Paper Menu >>
Journal Menu >>
![]() J. Biomedical Science and Engineering, 2011, 4, 297-314 JBiSE doi:10.4236/jbise.2011.44040 Published Online April 2011 (http://www.SciRP.org/journal/jbise/). Published Online April 2011 in SciRes. http://www.scirp.org/journal/JBiSE Identification and low-complexity regime-switching insulin control of type I diabetic patients Ali Hariri1, Le Yi Wang2 1DTE Energy, One Energy Plaza, Detroit, Michigan, USA; 2ECE Department, Wayne State University, Detroit, USA. Email: hariria@dteenergy.com, U.S.lywang@wayne.edu Received 3 June 2010; revised 19 June 2010; accepted 21 June 2010. ABSTRACT This paper studies benefits of using simplified re- gime-switching adaptive control strategies in im- proving performance of insulin control for Type I diabetic patients. Typical dynamic models of glu- cose levels in diabetic patients are nonlinear. Using a linear time invariant controller based on an oper- ating condition is a common method to simplify control design. On the other hand, adaptive control can potentially improve system performance, but it increases control complexity and may create fur- ther stability issues. This paper investigates patient models and presents a simplified switching control scheme using PID controllers. By comparing dif- ferent switching schemes, it shows that switched PID controllers can improve performance, but fre- quent switching of controllers is unnecessary. These findings lead to a control strategy that utilizes only a small number of PID controllers in this scheduled adaptation strategy. Keywords: Insulin Control; Diabetes; Switching Control; Modeling; Adaptation 1. INTRODUCTION Insulin is a hormone that is necessary for converting the blood sugar, or glucose, into usable energy. The human body maintains an appropriate level of insulin. Diabetes are caused by lack of insulin in the body. There are two major types of diabetes, called type ‘I’ and type ‘II’ dia- betes. Type ‘I’ diabetes are called Insulin Dependent Diabetes Mellitus (IDDM), or juvenile onset diabetes mellitus. Type ‘II’ diabetes are known as Non-Insulin Dependent Diabetes Mellitus (NIDDM) or Adult-Onset Diabetes (AOD) [1-7]. This paper is focused on type ‘I’ diabetes. Type ‘I’ diabetes develop when the pancreas stop producing insulin. Consequently, insulin must be provided through injection or continuous infusion to control glucose levels. The insulin infusion rate to a diabetic patient can be administrated based on the glucose (sugar) level inside the body. Over the years many mathematical models have been developed to describe the dynamic behavior of human glucose/insulin systems. The most commonly used model is the minimal model introduced by Berg- man, et al. [6,8-13]. The minimal model consists of a set of three differential equations with unknown parameters. Since diabetic patients differ dramatically due to varia- tions of their physiology and pathology characteristics, the parameters of the minimal model are significantly different among patients. Based on such models, a vari- ety of control technologies have been applied to glucose/ insulin control problems [14-17]. This paper studies benefits of using simplified adapta- tion control strategies in improving performance of insu- lin control for Type I diabetic patients. Typical dynamic models of glucose levels in diabetic patients are nonlin- ear. Using a linear time-invariant controller based on an operating condition is a common method to simplify control design. On the other hand, adaptive control can potentially improve system performance, but it increases control complexity and may create further stability is- sues. This paper investigates patient model identification and presents a simplified switching control scheme using PID controllers. By comparing different switching sche- mes, it shows that switched PID controllers can improve performance, but frequent switching is unnecessary. These findings lead to a control strategy that utilizes only a small number of PID controllers in this scheduled adap- tation strategy. Many methods and techniques have been investigated, tested, and studied for controlling the glucose level in type ‘I’ diabetes patients. Lynch and Bequette [14] tested the glucose minimal model of Bergman [10] to design a Model Predictive Control (MPC) to control the glucose level in diabetes patients. Also in that study the nonlinear ![]() A. Hariri et al. / J. Biomedical Science and Engineering 4 (2011) 297-314 Copyright © 2011 SciRes. JBiSE 298 mathematical model was linearized about the steady- state values of time variant variables. Fisher [15] used the glucose insulin minimal model to design a semi closed-loop insulin infusion algorithm based on three hourly plasma glucose samplings. The study concen- trated on the glucose level, and did not take in consid- eration of some important factors such as the free plasma insulin concentration and the rate at which insulin is produced as the level of glucose rises. Furler [16,18] modified the glucose insulin minimal model by remov- ing the insulin secretion and adding insulin antibodies to the model. The algorithm calculates the insulin infusion rate as a function of the measured plasma glucose con- centration. The linear interpolation was used to find the insulin rate. The algorithm neglected some important variations in insulin concentration and other model variables. Ibbini, Masadeh and Amer [17] tested the glu- cose minimal model to design a semi closed-loop opti- mal control system to control the glucose level in diabe- tes patients. The rest of this paper is organized as follows. Section 2 discusses basic model structures, experimental data, and model simulations. Section 3 is focused on model parameter identification. The Levenberg-Marquardt al- gorithm is employed to obtain model parameters itera- tively. Control design and switching strategies are pre- sented in Section 4. By comparing fixed PID controllers with switching strategies, we first show that adaptation can be beneficial. Further studies show that using a small set of PID controllers in switching control is a feasible and desirable approach, which simplifies switching sche- me without performance degeneration. 2. MODELS Since the level of the glucose inside the human being body changes significantly in response to food intake and other physiological and environment conditions, for control design it is necessary to derive mathematics models to capture such dynamics [8-10,19-21]. Many mathematical models have been developed to describe the human glucose/insulin system. Such models are highly nonlinear and usually very complex. The most com- monly used and simplified model is the Minimal Model introduced by Bergman, et al. [8-10], which is still non- linear. To further simplify the model for control design, a common practice is to locally linearize the Minimal Model under a given operating condition. 2.1. Model Structures The minimal model of the glucose and insulin is perhaps the simplest model that is physiologically based and represents well for the observed glucose/insulin dynam- ics of a diabetic patient. The insulin enters or exits the interstitial insulin compartment at a rate that is propor- tional to the difference I tIb of plasma insulin I t and the basal insulin level Ib [22,23]. If the level of insulin in the plasma is below the insulin basal level, insulin exits the interstitial insulin compartment; and if the level of insulin in the plasma is above the insulin basal level, insulin enters the interstitial insulin com- partment. Insulin also can flee the interstitial insulin compartment through another route at a rate that is pro- portional to the insulin amount inside the interstitial in- sulin compartment. On the other hand, glucose enters or exits the plasma compartment at a rate that is propor- tional to the difference b Gt G of the plasma glu- cose level Gt and the basal glucose level b G. If the level of glucose in the plasma is below the glucose basal level, the glucose exits the plasma compartment; and if the level of glucose in the plasma is above the glucose basal level, glucose enters the glucose compartment. Glucose also can flee the plasma compartment through another route at a rate that is proportional to the glucose amount inside the interstitial insulin compartment. Currently, the most widely used model in physiologi- cal research on the metabolism of glucose is the Minimal Model. This model structure describes experimental data well with the smallest set of identifiable and meaningful parameters. The Minimal Model consists of two parts: the minimal model of glucose disappearance and the minimal model of insulin kinetics. 