confirmed that there is overlap between spectrum of light sources with center wavelength of 760 and 780 nm and emission spectrum of ICG.
In addition, spectral measurement results of the interior light and the surgical light show that they have spectral region more than 810 nm although optical intensity at the spectral region is minute.
For analysis of optical noise in fluorescence imaging, output power of excitation light sources with center wavelength of 740, 760 and 780 nm was identically prepared as 270 μW and output power of the surgical light output power was prepared to control as 0.85, 1.08, 1.81, 2.41, 3.14 and 3.68 mW/cm2.
To measure and analyze optical noise factors in fluorescence imaging, we acquired NIR fluorescent images by using LED excitation light sources with center wavelength of 740, 760 and 780 nm in dark room and we repeated NIR fluorescent imaging under a interior light and a surgical light, respectively.
In case of the excitation light source with center wavelength of 740 nm, the fluorescence image had an average intensity of 97.67 and contrast of 0.94 in the dark room.
In case of the excitation light sources with center wavelength of 760 and 780 nm, each average intensity increased to 209.17 and 255.0 and each contrast decreased to 0.45 and 0.00 beside case of the excitation light sources with center wavelength of 740 nm.
In case of the excitation light source with center wavelength of 740 nm, the fluorescence image had an average intensity of 113.16 and contrast of 0.75 under the interior light.
In case of the excitation light sources with center wavelength of 760 and 780 nm, each average intensity was 228.79 and 255.0 and each contrast was 0.41 and 0.00, respectively.
The fluorescence images were saturated from 1.81 mW/cm2 intensity in case of the excitation light source with center wavelength of 740 nm and were saturated from 0.85 mW/cm2 intensity in case of the excitation light source with center wavelength of 760 and 780 nm under the surgical light.
We placed the short-pass filter in front of excitation light sources (740, 760 and 780 nm) and acquired fluorescence images of the circular gauze coated with ICG solution by using LED excitation light sources with center wavelength of 740, 760 and 780 nm in the dark room, respectively.
Average fluorescence intensity from each fluorescent image was 77.52 (in case of the 740 nm light source), 87.83 (in case of the 760 nm light source), 103.74 (in case of the 780 nm light source) and each contrast was 0.97 (in case of the 740 nm light source), 0.97 (in case of the 760 nm light source), 0.97 (in case of the 780 nm light source). We could confirm that each optical noise decreased 3% (in case of the 740 nm light source), 49% (in case of the 760 nm light source), 93% (in case of the 780 nm light source) by using the short-pass filter, respectively.
In addition, we placed the short-pass filter in front of excitation light sources (740, 760 and 780 nm) and the interior light and then we acquired fluorescence images of the circular gauze coated with ICG solution by using LED excitation light sources with center wavelength of 740, 760 and 780 nm under the interior light, respectively.
Finally, we placed the short-pass filter in front of excitation light sources (740, 760 and 780 nm) and the interior light and then we acquired fluorescence images of the circular gauze coated with ICG solution by using LED excitation light sources with center wavelength of 740, 760 and 780 nm under the surgical light, respectively.
From
that each contrast was in value from 0.55 to 0.8 and that it is possible to distinguish the background from the fluorescence at 0.85 mW/cm2 power of surgical light when the short-pass filter referred to the light source and the surgical light.
In this paper, we have suggested that LEDs as the excitation light, the interior light and surgical light can degrade light source and the external.
To analyze influence of the excitation light, the interior light and surgical light as optical noise factor, we acquired and compared fluorescence images by using the ICG and NIR fluorescence imaging system under various experimental conditions such as no light, interior light and surgical light.
An experiment was carried out by using LEDs with center wavelength of 740, 760 and 780 nm as the excitation light source.
From measured results of spectrum and intensity of the excitation light source, the interior light and the surgical light with review of the optical noise occurrence factor, we could confirmed that the optical noise is generated because of the overlapping between wavelength band of the optical noise factors and wavelength band of fluorescence signal. To solve the optical noise problem, we applied the short-pass filter to optical noise factors.
We used the short-pass filter which has cut-off edge at 775 nm wavelength to minimize influence by optical characteristics of the optical factors such as intensity decrease.
We could confirm that contrast of fluorescence images improved 49%, 56% and 66% by using the short-pass filter as compared with contrast of fluorescence imaging in case of no short-pass filter.
In research results of other groups, the LED excitation light source with center wavelength of 760 nm or 770 nm was considered as an optimal light source in case of ICG based on NIR signal detection [