Mathematical Models for Flow of Chyme during
Gastrointestinal Endoscopy
Ranadhir Roy, Fabiola Rios, Daniel N. Riahi
Department of Mathematics, University of Texas-Pan American, Edinburg, USA
E-mail: rroy@utpa.edu
Received March 11, 2011; revised March 25, 2011; accepted March 29, 2011
Abstract
Intestinal infection has become a common disease in human and endoscopy can be a powerful means in di-
agnosis of intestinal illnesses. Mathematical models are developed for an inserted endoscope on the flow of
chyme in the small intestine considering a Newtonian incompressible fluid flow, under an axisymmetric
condition, in a cylindrical annulus between the small intestine and the endoscope. We obtain novel mathe-
matical expressions for the pressure drop, forces exerted by the endoscope on the flow of chyme, and the
force exerted by the chyme on the intestine for one wave length of the peristaltic rush wave. We also inves-
tigate and calculate the flow velocity and pressure for different flow rates and the wave lengths. The results
are presented, and discussed for the cases and conditions under which pressure, pressure drop can be positive
or negative and the forces can be acted either by the intestine or endoscope on the flow or vice-versa.
Keywords: Mathematical Models, Gastrointestinal, Chyme Flow, Endoscopy, Flow Modeling, Newtonian
Incompressible Fluid Flow
1. Introduction
Intestinal infection has become a rather common disease
in living systems such as that of human. This illness is
often caused by the environmental pollution like water
pollution that can lead to constipation, distention, etc.
Due to such infectious conditions, a strong wave called
peristaltic rush develops and it travels relatively long
distances in a few minutes in small intestine, which is a
convoluted tube of about 6-7 m in length and average
radius about 1.25 cm lying in the central and lower parts
of abdomen. The functions of the small intestine in hu-
man, which is called gastrointestinal tract, are for diges-
tion and absorption. It is known that endoscopy can be a
powerful means in diagnosis and management of various
types of intestinal illnesses.
Although there have been studies about transport of
chyme flow by the peristaltic rush wave [1-6], but rela-
tively little is known about the adverse effect of an in-
serted endoscope on the chyme flow in the small intes-
tine. Srivastava [7] investigated a particular form of this
problem under the approximation of purely axisymmetric
and Newtonian fluid flow in a cylindrical annulus be-
tween the small intestine and the endoscope and re-
stricted to case of a peristaltic rush wave as a pure single
sine wave of long wavelength along the outer boundary
of the annulus. The author calculated the pressure drop
and the friction forces as functions of some parameters.
In the present study, we reformulate the forces acted
by endoscope and the intestine on the flow of chyme in
the presence or absence of the endoscope and calculate
the pressure, pressure drop, leading order flow velocity
and the forces due to the boundaries. Results are ob-
tained with and without endoscope in the intestine and
verify the results in particular operative regime which
can be appropriate to the case flow of chyme during gas-
trointestinal endoscpy procedure in a human.
The present study is the first step of our plan to inves-
tigate chyme flow during gastrointestinal endoscopy. We
are presently working on related problems but more real-
istic cases of non-axisymmetric flow, a non-Newtonian
power-law fluid and superposed sine waves of different
wavelengths, and the results of such studies will be re-
ported elsewhere. This is consistent with our plan to un-
derstand the way intestinal infected disease can be diag-
nosed more effectively and thus improve the patient’s
health more quickly and with no possibility for compli-
cation due to lack of clear understanding of the factors
that can contribute to the patient’s illness.
R. ROY ET AL.601
2. Model Formulation
We consider the problems of axisymmetric incompressi-
ble Newtonian fluid flow in a cylindrical annulus boun-
ded at the outer boundary by the small intestine and at
the inner boundary by the inserted endoscope. We as-
sume that a sinusoidal peristaltic wave travels down the
wall of the outer tube, which corresponds to the bound-
ary of the small intestine (Figure 1). We describe the
geometry of the outer tube surface as

 
sin,
o
H
z, trA2zct
 
(1)
where ro is the radius of the outer tube (intestine), A is
the amplitude of the wave, c is the wave speed, t is the
time variable and z is the axial variable.
Due to the geometry of the physical problem, we con-
sider cylindrical coordinate system. The governing equa-
tions for the continuity and the momentum of an axi-
symmetric Newtonian fluid flow in a fixed frame are
given by White [8]
 
0rurrw z (2a)


 
22
22
,
1
utuurwuz
pr uur
rrrr z



2
(2b)


2,
wtuwrwwz
pz w

 (2c)
where r is the radial variable, z is the axial variable, t is
the time variable,
is the fluid density,
is the dynamic
viscosity of the fluid, u is the radial velocity, w is the
axial velocity and p is the pressure.
We now consider the governing system in the peristal-
tic wave frame of reference moving at speed c of the
wave, so that (2a)-(2c) become
 
0ru rrWZ 
, (3a)
Figure 1. This is a schematic diagram for the physical sys-
tem.