11 d db Gt Pxt GtPG t (1) 23 d db xt Px tPItI t (2) d + d It nItG tht t (3) where Gt (mg/dL) is the blood glucose level in plasma; I t (µU/mL) is the insulin concentration level in plasma; x t (min−1) is the variable which is proportional to insulin in the remote compartment, Gb (mg/dL) is the basal blood glucose level in plasma; Ib (µU/mL) is the basal insulin level in plasma; t (min) is the time interval from the glucose injection. Eqs.1 and 2 represent the glucose disappearance and Eq.3 represents the insulin kinetics. The initial conditions of the above three differential equations are as the following: 00 0,00,0GGxI I The model parameters carry some physiological meanings that can be summarized as follows. P1 (min−1) describes the “glucose effectiveness” which represents the ability of blood glucose to enhance its own disposal at the basal insulin level. P2 (min−1) describes the ![]() A. Hariri et al. / J. Biomedical Science and Engineering 4 (2011) 297-314 Copyright © 2011 SciRes. JBiSE 299 decreasing level of insulin action with time. P3 (2 min 1 UmL ) describes the rate in which insulin action is increased as the level on insulin deviates from the corre- sponding baseline. ((µU/mL)(mg/dL)−1min−1) de- notes the rate at which insulin is produced as the level of glucose rises above a “target glycemia” level. n (min−1) represents fractional insulin clearance. h (mg/dL) is the pancreatic “target glycemia” level. G0 (mg/dL) is the theoretical glucose concentration in plasma extrapolated to the time of glucose injection t = 0 [8-10,24]. I0 (µU/mL) is the theoretical plasma insulin concentration at t = 0. µU/mL is the conventional unit to measure the insulin level and has the following conversion formula: 1 micro-unit/ milliliter = 6 picomole/liter (1µU/mL = 6 pmol/L) [25,26]. P1, P2, P3, n, , h, G0 and I0 are the pa- rameters to be estimated. A fourth differential equation will be added to the set of the Minimal Model equations to represent a first-order pump dynamics: 1 1 d1 d Ut Ut ut ta (4) where 1 Ut is the infusion rate, ut is the input command, and a is the time constant of the pump. 2.2. Experimental Data A new approach was developed by Bergman, et al. [8-10] to compute the pancreatic responsiveness and insulin sensitivity in the intact organism. This approach uses computer modeling to investigate the plasma glucose and insulin dynamics during a Frequently Sampled In- travenous Glucose Tolerance (FSIGT). An amount of glucose was injected at t = 0 over a period of time equal to 60 seconds [8-10,27]. The blood samples were taken from a fasting subject at regular intervals of time, and then analyzed for glucose and insulin content. Glucose was measured in triplicate by the glucose oxidize tech- nique on an automated analyzer. The coefficient of vari- ation of a single glucose determination was about ± 1.5%. Insulin was measured in duplicate by radioimmu- noassay, with dextrin-charcoal separation using a human insulin standard. Table 1 shows the FSIGT test data for a normal individual. 2.3. Model Simulation Implementation of the minimal model can be achieved by using computer simulation tools. The two differential Eqs.1 and 2 of the minimal model that correspond to the glucose kinetics are modeled here by using the MAT- LAB/Simulink software. In this model, the insulin I t is considered as an input and the glucose Gt as an output. The values of the input I t at a time interval are given in Table 1. The simulation diagram of the minimal model for the glucose kinetic is shown in Fig- ure 1. The output of the system, glucose Gt , is Table 1. FSIGT test data for a normal individual. Sampling time (minutes) Glucose level (mg/dL) Insulin level (µU/mL) 0 2 4 6 8 10 12 14 16 19 22 27 32 42 52 62 72 82 92 102 122 142 162 182 92 350 287 251 240 216 211 205 196 192 172 163 142 124 105 92 84 77 82 81 82 82 85 90 11 26 130 85 51 49 45 41 35 30 30 27 30 22 15 15 11 10 8 11 7 8 8 7 shown in Figure 2 for a normal individual with the fol- lowing parameters: P1 = 3.082 × 10–2, P2= 2.093 × 10–2, P3 = 1.062 × 10–5, G0 = 350, Gb = 92, Ib = 11, [8-10]. 