The Impact of Small Field’s Off-Set on Output Factor in Intensity Modulated Radiation Treatment

Hulya Ozdemir1,2, Nina Tuncel1*, Adem Unal Kizildag3

1Physics Department, Faculty of Science, Akdeniz University, Antalya, Turkey

2Gammaray Private Health Services, Istanbul, Turkey

3Elekta Medical Systems, Istanbul, Turkey

Copyright © 2017 by authors and Scientific Research Publishing Inc.

This work is licensed under the Creative Commons Attribution International License (CC BY 4.0).

http://creativecommons.org/licenses/by/4.0/

Received: October 22, 2017; Accepted: November 20, 2017; Published: November 23, 2017

ABSTRACT

In intensity modulated radiation treatment (IMRT) planning, the use of asymmetrically collimated fields that are placed on central axis or its off-set is mostly required. Output is the main topic discussed today for extremely small and/or severe irregularly shaped fields. The air scatter data are involved directly or indirectly in obtaining the output. Despite the fact that extensive data have been published in many studies to provide a guide on the magnitude of output factor for clinical accelerators, there are very few data reviewed about output factor in-air or phantom for off-set fields. This study was aimed to investigate the impact of these conditions for small fields. This study was conducted in Elekta Synergy linear accelerator which produces 6 MV X-ray energy. The in-air output factor (Sc) has been measured by CC04 ion chamber with brass-alloy “build-up” cap and Dose-1 electrometer, and the total output (Scp) measurements were carried on at dose maximum depth in phantom by the same chamber and Thermoluminescence dosimeter (TLD) for 1 - 10 cm2 fields. The all measurements at center of isocenter and off-set fields at three directions (X2, Y1, Diagonal) were done. By decreasing field size from 10 to 2 cm2 at isocenter, the Sc value using CC04 was decreased to 5.4% and Scp using CC04 and TLD to 14.5% and 11% respectively. By increasing off-set value, the Sc and Scp values were increased in all directions comparing to central fields. The maximum increase was obtained in Y1 direction for Sc and Scp. TLD results for Scp is slightly higher than CC04. The dosimetric properties of small fields and their off-set should be evaluated and modelled appropriately in the treatment planning system to ensure accurate dose calculation in Intensity Modulated Radiation Treatment.

Keywords:

Small Field, Off-Set Fields, Total Output Factor, In-Air Output Factor,

1. Introduction

Generally, in radiation therapy (RT), the reliability of treatment planning system that based on factors or model is affected by many factors. These may include such phenomena as scattered photons by structures in the linear accelerator head (head-scatter), backs catered photons and electrons into the monitor chamber (monitor backscatter), and the partially obscuring of the X-ray source by the collimators (source-obscuring effect) especially in very small fields [1] . Various sources of head-scatter, which include in the primary collimator, the flattening filter, the secondary collimators, the monitor chamber and a wedge, if used, have been considered [2] [3] [4] . The size of collimator aperture describes the X-ray output and multi-leaf collimator (MLC), which has a critical role for shaping the radiation field.

The concept of output for a field size describes the absorbed dose for X-ray produced by linear accelerators (LINAC). The in-air X-ray scatter component composes the main part of output. But in dosimetry procedure of RT, this concept refers to absorbed dose to medium or water. It could be claimed that the medium (water or phantom) scatter contains a minor part of output. Output is the main topic discussed today for extremely small and/or severe irregularly shaped fields [5] . The air scatter data are involved directly or indirectly in obtaining the output. Generally, the collimator arrangement of any LINAC consists of fixed and movable structures. The Flattening Filter which constitutes one of the fixed structures in any LINAC is placed between the primary collimator and the monitor chamber, and its main role is to make the photon beam dose distribution uniform at a reference depth [4] . The X-Y jaws and MLC with unidirectional motion that play a major role in field shaping constitute the moving structures of LINACs. The concept of in-air output ratio [6] [7] [8] [9] is characterized by the variation in the incident photon fluence per monitor unit according to collimator settings, which is known as collimator scatter factor [10] or head scatter factor (Sc) for photon [11] [12] . The “In-air output factor” that contains the latest two definitions is commonly used and usually symbolized as Sc. It includes the source-obscuring effect, the head-scatter and the monitor backscatter effect. Many recommendations are given for using small size detector for fields smaller than 3 × 3 cm2, because significant differences can be found among using different detectors. The underestimation of output factors due to the increase of lateral electron disequilibrium and the volume effect of detectors is expected [13] [14] [15] .

