Intelligent Control and Automation
Vol.05 No.04(2014), Article ID:51353,8 pages
10.4236/ica.2014.54027

Feasibility Assay for Measure of Sternocleidomastoid and Platysma Electromyography Signal for Brain-Computer Interface Feedback

Juliet Sánchez Galego, Omar Valle Casas, Alexandre Balbinot

Electrical Department, Federal University of Rio Grande do Sul, Porto Alegre, Brazil

Email: aulaeng@gmail.com, alexandre.balbinot@ufrg.br

Copyright © 2014 by authors and Scientific Research Publishing Inc.

This work is licensed under the Creative Commons Attribution International License (CC BY).

http://creativecommons.org/licenses/by/4.0/

Received 18 September 2014; revised 8 October 2014; accepted 17 October 2014

ABSTRACT

A feasibility assay is conducted for electromyography measure in sternocleidomastoid and platysma, tenting to use it on Brain-Computer Interface (BCI) feedback. It is proposed a case of study for four healthy subjects with an average of 35 years old, two females and two males. Methodology proposed includes signal acquisition and processing with feature extraction of RMS, Mean and Variance. The data are acquired with the AD board NI USB-6009, interfaced with LabView and pro- cessed in MatLab. An uncertainty analysis was made obtaining a system uncertainty of ±2.31 mV. ANOVA analysis was done, with a Randomized Complete Block Design (RCBD) experiment and in- teraction of factors and residues obtained with the software Minitab.

Keywords:

Electromyography, Feature Extraction, Uncertainty, Statistical Analysis, Factors Interactions

1. Introduction

Electromyography (EMG) is an experimental technique concerned with the development, recording and analysis of myoelectric signals. Myoelectric signals are formed by physiological variations in the state of muscle fiber membranes. The smallest functional unit to describe the neural control of the muscular contraction process is called a Motor Unit. The EMG-signal is based upon action potentials at the muscle fiber membrane resulting from depolarization and repolarization processes. The extent of this Depolarization zone is described in the lite- rature as approximately 1 - 3 mm2. Because a motor unit consists of many muscle fibers, the electrode pair “sees” the magnitude of all innervated fibers within this motor unit—depending on their spatial distance and resolution. Typically, they sum up to a Motor unit action potential (MUAP), which differs in form and size depending on the geometrical fiber orientation in ratio to the electrode site [1] .

A Brain-Computer Interface (BCI) system allows to communicate humans and devices without need of sen- sors in others parts of the body or the muscular system. Although, they are a promising tool for persons with se- vere palsy conditions like neurodegenerative. Facial electromyography (EMG) contamination of the electroen- cephalography (EEG) signals is a largely unresolved issue in brain-computer interface (BCI) research. Mu and beta rhythms are widely used in the literature and they lie in the frequency range that is susceptible to electro- myography (EMG) contamination [2] . Also, signals from the muscles related to the movement can provide a feedback system, with precious information for taking right decisions concerning for example, to a wheelchair movement. That’s why the main objective of this research is to measure the EMG signal of the sternocleido- mastoid (SCM) and platysma (PM) muscles in order to validate the feasibility of their use. The MUAPs within the pick-up area of electrodes can be observed noninvasively by using electrodes affixed to the surface of the skin. These signals are stochastic in nature and can be represented by a Gaussian distribution function. The am- plitude of signal ranges from 0 to 10 mV (peak to peak) or 0 to 1.5 mV (RMS). The usable energy of signal is limited to 10 to 500 Hz frequency range, with dominant energy being 50 to 150 Hz range [3] .

Literature recommends band pass filtering settings from 10 Hz high-pass up to at least 500 Hz low pass; most of the surface EMG frequency power is located between 10 and 250 Hz. After this, other techniques are applied to prepare the signal for feature extraction; therefore these characteristics are introduced in computational intel- ligence systems (e.g., Support Vector Machines, Artificial Neural Networks and Linear Discriminant Analysis) that turns the signal information into the movement that is been performing [4] [5] .