3. PARAMTERS ESTIMATION 3.1. Parameter Estimation Parameter estimation is to determine the values of model parameters that provide the best fit to measured data, based on some error criteria such as least-squares. The method of least squares assumes that the best-fit curve of a given set of data is the curve that has the minimal sum of the deviations squared from a given set of data [28-31]. Given a set of data (x1, y1), (x2, y2), (x3, y3), , (xN, yN), where the independent variable is x and the dependent variable is y, under a se- lected model function form f x the least squares (LS) estimation seeks to minimize 2 1 N ii i yfx (5) When the function is an m-th degree polynomial 2 01 2 m m fxaax axax (6) we have 2 1 2 2 01 2 1 N ii i Nm iiimi i yfx yaaxax ax (7) The unknown coefficients 012 ,, , , m aaa a , can be estimated to yield a least squares error. ![]() A. Hariri et al. / J. Biomedical Science and Engineering 4 (2011) 297-314 Copyright © 2011 SciRes. JBiSE 300 Figure 1. Simulation diagram of the glucose kinetic model. Figure 2. The output of the minimal model for the glucose kinetics Model parameters can be obtained iteratively to re- duce computational complexity. It starts with an initial guess of the unknown parameters. Each iteration updates the current estimate based on new observations. Suppose there are m base functions12 , ,,m ff f of n parameters 12 , ,,n pp p . The functions and the parameters can be represented as follows: T 12 T 12 , ,, , ,, m n f ff pp p f p (8) The least square method is to find the values of the unknown parameters 12 , ,,n pp p for which the cost function is minimum, i.e. T2 1 11 22 m i i Sf Pff p (9) The Levenberg-Marquardt algorithm is an iterative technique that seeks the minimum of a multivariate function that is expressed as the sum of squares of nonlinear real-valued functions [28]. It has become a standard technique for nonlinear least-squares problems. Levenberg-Marquardt can be thought of as a combina- tion of steepest descent and the Gauss-Newton method. When the current solution is far from the correct one, the algorithm behaves like a steepest descent method which is guaranteed to converge. When the current solution is close to the correct solution, it becomes a Gauss-Newton method. The Levenberg-Marquardt algorithm is an iterative procedure. Let ˆfxp be the parametrized model function. The minimization starts after an initial guess for the parameter vector p is provided. The algorithm is locally convergent, namely it converges when the initial guess is close to the true values. In each iteration step, the parameter vector p is updated by a new esti- mate p, where is a small correction term that can be determined by a Taylor Series expansion which leads to the following approximation: p p ff ppJ (10) ![]() A. Hariri et al. / J. Biomedical Science and Engineering 4 (2011) 297-314 Copyright © 2011 SciRes. JBiSE 301 where, J is the Jacobian of f at p f p Jp (11) Levenberg-Marquardt iterative initiates at the starting point 0 p, and produces a series of vectors 123 , ,pp p etc, that converge towards a local minimizer p of f. At each step, it is required to find the which mini- mizes the value of p p ff xp xpJ That gives the following: ˆ p pp fe xp xxJJ (12) where p is the solution to a linear least squares prob- lem. The minimum is achieved when the term pe J is orthogonal to column space J. Based on that, the fol- lowing can be concluded T0 pe JJ (13) Eq.13 can be rearranged as the following TT pee J JJ (14) The Levenberg-Marquardt algorithm solves a slight variation of Eq.14, which is known as the augmented normal equation T pe NJ (15) where the diagonal elements of N are computed as T ii ii NJJfor 0 , while the rests of the matrix N are identical to those of the matrix T J J. is called the damping parameter. If the updated parame- ter vector, p p, where p is computed from Eq.15, yields a reduction in the residual value or error e, then the update is valid and the process repeats with a de- creased damping parameter . Otherwise, the damping parameter is increased and the augmented normal Eq.15 is solved again. Then the process iterates until a value of p that reduces error is found. The MATLAB Software has the Optimization Toolbox which has a command called Lsqnonlin for this algorithm. This algorithm is applied to our problem here. The FSIGT data sample in Table 1 consists of 24 samples. The FSIGT samples were taken over a period of 182 minutes. The unknown parameters of the minimal model Eq.1, Eq.2, Eq.3 were estimated by utilizing the Le- venberg-Marquadrt Algorithm. The parameters to be es- timated were given an initial guess, then the algorithm was used to update the parameters using the sequential data in Table 1. A MATLAB program was written to estimate