In intensity modulated radiation treatment (IMRT) planning, the use of asymmetrically collimated fields that is placed on central axis or its off-set is mostly required. Generally, the initial part of IMRT dose calculations is an energy fluence optimization that is directly related to Sc measurements. The head scatter factor (in-air output ratio, Sc) is defined as the ratio of primary collision water kerma in air per monitor units (MU) at isocentric distance (100 cm) for a given field setting to that of a reference 10 × 10 cm2 field. Accurate determination of Sc for IMRT is much more challenging, where extremely small and/or severe irregularly shaped fields are more commonly used [5] . The Sc emphasizing a component of the output concept varies with field size at related photon energy.

The treatment planning system’s software that could calculate dose distributions for more complex treatment containing configurations is directly related to head scatter modelling, output factor determination, and monitor unit calculations, respectively. The applied algorithms are based on either empirical methods, analytical models, or a combination of both [16] . The extensive data have been published in many studies to provide a guide on the magnitude of output factor for clinical accelerators. But, there are very few data have been reviewed about output factor in-air or phantom for off-set fields [17] . This study was aimed to investigate the impact of off-set conditions on in-air and total output factor for small fields in IMRT.

2. Methods and Materials

2.1. Dosimetric Equipment and Processes

This study was conducted in Elekta Synergy linear accelerator that produced 6 MV X-ray energy. The head diagram of Elekta (TM) linear accelerator (Elekta AB, Stockholm, Sweden) is equipped by MLC with 1 cm width leaf for providing shaped fields (Figure 1) and also small fields for performing step and shot intensity modulated radiation treatments. Each leaf can travel 12.5 cm over the beam central axis regarding shifting to ×2 axis direction. And it is contained flattening filter.

The small volume compact ionization chamber was preferred to use in small fields and high dose rates measurements.

The high spatial resolution cylindrical ion chamber (Figure 2), CC04 (IBA, GmBH, Scanditronix Wellhofer, Germany) with cavity volume of 0.04 cm3 was used for measurements [19] .

The Sc measurements were performed by this chamber with brass-alloy “build-up” cap that designed for 6 MV X-ray energy and Dose-1 electrometer. This “build-up” cap was made of brass alloy (Copper, Nickel, Zinc, Lead, Tin, Manganese and Iron) with 8.62 g/cm3 density. Its outer diameter is 13.6 mm with 4.4 mm wall thickness.

The Scp measurements at dose maximum depth (1.6 cm) in RW3 solid-water phantom (IBA, GmBH, Germany) by the same ion chamber and electrometer were carried on for 1 × 1 - 10 × 10 cm2 fields. This arrangement was used for Scp

Figure 1. The head diagram of Elekta (TM) Linear accelerator head [18].

(a)(b)

Figure 2. (a) Scanditronix-Wellhöfer CC04 compact ion chamber and its diagram; (b) Brass-alloy “build-up” cap [19] .

measurement by Thermoluminescence dosimeter (TLD). LiF-100 (LiF: Mg, Ti) disc-shaped crystals have 4.5 mm in diameter, 0.9 mm in thickness (MTS-N Poland). The two TLD discs were embedded in a RW3 solid-water phantom plate for obtaining adequate measurement depth. RADOS RE-2000RT (RadRpo Int. GmbH Germany) with RADOS TLD Server software was used for TLD readings. In this study, TLDs’ with ±3% sensitivity were selected. TL-count conversion to dose was done for 6 MV energy using the dose of 10 × 10 cm2 field, SSD = 100 cm at dDmax.

2.2. In-Air and Total Output Factor Measurements for Central Fields and Off-Set Fields

Computer-aided water phantom was used to minimize the displacement error and the geometric shift for ion chamber matching with isocenter for Sc. The chamber with brass-alloy cap that paralleled to beam was placed in the empty phantom at the center of field (Figure 3). The source and tip of “build-up” cap distance was set to 100 cm. The output measurements per 100 MU for 1 × 1, 2 × 2, 3 × 3, 4 × 4, 5 × 5, 6 × 6 and 10 × 10 cm2 fields were performed. The corrected for both pressure and temperature readings by CC04 chamber with Dose-1 electrometer were obtained as a Gray (Gy).