22
utuurWuZ
pr uur
 
 (3b)

2
WtuWrWWZ
pZ W,
 
 (3c)
where
,
Z
zctW wc
 (3d)
The boundary conditions in the moving frame of ref-
erence are
0,at,0
i
uWcrru
 (4a)

,, ,
at
uHtWrHZtHZ
WcrH
 
  (4b)
where ri is the radius of the inner cylinder (endoscope).
Next, we non-dimensionalize the governing Equations
(3a)-(3c) and the boundary conditions (4a)-(4b) using ro,
,
/(roc), c,
/c, (
c
/ro) as scales for radial distance,
axial distance, radial velocity, axial velocity, time and
pressure, respectively. The non-dimensional form of the
governing equations and the boundary conditions are
then given below for simplicity using the same symbols
that were used before for the independent variables, flow
velocity and pressure
0ru rrWZ
 (5a)
 
3
22
1
e
RutuurWuZ
222
p
rrrrururu


Z
 
(5b)
 
1
e
22 2
eR(W tuW rWW Z)
pZr /rrWrWZ

  
(5c)
0,1 at=uW r
 (5d)

,, ,
1at1 sin2
uWrhZtHZ
Wrh
 
Z

(5e)
where R =
c ro/
is the Reynolds number,
= ro/
is
ratio of the length scale in the radial to that in the axial
direction,
= ri/ro is the annulus aspect ratio, and =
A/ro is the non-dimensional amplitude of the wave.
We shall assume that the Reynolds number Re is suffi-
ciently small and
is also a small parameter (wave
length is large), so our leading order investigation ne-
glects terms in (5b)-(5c) that contain such parameters.
Such assumptions are consistent with the available ob-
servation by Lew et al. [9] that Re is very small and long
wavelength approximation also exists in the small intes-
tine. Under such assumptions, the Equation (5b) implies
that the pressure p is independent of the radial variation
Copyright © 2011 SciRes. AM
R. ROY ET AL.
Copyright © 2011 SciRes. AM
602



1
222
i
0
2ln d
F
GZhh Z
 

 
(10a)
to the leading terms, and the Equation (5c) can be inte-
grated twice with respect to r leading to the result



22 22
10.25d dlnln
W
pZrh rh
 

 





1
0
ln d
222
o
F
GZ hh2hZ

 

(10b)
It should be noted that the main results in the absence
of the endoscope can be found by taking the limit
0
in (9)-(10). This leads to
(6)
after the necessary boundary conditions given in (5d)-(5e)
are used.
Next, the dimensionless volume flow rate q = S WdS,
which is assumed to be given, in the wave frame along
the axial direction is determined. Here S is the cross-
sectional area of the annulus. We then find




10 125ddln
2222 22
q
h.pZhh h

 
 


1
22
00
0
1
2
0
0
d,81 2,
0, d
io
pGZZGZQ hh
FF hGZZ




4
(11a-d)
(7)
which are similar to the results due to Srivastava and
Saxena [11] for the single tube case after the relation (8c)
is used to write Q in terms of q.
Now, the relation between the q in the wave frame and
qL in the laboratory frame, which is the instantaneous one,
is
3. Results
2d
h
22
LL
qqrrq h

(8a)
The relation between the mean volume flow rates
given by
1
0
d
L
QqZ (8b)
in the laboratory frame and q can then be found by
taking the average of (8a), which yields the following
equation:
2
12qQ
2
 (8c)
The result (8c) is analog to that in the absence of en-
doscope [3]. Using (7), we can solve for dp/dZ and then
integrate over one wavelength in the axial direction and
make use of (8c) to find the following expression for the
pressure drop p across one wavelength:
 
 

1
0
2
24422
d
812 ln
p=GZ Z,GZ
Qhh hh
 


 