the unknown parameters. The estimated values of those parameters are shown in Table 2. Table 2. Estimated Parametres. Parameters Normal Individual #1 Normal Individual #2 P1 P2 P3 n γ h G0 I0 0.032299 0.0092644 5.3004e–006 0.29858 0.0068676 90.3709 295.6801 401.7177 0.049519 41.5953 1.8577e–004 0.14653 1.0113e–005 196.0531 318.84 203.2434 The values of the parameters shown in Ta b l e 2 were implemented in the simulation diagram of the minimal model shown on Figure 1. The values of the glucose levels of both individuals are shown in Table 3. The graphs of both experimental and simulated data for normal individual #1 and #2 are shown in Figure 3 and Figure 4 respectively. These plots show that the two graphs (experimental and simulated) are close to each other, leading to the conclusion that the estimated values of parameters are close to the actual values. In general, the relative errors indicate how good an estimate is, relative to the true values. Although absolute errors are useful, they do no necessarily give an indication of the importance of an error. If the experimental value is denoted by G, and the estimated (or simulated) value is denoted by G, then the relative error is defined as Relative Error ˆ GG G (16) Table 3. Simulated data for normal individuals #1 and #2. Normal Individual #1 Normal Individual #2 Sampling time during the test (minutes) Glucose Level (mg/dL) Glucose Level (mg/dL) 0 2 4 6 8 10 12 14 16 19 22 27 32 42 52 62 72 82 92 102 122 142 162 182 94 295.6801 282.1308 268.2993 254.8580 242.1020 230.1353 218.9702 208.5776 198.9116 185.6659 173.7810 156.6159 142.2917 120.5167 105.8327 96.35277 90.67314 87.73594 86.65621 86.77255 89.03143 92.55181 95.65837 94 318.84 297.2223 277.7749 260.2561 244.4545 230.1878 217.2973 205.6436 195.1034 181.1443 169.1246 152.6775 139.8434 122.0036 111.1272 104.4947 100.4500 97.98329 96.47894 95.56149 94.66075 94.32573 94.20113 ![]() A. Hariri et al. / J. Biomedical Science and Engineering 4 (2011) 297-314 Copyright © 2011 SciRes. JBiSE 302 020 4060 80100 120 140160 18020 0 0 50 100 150 200 250 300 350 400 time (min) Glucose mg/ dL) p Basal Level E x perim ental Data Simulated Data Figure 3. Plot of glucose level G(t) for normal individual #1. 020 40 6080100120 140160180 20 0 0 50 100 150 200 250 300 350 400 time (min) Glucose mg/ dL) p B asal Level E xperim ental Data Simulated Data Figure 4. Plot of glucose level G(t) for normal individual #2. And the Square Relative Error can be expressed as Square Relative Error 2 ˆ GG G (17) When the data is sampled over a certain period of time, the mean squared relative error (MSRE) can be used. The MSRE is defined as 2 1 ˆ 1 MSRE, for 1, 2,, n ii ii GG in nG (18) where i G is the experimental value at sample i. ˆ i G is the estimated value at sample i. n is the number of sam- ples of a data set. The Square Relative Error between the experimental data and the simulated data of the glucose level for indi- vidual #1 and # 2 are calculated and shown in Ta ble 4 and Table 5 respectively. Normally the Mean Square Relative Error is expressed in percentage format. Below is the percentage error for both individuals: the percent- Table 4. Error data for normal individuals #1. Experimental Data, G(t) Simulated Data, ˆ()Gt Square Relative Error 94 298 284 272 253 248 235 217 208 205 191 172 164 141 132 120 116 108 106 104 105 109 107 110 94 295.6801 282.1308 268.2993 254.8580 242.1020 230.1353 218.9702 208.5776 198.9116 185.6659 173.7810 156.6159 142.2917 120.5167 105.8327 96.35277 90.67314 87.73594 86.65621 86.77255 89.03143 92.55181 95.65837 0 6.060466e–005 4.3317e–005 0.0001851148 5.393524e–005 0.0005656016 0.0004285321 8.243416e–005 7.710311e–006 0.0008820624 0.0007799362 0.0001072176 0.002027272 8.391868e–005 0.007568141 0.01393832 0.0286871 0.02573902 0.02968814 0.02781129 0.03013514 0.03356148 0.01823304 0.01699855 Table 5. Error data for normal individuals #2. Experimental Data, G(t) Simulated Data, ˆ() Gt Square Relative Error 94 320 303 289 272 258 244 223 205 194 182 169 152 139 122 112 105 100 98 97 97 95 94 93 94 318.84 297.2223 277.7749 260.2561 244.4545 230.1878 217.2973 205.6436 195.1034 181.1443 169.1246 152.6775 139.8434 122.0036 111.1272 104.4947 100.4500 97.98329 96.47894 95.56149 94.66075 94.32573 94.20113 0 1.314063e–005 0.0003635986 0.001508629 0.001864179 0.002756472 0.003204405 0.0006539557 9.857924e–006 3.235126e–005 2.210359e–005 5.439401e–007 1.986409e–005 3.681781e–005 8.715616e–010 6.073247e–005 2.315634e–005 2.024906e–005 2.909034e–008 2.88559e–005 0.0002199282 1.275242e–005 1.200794e–005 0.0001668063 age MSRE for individual # 1 = 0.99028%; the percent- age MSRE for individual # 2 = 0.04596%. 