The output measurements for off-set fields called OSF procedure was the same as Sc in-air output factor measurement while the field location was placed on centre of off-set fields at three directions (X2, Y1 and Diagonal) (Figure 4). The applicable off-set conditions at each direction for Sc measurements were given by “√” symbol that inserted in Table 1.

The Scp measurements at dose maximum depth (1.6 cm) in RW3 phantom using the same chamber and TLD were carried on fields that mentioned above for Scp. The source-surface distance was set to 100 cm and irradiation of 100 MU per field was performed. The total output measurements for OSFs procedure was the same as Scp total output factor measurement while the field location was placed on centre of OSFs at three directions (X2, Y1 and Diagonal). This measurement was done by both CC04 ion chamber and TLD pairs. The applicable OSF conditions at each direction for Scp measurements were given by “*” symbol that inserted in Table 1.

Figure 3. The in-air output measurement set-up.

Figure 4. The field locations depend on off-set fields at (a) X2; (b) Y1; and (c) Diagonal directions.

(a) (b) (c)

Table 1. The Sc and Scp measurement conditions: off-set value along (a) X2; (b) Y1; and (c) Diagonal direction for each field.

3. Results

This study was conducted in Elekta Synergy linear accelerator that produced 6 MV X-ray energy. Before measurements, the linear accelerator was calibrated to deliver 1 cGy/MU for 10 × 10 cm2 at dDmax and 100 cm SSD. All measurement results were obtained by applying 100 MU per fields. The geometric uncertainty related to field aperture arrangement in LINAC and detector position in water phantom were less than 1 mm totally.

3.1. In-Air and Total Output Factor Measurements for Central Fields

The central field in-air output measurements per 100 MU for 1 × 1, 2 × 2, 3 × 3, 4 × 4, 5 × 5, 6 × 6 and 10 × 10 cm2 fields were performed and normalized to 10 × 10 cm2 reference field for Sc. Generally, by decreasing field size from 10 × 10 to 1 × 1 cm2 the Sc value decreased. While field size changing from 10 to 2 cm2 this reducing rate was 5.4%, and by adding 1 × 1 cm2 field to this range it was dramatically drop to 12.5%. Sc, in-air output, values for central fields were shown in the first row (off-set = 0) of Figure 5.

The results for total output factor by CC04 and TLD were shown in the first column (off-set = 0) of Figure 6(a). By decreasing field size from 10 to 2 cm2 the Scp value was decreased by using both detectors. The decreasing ratio in the results by CC04 and TLD were reached to 14.5% and 11% respectively. When including 1 × 1 cm2 field to this range the discrepancy between CC04 and TLD results was seen as 58.5% and 27.8%, respectively.

The output factors measured by TLD and CC04 were comparable and showed close agreement with each other regarding to field sizes up to 2 × 2 cm2. While for fields smaller than 2 × 2 cm2 these results got deviations (Figure 6(a)).

3.2. In-Air and Total Output Factor Measurements for Off-Set Fields

The in-air output measurement values for different off-set distances and indicated fields were obtained (Table 1). For calculating Sc factor for them, the in-air

Figure 5. The normalized Sc values according field size and its off-set value on each X2, Y1 and diagonal directions.

(a)(b)(c)

Figure 6. The normalized Scp values were obtained from CC04 ion chamber and TLD according field size and its off-set distance on each (a) X2; (b) Y1; and (c) Diagonal directions.

output of 10 × 10 cm2 field that placed on central axis was used. The normalized Sc values according to 10 × 10 cm2 for three directions X2, Y1 and Diagonal are shown in Figure 5. By increasing off-set value on any direction for each field, the Sc value was increased (excluding 1 × 1 cm2 field size). For example, the Sc value for 4 × 4 cm2 field was raised to 5.9%, 5.6% and 3% on X2, Y1 and Diagonal direction, respectively. The maximum increase in Sc was seen on the Y1 Direction when all results were evaluated.