This paper presents an experimental section with the methodology used to obtain the data, a short overview of the materials employed and the data processing, an approaching to the uncertainty estimation of the measuring system and the discussion of the results.

2. Experimental Section

In this research we want to validate if three of the characteristics used in computational methods, obtained by EMG feature extraction techniques, change in presence of a paradigm, and by changing the electrode’s position. Let’s consider three factors that may influence the response variable (EMG characteristic): Paradigm (A), Sub- ject (B) and Position (C).

We propose a Randomized Complete Block Design (RCBD) experiment, in which we fix the B factor and produce a complete randomization for each block, containing all the treatments [6] .

The paradigm factor has four levels corresponding to the movements of the neck: Right (1), Left (2), Ahead (3) and Back (4); and related with the basics actions that a wheeling chair user has to accomplish. The third fac- tor, position of the electrodes has four levels because the interest is to validate different positions related to the head’s movement like: PM—Right (1), PM—Left (2), SCM—Left (3) and SCM—Right (4). The second factor is subjects and is blocked because the lack of time between measures that could compromised the veracity of the experiment, that’s why this factor was blocked and the trials randomized for each subject: B1 (1), B2 (2), B3 (3) and B4 (4).

Finally, is proposed a case of study for four healthy subjects with an average of 35 years old, two females and two males having no history of muscle problems, as described by the methodology of the essay (Figure 1).

2.1. Materials Description

To obtain useful information from EMG signal it is required that any detecting and recording device processes the signal linearly, that’s why after the surface electrodes, the signal is presented to an active electrode with dif- ferential amplification (Instrumentation Amplifier) for more than 500 times before the analog filtering stage (Electromyograph). At this moment, the signal is digitalized by an A/D board for the computer acquisition con- trolled by an interface application (LabView) as seen on Figure 2.

Eight electrodes are located in the muscles of interest, two at right and two at left by muscle (SCM, PM), adding the reference electrode in the forehead (Figure 3). At the moment of acquisition, it’s only recorded the channel of the combination that it’s been evaluating. The application interface is set to a sampling frequency of 1000 Hz and about nine seconds of time are recorded (9000 samples). Using a function implemented in MatLab,

Figure 1. Methodology diagram of the assay propose for the EMG measure.

Figure 2. Block diagram of measuring channel of EMG.

Figure 3. Subject’s neck showing electrodes positions.

the measuring combinations are rearranged inside a single subject’s trial, granting the randomization inside the block of subject’s trial (e.g., Paradigm Right with SCM-Left Position measuring). The movement to be fallow by the subject is displayed in the center of the interface, and the muscle reaction is recorded.

All the data is recorded in archives and they are processed later by the MatLab function created for the test. This function loads the archive containing the register and at first it’s filter with a FIR filter band pass with but- ter approximation. Fourier transform is applied to the register for obtain the spectrum before and after filtering.

The register’s offset component is eliminated by its calculation and subtracted from it (Equation (1)); also this vector is normalized by the computation of the register’s maximum and its division (Equation (2)). The mean is obtained and used for the threshold’s determination according with the literature and the research group results [7] .

(1)

(2)

Depending of the value of the register’s maximum the threshold is establish and used a posteriori for the valid windows determinations by comparing the window’s power with the threshold obtain [8] . After eliminate the offset and normalize, we have the next register in time (Figure 4). For the windowing method is selected a window of 250 ms, and the validation of movement in the window is possible by comparing the threshold with the power calculated in Equation (3). Then, the EMG’s characteristics: RMS, Mean and Variance are obtained from the valid windows. At the end we get these values as we can see in Figure 5. Finally the features are cal- culated and store into a table for ANOVA and statistical analysis.

Figure 4. Register of EMG normalize with Mean (Xmed) and Threshold (Th).

Figure 5. Window of EMG of valid movements with RMS, Mean (Xmed) and Variance (XVar).