3.2. Model Linearization Now let us recall the four differential Eqs.1-4 that define the proposed mathematical model and denote them as 1 f G, 2 f x, 3 f I, and 41 f U. The result is ![]() A. Hariri et al. / J. Biomedical Science and Engineering 4 (2011) 297-314 Copyright © 2011 SciRes. JBiSE 303 111 d db Gt f GPxtGtPG t (19) 223 d db xt f xPxtPItI t (20) 31 d d It f InItGthtUt t (21) 1 41 1 d1 d Ut f UUtUt ta (22) The above equations can be written and arranged as follows. The above equations can be further simplified as 11 1b f GPGtxtGt PG (23) 2233b f xPxtPItPI (24) 31 f InIttGthtUt (25) 41 1 11 f UUtUt aa (26) The above system is a nonlinear system due to the presence of the nonlinear term that appears in Eq.23. The nonlinear term is x tGt. Now let us make the following definitions 12341 ,,, x t Gtxt xtxtItxt ut Then Eqs.23 to 26 become 1 111 21 d db xt P xt xtxt PG t (27) 2 2233 3 d db xt Px tPxtPI t (28) 3 134 d d xt tx tnxtxtht t (29) 4 4 d11 d xt x tut taa (30) The Jacobian matrices ( x J and u J ) of the model can be derived as 12 1 23 00 0 0 0 0 0 1 1 0 0 0 , x Px x PP tn a x xuu J and 00 0 0 0 1 , u a x xuu J (31) where the point (x0, u 0) is the equilibrium point. The equilibrium point can be calculated by setting the state equations to zero and solving 11010 2010 b PxxxPG (32) 220 330 30 b PxPx PI (33) 1030 400tx nxxht (34) 40 0 11 0xu aa (35) where x10, x20, x30, x40 and u0 are the values of the state variables and the input at the operating point (i.e. the equilibrium point). At the equilibrium point, u0 = 0, Eq.35 becomes 40 10x a , that gives x40 = 0 (36) Substituting the value of x40 in Eq.34 results in 10 300tx nxht and 10 30 x ht xn (37) The value of x30 can be substituted in Eq.33 10 220 330 b xht Px PPI n to obtain 31033 20 222 b PtxPth PI xPnPn P (38) Now, by substituting the value of 20 x in Eq.32 we have 2 333 10110 1 222 0 b b PtPthPI xP xPG PnPn P (39) The above equation is a 2nd order equation and can be solved by using the quadratic formula 2 10 4 2 bb ac xa (40) where 333 11 222 ,, b b PtPth PI abP cPG PnPn P There are 2 possible values of x10. Since x20 and x30 are expressed in term of x10, there will be 2 possible values for each. Based on that, the controllability test will be studied to check which value of x10 is accepted. 3.3. A Case Study The simulation diagram shown in Figure 1 was used to simulate the data of a diabetic patient. The value of the parameters for a diabetic patient is shown below. P1 = 0, P2 = 0.81/100, P3 = 4.01/1000000, I0 = 192, G0 = 337, = 2.4/1000, h = 93, n = 0.23, a = 2, Gb = 99, Ib = 8, [10]. ![]() A. Hariri et al. / J. Biomedical Science and Engineering 4 (2011) 297-314 Copyright © 2011 SciRes. JBiSE 304 The output of the simulated system is shown in Figure 5. By examining Figure 5, it can be clearly seen that the glucose level does not come down to the basal level after injecting an amount of 337 mg/dL of glucose inside a diabetic patient. Figure 5 shows that the level of the glucose inside a diabetic patient decreases for the first 100 minutes, starts increasing afterward, and reaches the value of around 310 mg/dL after 3 hours from the time the glucose was injected. The goal is to bring down the value of the glucose inside a diabetic patient to the nor- mal level or at least into a small neighborhood of the basal level. The above goal can be achieved by design- ing a PID feedback controller. The controller is to regu- late the infusion rate and inject the required amount of the insulin inside the diabetic patient, and in turn the insulin will work inside the patient to bring down the level of glucose to the normal level or at least to the neighborhood of the normal level. 