The main numerical results for the pressure drop and the
frictional forces given by the analytical expressions
(9)-(10) are determined using a computer code developed
for the numerical integrations of these equations. The
data were generated for the pressure drop () and the
frictional forces (i
p
F
ando
F
) for given values of the
mean volume flow rate (Q), amplitude of the wave (
)
and the aspect ratio of the annulus (
). Here a positive
pressure drop (p > 0) means that pressure in the chyme
flow decreases with increasing Z, while a negative pres-
sure drop (p < 0) means that pressure in the chyme
flow increases with Z. Due to the nature of the annulus
flow and the way the expressions for Fi and Fo are de-
termined in the previous section, it should be noted that a
positive value of the inner friction force (Fi > 0) means
this force is acted by the endoscope on the chyme flow,
while Fi <0 means there such force acted by the chyme
flow on the endoscope. However, in the case of the outer
friction force, a negative value of Fo (Fo < 0) means that
such force is acted by the intestine on the chyme flow,
while a positive value of such force means that the
chyme flow applies such force on the intestine.
Figures 2-4 present some typical results for the pres-
sure drop, the inner friction force and the outer friction
force versus the flow rate for different values of the am-
plitude of the wave and the aspect ratio of the annulus. It
can be seen from the Figure 2 that the pressure drop is
generally positive, provided the flow rate is not suffi-
ciently small. Otherwise, the pressure drop is negative. In
addition, for a given moderate value of Q, the pressure
drop is positive and larger for larger values of aspect
ratio and the wave amplitude. Thus, there is a pressure
gradient force along the positive Z-direction, which acts
(9a-b)
Frictional forces Fi and Fo at the inner and outer
boundaries, respectively, can be found from the integra-
tion of the shear stress around each tube and over one
wavelength in the axial direction [10]. In analogy to the
flow over a planner surface, we consider integral over
one wavelength along the Z-direction of the dimen-
sionless dominated term W/y for the shear stress over
the plate, where y+r = ro. This leads to the expressions
for these forces that are given below
R. ROY ET AL.603
Figure 2.Pressure drop p versus the flow rate Q with the annulus aspect ratio
= 0, 0.44 and the wave amplitude = 0.2, 0.4.
Figure 3. The inner friction force Fi versus flow rate Q with annulus aspect ratio
= 0.32, 0.44 and wave amplitude = 0.2,
0.4.
Figure 4. The outer friction force Fo versus flow rate Q with annulus aspect ratio
= 0.32, 0.44 and wave amplitude = 0.2,
0.4.
Copyright © 2011 SciRes. AM
R. ROY ET AL.
604
on the flow, if the flow rate of the chyme flow is not too
small, while such force takes opposite direction along
negative Z-direction if the flow rate is too small. The
magnitude of the pressure drop appears to increase sig-
nificantly as either wave amplitude or the endoscope’s
cross sectional size increases. In analogy to that of flow
over a surface, it is expected that negative pressure drop
with sufficiently large magnitude can destabilize the
chyme flow located between the intestine and the endo-
scope.
It can be seen from the Figure 3 that the inner friction
force, i
F
, is generally negative (i.e. the force acted by
the chyme flow is on the endoscope), provided the flow
rate is not sufficiently small. Otherwise, Fi is positive. In
the regimes where Fi is either positive or negative, the
magnitude of such force increases with either the wave
amplitude or the endoscope’s cross-sectional size. Thus,
if the flow rate in the chyme flow is sufficiently small,
then the force exerted by the endoscope on the chyme
flow can have stronger effect if either the wave ampli-
tude or the endoscope’s cross-section are larger, while
for sufficiently large values of the flow rate the force by
the endoscope is replaced by that due to the chyme flow,
which applies on the endoscope.
The results presented in the Figure 4 indicate that the
outer friction force, o
F
, is generally positive (i.e., the
force is acted by the chyme flow on the intestine) pro-
vided Q is not too small. Otherwise, o
F
becomes nega-
tive. For sufficiently large Q, o
F
is larger with larger
value of either amplitude or the aspect ratio. For suffi-
ciently small Q, where o
F
is negative, the magnitude of
such forces increases with increasing either the ampli-
tude or the aspect ratio. Thus, for sufficiently large flow
rate of the chyme flow, the flow exerts a force on the