2.2. Uncertainty Estimation

The uncertain analysis for the measuring channel is done taking into account the principals sources uncertain and errors. A classical analysis is proposed making an estimation of the majority of the source represented in the cause-effect diagram from Figure 6. As seen in diagram of Figure 6, there are five principal sources that con- tribute to the uncertain estimation on the system. For the estimation of the Instrumentation Amplifier (IA) con- tained in each active electrode (four wires for the experiment) and the electromyograph voltage characterizations were developed, measuring with a Digital Oscilloscope which uncertainty is too evaluated. The A/D board is a 13 bits USB data acquisition, which resolution is reviewed. The electrodes are MediTrace ECG Conductive Ad- hesive Electrodes disposables and besides are identified as an error source, and knowing that has a normal dis- tribution, we don’t have information for their uncertain estimation.

First of all let’s analyze the IA:

Affectation of common mode voltage to 60 Hz (CMRR): This source has it origin in the common mode voltage increasing when is in presence of 50 Hz or 60 Hz of industrial national power supply [8] . Let’s assume an un- certainty Type B with rectangular distribution:

(3)

Repeatability of the measure (IA): One group of measures is analyzed for the IA characterization at 150 Hz. The entrance had been changed to obtain the group of values of. Let’s assume an uncertainty Type A with normal distribution:

(4)

Repeatability of the electromyography (EMG): For this device that is compound by four operational in filter implementation, a characterization for five frequencies (50, 70, 100, 150 and 200 Hz) is developed obtaining data at that support a gain very close to 1, and maintain plane response at these frequencies. Let’s assume an uncertainty Type A with normal distribution:

(5)

Resolution of the Oscilloscope vertical channel (OSCRES): only is study this incidence because that the kind of measures are been made are in voltage. Let’s us assume an uncertainty Type B with rectangular distribution.

(6)

Calibration Certificate of the Oscilloscope vertical channel (OSCCAL): it owns a calibration certificate that introduces to the system an uncertainty Type B with normal distribution [9] . For a 95%:

Figure 6. Cause-effect diagram of the uncertainty sources from the EMG measuring system.

(7)

Resolution of the board NI USB-6009 (AD): For this analysis we calculate the minimum detectable change (Mdc) due to the resolution of the system [10] . As the board is used as single-ended, it has a resolution of 13 bits, and a dynamic range of ±10 V (Edr). Let’s us assume an uncertainty Type B with rectangular distribution.

(8)

The combined uncertainty was obtained by:

(9)

(10)

As the is superior to 30, we calculate the coverage factor by the normal table, looking for the value for a confidence level of 95%, it will be for an expanded uncertainty like Equation (11):

(11)

After this we can elaborate a table with the principal uncertainties of the system (Table 1). Finally the data acquire by the system can be expressed with an uncertainty of.

3. Results and Discussion

First of all, we corroborate that data obtained fallows a Normal distribution and are independents. To achieve this, we build a histogram for each of the features (Figure 7 shows only one analysis). The graph’s likeness to normal behavior is considered sufficient as premise to perform an ANOVA analysis. According with the ob- tained data from the feature extraction we can form tables for the three characteristics. This analysis, also named variance analysis, allows determining which factors and combinations of they, are related [6] . To have a better picture of data behavior, residual plots are obtained (Figure 8 shows only one analysis).

In the graphic of Figure 8, the residual histograms show a normal behavior. The residual plots versus fits, systematic patterns are not evident and random pattern of residuals is seen on both sides of 0. As was defined in

Table 1. Uncertainty data estimation resumes of the system sources.

Figure 7. Histogram of the EMG extracted features: RMS.

Figure 8. Residual plots for the EMG extracted features: RMS.

Section 2, a RCBD experiment is made were the B factor (Subjects) is blocked. In this case, the response variable RMS of the EMG signal change with the A, C and AC factor’s combination. Other evaluations are made for the EMG Mean and EMG Variance. The response variables Mean and Variance of the EMG signal, change with the A, C and AC factor’s combination, that means, the A, C factors and their interactions are significant, and further statistical analysis can be taking for the three characteristics obtained. The factors Paradigm (A) and Position (C) can be plotted for their interactions analysis for the three characteristics.