The total output measurement values using CC04 and TLD for off-set fields was obtained and the total output of 10 × 10 cm2 field that placed on central axis was used for calculating Scp factor. The normalized Scp values according to 10x10 cm2 for three directions X2, Y1 and Diagonal are shown in Figures 6(a)-(c) respectively. By increasing off-set value on any direction for each identified fields the Scp value was increased for both detectors (excluding 1 × 1 cm2 field). For example, the Scp values by CC04 for 4 × 4 cm2 field was raised to 7.2%, 11.6% and 10.7% on X2, Y1 and Diagonal direction, respectively. These values were 11.2%, 13.7% and 10.8% respectively from results by using TLDs. Total output factors measured by TLD showed close agreement with those measured using the ion chamber for field sizes of 4 × 4 cm2 and above. It is recognized that TLD’s were more sensitive for small fields especially 1 × 1 cm2 and 2 × 2 cm2 in off-set measurements compared to CC04. The maximum increase in Scp was seen on the Y1 Direction, when all results were evaluated.

4. Discussion

Khan et al. [10] has declared that in the case of static MLC in conventional radiotherapy, Sc for a given jaw opening is affected very little by the MLC setting for fields larger than 4 × 4 cm2. On the other hand, when the MLC aperture is reduced to 1 × 1 cm2 field this factor drops to 5%. The Sc characteristics which obtained from this study were consistent with findings of Zhu et al. [12] , Jaffray et al. [2] and Sharpe et al. [3] . The Sc value for square fields of sides 10−1 cm in many studies demonstrated a decreasing with different ranges related to energy, these were 0.823 for 10 MV [19] and 0.91 for 6 MV X-ray energy [20] . When the field size decreased, the direct-beam source was partially blocked, so the number of direct-beam (primary photons) reaching to the measure-point was reduced and Sc value decreased [21] . However, at field sizes smaller than 2 × 2 cm2 the direct beam source was shielded by the collimating structures and source occlusion became important. So, the sharply decrease of Sc was seen at 1 × 1 cm2.

Das et al. showed that the output factor for small fields at 6 MV X-ray energy was strongly depend on detector type and a rapid drop in output with a certain detector was observed when the field size was decreased especially including 1 × 1 cm2 field [22] [23] . Also, Cranmer-Sargison et al. found the same measurement results using different ion chamber [14] . Because of the volume effect and water-equivalent property of TLD, the results obtained from them were more reliable and higher then CC04 ion chamber results (Figure 6(a)).

Shih et al. proposed a method for the calculation of head scatter or in-air output factor for an arbitrary jaw setting. They found that the head scatter factors at isocenter for asymmetric fields are lower than for the same jaw setting that placed at centre of field by up to 4% [24] . It can be recognized that the off-set of fields that generally used in creating of asymmetric fields by collimator settings also affect the Sc value.

There is a quite little information about evaluations regarding off-set of fields and their Scp value. Only one study contained the evaluation of 4 cm shift on leaf axis for square fields from 0.5 up to 10 cm2. The Scp ratio regarding off-set field to no off-set field did not show any fluctuation from 10 to 4 cm2 fields while by decreasing of field size to 0.5 cm, this ratio dropped to 13% in 6 MV X-ray energy [25] . When the results of this study were compared to our results, the same outcomes were driven for shifts up to 6 cm at each direction for fields larger than 4 cm2. For larger off-set or shift values on this range of fields Scp ratio raised to maximum 10%. On the other hand, for 4 cm2 and 2 cm2 fields the large off-set distance caused increasing slightly higher than 10%. An additional comparison could be made between the results regarding to use of the different detector types. The difference in the active volumes of the detectors used in this study represents the most likely cause of large differing in Scp values especially for fields smaller than 4 cm2.