intestine, and this force increases in its intensity for lar-
ger values of either the wave amplitude or the endo-
scope’s cross-sectional size. However, if the flow rate is
sufficiently low, then there is a force acted by the intes-
tine on the flow whose magnitude can increase with in-
creasing either the wave amplitude or the endoscope’s
cress-sec- tion.
Figures 5-7 present some typical results for the pres-
sure drop p
, the inner friction force Fi and the outer
friction force Fo versus the wave amplitude for different
values of the flow rate Q and the aspect ratio of the an-
nulus
. It can be seen from the Figure 5 that for suffi-
ciently small amplitude, the pressure drop is positive but
is smaller for smaller aspect ratio and higher Q. However,
for sufficiently larger amplitude, the pressure drop can be
negative. Thus, the pressure gradient force in the chyme
flow is in the positive direction of the Z-axis if the wave
amplitude is sufficiently small, while such force has a
direction in the negative Z-axis if the wave amplitude is
sufficiently large, which can destabilize the flow system.
Also such destabilizing effect can be intensified if the
endoscope’s cross-section is sufficiently large.
The results presented in the Figure 6 for the inner
friction force indicate that for sufficiently small ampli-
tude, the inner friction force is negative, and the magni-
tude of this force is larger for smaller Q and larger aspect
ratio. For sufficiently large amplitude, the inner friction
force is positive and has larger magnitude for larger val-
ues of either Q or
. Thus, the force acted by the flow on
the endoscope intensifies if the flow rate is smaller and
the endoscope’s cross-section is larger. The force acted
by the endoscope on the chyme flow is stronger if either
volume flow or the endoscope’s cross section is larger.
From the results presented in the Figure 7, we find that
for sufficiently small amplitude, the outer friction force is
positive which can have larger magnitude for either
smaller flow rate Q or larger annulus aspect ratio. For
sufficiently large amplitude, the outer friction force is
Figure 5. The pressure drop p versus wave amplitude for flow rate Q = 2, 6 and annulus aspect ratio
= 0.32, 0.34.
Copyright © 2011 SciRes. AM
R. ROY ET AL.605
Figure 6. The inner friction force Fi versus wave amplitude
for flow rate Q = 2, 6 and annulus aspect ratio
= 0.32, 0.34.
Figure 7. The outer friction force Fi versus wave amplitude
for flow rate Q = 2, 6 and annu lus aspe ct ratio
= 0.32, 0.34.
negative, and its magnitude is again larger for smaller Q
or larger
. Thus, the force exerted by the flow on the
intestine is intensified either for larger endoscope’s
cross-section or for smaller flow rate. The intensification
of the force acted on the flow by the intestine is similar
to the case for force acted by the intestine of the chyme
flow.
We have investigated leading order velocity flow and
pressure versus axial variable for both annulus and tube
flows for different flow rate, the annulus aspect ratio and
the wavelength parameters. Figures 8-10 present some
typical results for the pressure and velocity flow versus
the axial variable. The parameters used for this analysis
are the wave amplitude Φ = 0.2, annulus aspect ratio δ =
0.0, 0.44 and flow rate Q = 0.6, 2.0. Figure 8 presents
results for pressure (p). It can be seen from this figure that
for the flow in the tube case, the pressure is lower for
higher flow rate, which physically is reasonable, and p
mostly decreases with increasing Z especially at higher
flow rate. For the annulus flow, again p is lower for high-
er flow rate and decreases with increasing Z for higher
flow rate. Figure 9 presents some typical results for the
leading order velocity w in the fixed frame versus Z. It
can be seen from the Figure 9 that as anticipated the flow
velocity is higher for higher flow rate. For the flow in the
tube case, the flow direction is from left to the right, ex-
cept in the case of low flow rate where the flow is to the
left with relatively small magnitude in an intermediate
zone in the axial direction. It is anticipated that in this
zone flow circulate slightly. Similar results are apparent
in the case of the flow in the annulus. Figure 10 presents
some characteristic results for axial velocity w versus
aspect ratio r for r between annulus aspect ratio δ and
outer tube surface h and at a axial distance Z = 0.5. The
parameter values are the same as those given for the Fig-
ures 8 and 9, respectively. It can be seen from the Figure
10, that w>0 and the value of the leading order velocity
flow is higher for higher flow rate Q. The leading order