As we can see, Figure 9(a) (RMS) shows that the Paradigm’s level: 3 and 4, move ahead and back respec- tively, doesn’t interact, changing equally with the position; opposed to the behavior of the others two Paradigms’

(a)(b)(c)

Figure 9. Interactions plots between the paradigm and the electrode’s position for the EMG extracted features: (a) RMS; (b) Mean; and (c) Variance.

level: move Right (1) and Left (2), that have a crossover interaction, showing variation with the electrode’s po- sition. It means that the lateral movements (Right and Left) are more influents than the other two (Ahead and Back). We can appreciate also how Positions 1 (PM—Right) and 2 (PM—Left), change equally with the four paradigms, contrary to the other two: 3 (SCM—Left) and 4 (SCM—Right) that interact in this values with all the paradigms. This last aspect is also appreciated in Figure 9(b) (Mean) and Figure 9(c) (Variance). It could indicate that SCM muscle is most influent for EMG signal acquisition for the paradigm’s translation, than PM.

4. Conclusion

This research tries to validate the feasibility of the electromyography of the study muscles and the paradigms tested. In the attempt, a new interface for signal acquisition was developed, as well as a code implemented, for feature’s extraction like RMS, Mean and Variance. Instead of the fact that the influence of the paradigm and po- sition factors over the three response variables is significant, the results showed that the obtained sternocleido- mastoid muscle signal is much stronger and more significant, than the one acquire from the Platysma. Also the paradigms of lateral movements have more interactions than the other levels, suggesting a paradigms change. It is recommended further assays with other paradigms including the lateral movement’s studies in this research, and the use of the SCM muscle for BCI feedback.

References

  1. Peter, K. (2005) ABC of EMG: A Practical Introduction to Kinesiological Electromyography. Noraxon INC, USA.
  2. Xinyi, Y., Rabab, K.W. and Gary, E.B. (2008) Facial EMG Contamination of EEG Signals: Characteristics and Effects of Spatial Filtering. Proceedings of the ISCCSP 2008, Ottawa, 12-14 March 2008, 729-734.
  3. Teena, G., Shalu, G.K. and Sivanandan, K.S. (2011) Processing and Application of EMG signals for HAL (Hybrid As- sistive Limb). Proceedings of the 2nd International Conference on Sustainable Energy and Intelligent System (SEISCON 2011), Chennai, 20-22 July 2011, 749-753.
  4. Kazuo, K. and Yoshiaki, H. (2011) Motion Estimation Based on EMG and EEG Signals to Control Wearable Robots. Proceedings of the 2013 IEEE International Conference on Systems, Man, and Cybernetics, Japan, 4213-4218.
  5. Elaine, A.C., Nicholas, A.S. and Eric, J.P. (2013) EMG Control of Robotic Reaching by People with Tetraplegia Im- proved through Proprioceptive and Force Feedback. Proceedings of the 6th Annual International IEEE EMBS Conference on Neural Engineering, San Diego, 6-8 November 2013, 1178-1181.
  6. Douglas, C.M. (2013) Randomized Blocks, Latin Squares, and Related Designs. In: Design and Analysis of Experiments, 8th Edition, John Wiley & Sons, Inc., Hoboken, 139-146.
  7. Calibration of Measuring Devices for Electrical Quantities Calibration of Oscilloscopes. EURAMET (European Asso- ciation of National Metrology Institutes) cg-7, Version 1.0 (06/2011). Calibration Guide. Bundesallee 100, D-38116 Braunschweig, Germany. http://www.euramet.org/
  8. Balbinot, A. and Favieiro, G. (2013) A Neuro-Fuzzy System for Characterization of Arm Movements. Sensors, 13, 2613-2630.
  9. Tektronix (1998) Measurement Products Catalog 1998/1999, Tektronix.
  10. (2012) USER GUIDE AND SPECIFICATIONS NI USB-6008/6009: Bus-Powered Multifunction DAQ USB Device, National Instruments.