5. Conclusions

This study was focused on the impact of off-set conditions of in-air and total output factor for 1 - 10 cm2 radiation fields at 6 MV X-ray photon energy by a linear accelerator with collimation device equipped by 1 cm leaf width. By increasing off-set distance, the Sc and Scp values were increased for any related fields regarding shift at all directions comparing to central fields. There are some commencements related to this matter: the particular configuration of a LINAC head and collimation device, the flattening filter and its geometry and etc. The contribution of scattered photons from the primary collimator was larger than that of the flattening filter, and backscattered particles were affected not only by the upper jaw but also the lower jaw. Therefore, the low secondary filter was correctly modelled, because the design of this filter plays a role in the variation of the accelerator output as a function of the off-set fields. In this case, it is considered that the increase of Sc and Scp values is due to the breakdown of the homogeneity at the off-set fields depending on the flattening filter design. So, the photons that arrived to the distal fields are supposed to pass from the edge of the flattening filter intensely and in the average, more energetic. Based on this fact, the Sc and Scp values related to off-set fields should be assessed during TPS quality control processes.

Also, the farther works need to perform for each LINAC head design to obtain additional factor on regard to off-set distance for Sc and Scp of small fields that hugely used in IMRT treatment plans.

There are now an increasing number of innovated detectors for small field dosimetry as miniature ionization chambers, diodes, synthetic diamonds, and plastic scintillators that these will seem to lead a solution for dosimetry of very small field as well dosimetry of their off-set position.

In conclusion, the dosimetric properties of small fields and their off-set should be evaluated and modelled appropriately in the treatment planning system to ensure accurate dose calculation in Intensity Modulated Radiation Treatment.

Acknowledgements

This work was supported by Akdeniz University Research Project Coordination Unit (Project No. 2014.02.0121.008). It was carried on Radiotherapy Centre, Denizli State Hospital, Turkey and the authors would like to thank the members of this centre.

Cite this paper

Ozdemir, H., Tuncel, N. and Kizildag, A.U. (2017) The Impact of Small Field’s Off-Set on Output Factor in Intensity Modulated Radiation Treatment. International Journal of Medical Physics, Clinical Engineering and Radiation Oncology, 6, 433-444. https://doi.org/10.4236/ijmpcero.2017.64039

References

  1. 1. Zhu, T.C. and Bjärngard, B.E. (1995) The Fraction of Photons Undergoing Head Scatter in X-Ray Beams. Physics in Medicine and Biology, 40, 1127-1134. https://doi.org/10.1088/0031-9155/40/6/011

  2. 2. Jaffray, D.A., Battista, J.J., Fenster, A. and Munro, P. (1993) X-Ray Sources of Medical Linear Accelerators: Focal and Extra-Focal Radiation. Medical Physics, 20, 1417-1427. https://doi.org/10.1118/1.597106

  3. 3. Sharpe, M.B., Jaffray, D.A., Battista, J.J. and Munro, P. (1995) Extrafocal Radiation: A Unified Approach to the Prediction of Beam Penumbra and Output Factors for Megavoltage X-Ray Beams. Medical Physics, 22, 2065-2074. https://doi.org/10.1118/1.597648

  4. 4. Sharma, S.D. (2011) Unflattened Photon Beams from The Standard Flattening Filter Free Accelerators for Radiotherapy: Advantages, Limitations and Challenges. Medical Physics, 36, 123-125. https://doi.org/10.4103/0971-6203.83464

  5. 5. Zhu, T.C., Ahnesjö, A., Lam, K.L., Li, X.A., Ma, C.M., Palta, J.R., Sharpe, M.B., Thomadsen, B. and Tailor, R.C. (2009) In-Air Output Ratio, Sc, for Megavoltage Photon Beams. Report of the AAPM Radiation Therapy Committee Task Group No. 74. Medical Physics, 36, 5261-5291. https://doi.org/10.1118/1.3227367

  6. 6. Dutreix, A., Bjärngard, B.E., Bridier, A., Mijnheer, B., Shaw, J.E. and Svensson, H. (1997) Monitor Unit Calculation for High Energy Photon Beams. Physics for Clinical Radiotherapy. ESTRO Booklet No. 3. Garant, Leuve, 36 p.

  7. 7. Young, M.E.J. (1957) Radiological Physics. 1st Edition, Academic, New York, 207 p.

  8. 8. Holt, J.G., Laughlin, J.S. and Moroney, J.P. (1970) The Extension of The Concept of Tissue-Air Ratios TAR to High-Energy X-Ray Beams. Radiology, 96, 437-446. https://doi.org/10.1148/96.2.437

  9. 9. Kim, S., Palta, J.R. and Zhu, T.C. (1998) A Generalized Solution for The Calculation of In-Air Output Factors in Irregular Fields. Medical Physics, 25, 1692-1701. https://doi.org/10.1118/1.598350

  10. 10. Khan, F.M. and Gibbons, J.P. (2014) The Physics of Radiation Therapy. 5th Edition, Lippincott Williams & Wilkins, Philadelphia, 152-153.