velocity flow in the tube case is higher than that for w in
the annulus flow only if r is sufficiently small.
Copyright © 2011 SciRes. AM
R. ROY ET AL.
606
Figure 8. Pressure p versus axial variable Z for wave amplitude0.2
, annulus aspect ratio
= 0.0, 0.44 and flow rate Q =
0.6, 2.0.
Figure 9. Axial velocity w versus axial variable Z at aspect ratio r = 0.5 for wave amplitude 0.2
, annulus aspect ratio
=
0.0, 0.44 and flow rate Q = 0.6, 2.0.
Figure 10. Axial velocity w versus aspect ratio r at Z = 0.5 for wave amplitude 0.2
, annulus aspect ratio
= 0.0, 0.44
and flow rate Q = 0.6, 2.0.
Copyright © 2011 SciRes. AM
R. ROY ET AL.
Copyright © 2011 SciRes. AM
607
4. Conclusions & Remarks
We have carried out mathematical modeling and compu-
tation of a Newtonian incompressible fluid flow, under
an axisymmetric condition, in a cylindrical annulus be-
tween the small intestine and an endoscope. We have
developed the expressions for the pressure, pressure drop,
leading order flow velocity in the flow and the integrated
forces at the intestine and the endoscope over one wave-
length of the peristaltic wave. We have computed the
results for the pressure, flow velocity, pressure drop and
forces for different flow rates, the amplitude of the wave,
and endoscope cross-sectional size. We have seen that
pressure decreases with increasing axial variables indi-
cating that driving the pressure gradient force along the
peristaltic wave and the pressure drop is generally posi-
tive. The magnitudes of the frictional forces are in-
creased with increasing the wave amplitude or increasing
the annulus aspect ratio or both. The leading order veloc-
ity flow is higher for higher flow rate, has higher ampli-
tude in the absence of endoscope, decreases with in-
creasing the radial variable r in the absence of endoscope,
while increases with r in the presence of endoscope.
The purpose of the present investigation is to do mod-
eling and computational study of the effect of an inserted
endoscope on the flow of chyme in the small intestine
using the simplest case by the consideration of axisym-
metric flow of Newtonian fluid. Since boundary of the
small intestine is not purely axisymmetric, a more realis-
tic consideration for such flow system will be to consider
non-axisymmetric flow system of either Newtonian fluid
or more realistic power-law fluid [5,8]. Such study will
be done in near future. Also as part of such future study,
we will consider the system that models more realisti-
cally the flow of chyme and the peristaltic rush waves by
assuming that such waves are composed of several sine
waves with different wavelengths of moderately large
size.
5. Acknowledgements
This research was supported by a grant from UTPA-URI.
6. References
[1] M. Y. Jaffrin and A. H. Shapiro, “Peristaltic Pumping,”
Annual Review of Fluid Mechanics, Vol. 3, 1971, pp.
13-36. doi:10.1146/annurev.fl.03.010171.000305
[2] T. W. Latham, “Fluid Motion in a Peristaltic Pump,”
Master’s Thesis, Massachusetts Institute of Technology,
Cambridge, 1966.
[3] A. H. Shapiro, M. Y. Jaffrin and S. L. Weinberg, “Peri-
staltic Pumping with Long Wavelength at Low Reynolds
Number,” Journal of Fluid Mechanics, Vol. 37, 1969, pp.
799-825. doi:10.1017/S0022112069000899
[4] L. M. Srivastava and V. P. Srivastava, “Peristaltic Tran-
sport of Blood. Casson Model II,” Journal Biomechanics,
Vol. 17, No. 11, 1984, pp. 821-829.
doi:10.1016/0021-9290(84)90140-4
[5] L. M. Srivastava and V. P. Srivastava, “Peristaltic Tran-
sport of a non-Newtonian Fluid: Applications to Vas De-
ferens and Small Intestine,” Annals Biomedical Engi-
neering, Vol. 13, No. 2, 1985, pp. 137-153.
doi:10.1007/BF02584235
[6] V. P. Srivastava, “Particle-Fluid Suspension Flow In-
duced by Peristaltic Waves in a Circular Cylindrical Tube,”
Bulletin of the Calcutta Mathematical Society, Vol. 5, No.
1, 2002, pp. 167-184.
[7] V. P. Srivastava, “Effects of an Inserted Endoscope on
Chyme Movement in Small Intestine-A Theoretical Mo-
del,” Applications and Applied Mathematics, Vol. 2, No.
2, 2007, pp. 79-91.
[8] F. M. White, “Viscous Fluid Flow,” McGraw-Hill, Inc.,
New York, 1991.
[9] H. S. Lew, Y. C. Fung and C. B. Lowenstein, “Peristaltic
Carrying and Mixing of Chyme in Small Intestine,” Jour-
nal Biomechanics, Vol. 4, No. 4, 1971, pp. 297-315.
doi:10.1016/0021-9290(71)90036-4
[10] L. D. Landau and E. M. Lifshitz, “Fluid Mechanics,” 2nd
Edition, Pergamum, London, 1987.
[11] V. P. Srivastava and M. Saxena, “A Two-Fluid Model of
non-Newtonian Blood Flow Induced by Peristaltic Waves,”
Rheologica Acta, Vol. 34, No. 4, 1995, pp. 406-414.
doi:10.1007/BF00367155