  11. 11. Kase, K.R. and Svensson, G.K. (1986) Head Scatter Data for Several Linear Accelerators 4-18 MV. Medical Physics, 13, 530-532. https://doi.org/10.1118/1.595857

  12. 12. Tatcher, M. and Bjärngard, B.E. (1992) Head-Scatter Factors and Effective X-Ray Source Positions in A 25-MV Linear Accelerator. Medical Physics, 19, 685-686. https://doi.org/10.1118/1.596811

  13. 13. Cranmer-Sargison, G., Weston, S., Sidhu, N.P. and Thwaites, D.I. (2011) Experimental Small Field 6 MV Output Ratio Analysis for Various Diode Detector and Accelerator Combinations. Radiotherapy and Oncology, 100, 429-435. https://doi.org/10.1016/j.radonc.2011.09.002

  14. 14. Sharma, S.D. (2014) Challenges of Small Photon Field Dosimetry Are Still Challenging. Journal of Medical Physics, 39, 131-132. https://doi.org/10.4103/0971-6203.138998

  15. 15. Laud, W.U. and Wong, T. (2003) The Volume Effect of Detectors in The Dosimetry of Small Fields Used in IMRT. Medical Physics, 30, 341-347. https://doi.org/10.1118/1.1544678

  16. 16. Olofssona, J., Georgb, D. and Karlssona, M. (2003) A Widely Tested Model for Head Scatter Influence on Photon Beam Output. Radiotherapy and Oncology, 67, 225-238. https://doi.org/10.1016/S0167-8140(02)00409-7

  17. 17. ICRU (2014) REPORT 91: Prescribing, Recording, and Reporting of Stereotactic Treatments with Small Photon Beams. Journal of the International Commission on Radiation Units and Measurements, 14, 31-53.

  18. 18. Houdek, P.V., Vanburen, J.M. and Fayos, J.V. (1983) Dosimetry of Small Radiation Fields for 10 MV X-Rays. Medical Physics, 10, 333-336. https://doi.org/10.1118/1.595262

  19. 19. IBA (2011) CC04, User’s Guide. http://www.iba-dosimetry.com/sites/default/files/brochure/RT-BR-E-Detectors%20for%20RD%20and%20AD%200211_0.pdf

  20. 20. Thomas, S.J., Fipem, Eaton, D.J., Tudor, G.S.J. and Twyman, N.I. (2008) Equivalent Squares for Small Field Dosimetry. The British Journal of Radiology, 81, 897-901. https://doi.org/10.1259/bjr/27713136

  21. 21. Aspradakis, M.M., Byrne, J.P., Palmans, H., Duane, S., Conway, J., Warrington, A.P. and Rosser, K. (2010) Small Field MV Photon Dosimetry. IPEM Report 103, IPEM.

  22. 22. Das, I.J., Ding, G.X. and Ahnesjö, A. (2008) Small Fields: Nonequilibrium Radiation Dosimetry. Medical Physics, 35, 206. https://doi.org/10.1118/1.2815356

  23. 23. Das, I.J., Morales, J. and Francescon, P. (2016) Small Field Dosimetry: What Have We Learnt? AIP Conference Proceedings, 1747, Article ID: 060001.

  24. 24. Shih, R., Li, X.A., Chu, J.C.H. and Hsu, W.L. (1999) Calculation of Headscatter Factors at Isocenter or at Center of Field for Any Arbitrary Jaw Setting. Medical Physics, 26, 506-511. https://doi.org/10.1118/1.598549

  25. 25. Klein, D.M., Tailor, R.C., Archambault, L., Wang, L., Therriault-Proulx, F. and Beddar, A.S. (2010) Measuring Output Factors of Small Fields Formed by Collimator Jaws and Multileaf Collimator using Plastic Scintillation Detectors. Medical Physics, 37, 5541-5549. https://doi.org/10.1118/1.3488981

NOTES

Intensity Modulated Radiation Treatment