4. CONTOL DESIGN We start with a linearized state space model for our sys- tems. The general form of the state space is defined in the following equation: x xu yxu AB CD (41) The proposed mathematical model at the equilibrium point (x0, u 0) can be written in the state space form as shown below: 120 10 23 0 00 0 00 0 0 1 1 1 0 0 0 Px x PP xx tn a a 1 0 0 0 u yx (42) Figure 5. Output of the simulated system for diabetic. where u is the input and y is the output of the system. The data of a diabetic person shown in Section 3.3 was used and the equilibrium point (x0, u0) was calculated as time varies from t = 1 min to t = 182 min. The two val- ues for x10 that were calculated before were substituted in Eq.42 and the controllability test was performed. It was found that only one value (the one obtained from the sign) makes the system to be controllable, hence only this value is used in the subsequent development. 4.1. Design of PID Controllers for Diabetic Patient When designing a controller, the designer must define the specifications that need to be achieved by the con- troller. Normally the maximum overshoot (Mp) of the system step response should be small. Commonly a range between 10% and 20% is acceptable. Also the set- tling time (ts), is an important factor. The objective here is to design a PID controller, so that the closed-loop sys- tem has the following specifications: small steady-state error for a step input; less than 10% overshoot; settling time less than 60 minutes. The patient dynamic system was expressed in the state space representation in (42). For an overshoot less than 10%, a damping ratio must be greater than 0.59, and a settling time less than 60 minutes implies that n must be greater than 0.067. The PID Controllers can be designed based on the models at different operating points. The following list contains the models at t = 1, 20, 40, 60, 90, 120, 150 and 182 minutes, and the corresponding PID controllers. Since B and C do not change with time, they are fixed in all cases as: 0 0 and 1 0 0 0 0 0.5 BC The control design is done by applying the root locus method and then evaluated by using the step response. For example, if the model at t = 1 min is used, after sub- stituting the above numerical values (42), the root locus plot can be generated by the Matlab functions MATLAB Program Plotting the open loop system Root locus using MATLAB [num, den] = ss2tf(A,B,C,D]; rlocus(num, den) axis([–0.6 0.1 –0.5 0.5]) sgrid(0.59,0) sigrid(0.067) ![]() A. Hariri et al. / J. Biomedical Science and Engineering 4 (2011) 297-314 Copyright © 2011 SciRes. JBiSE 305 The open loop poles are shown in Figure 6. These poles are located at –0.0040 + 0.0045j, –0.0040 – 0.0045j, –0.2302, –0.5. The four poles are stable, but the first two poles are very close to the imaginary axis and hence represent the slowest dynamics. The controller takes the form 12 K Sz Sz Gc SS where K is the value of the gain where the root locus intersects with the line of the damping ratio. The z1 and z2 represent the value of the zeros to be added and may be selected to cancel the slowest poles of the dynamic system. Hence, select z1 = –0.004 + 0.0045j, z2 = –0.004 – 0.0045j. The controller is * (please see the equation below) resulting in the system matrix A and the PID controller as 1 0 859.6667 0 0 0 0.0081 0.00000401 0 0.0024 0 0.23 1 0 0 A 2 1 , 0 0.5 0.008 0.00003625 C KS S GS S The PID Parameters are: 4 0.0444, 2.009410, 5.59 pi d KK K The design specifications of the system require the maximum overshoot to be less than 10% and the settling time to be less than 60 minutes. After inserting the PID controller in series with the patient system and connect- ing them in a unity feedback, we get the maximum overshoot 5.71% and the settling time 47.5 minutes (see Figure 7). t = 20 minutes: 20 0 131.33 0 0 0 0.0081 0.00000401 0 0.048 0 0.23 1 0 A 2 20 , 0 0 0.5 0.0076 0.0001 C KS S GS S The PID Parameters are: = 0.2160, 0.0031, = 28.4 pi d KKK Figure 6. Output of the simulated system for Root Locus. Figure 7. Output of the unit step response, using the model at t= 1 minute with K = 5.5. t = 40 minutes: 40 0 112.17 0 0 0 0.0081 0.00000401 0 0.096 0 0.23 1 0 A 2 40 , 0 0 0.5 0.0072 0.0002 C KS S GS S The PID Parameters are: = 0.2374, 0.0061, = 32.7 pid KKK * 0.0040 0.00450.0040 0.0045 K SjSj Gc SS ![]() A. Hariri et al. / J. Biomedical Science and Engineering 4 (2011) 297-314 Copyright © 2011 SciRes. JBiSE 306 t = 60 minutes: 60 0 105.78 0 0 0 0.0081 0.00000401 0 0.1440 0 0.23 1 0 A 2 60 , 0 0 0.5 0.0070.0003 C KS S GS S The PID Parameters are: 4 0.0444, 2.009410, 5.59 pi d KKK t = 90 minutes: 90 0 101.52 0 0 0 0.0081 0.00000401 0 0.2160 0 0.23 1 0 A 2 90 , 0 0 0.5 0.0064 0.0004 C KS S GS S The PID Parameters are: = 0.2331, 0.0138, = 36.4 pi d KK K t = 120 minutes: 120 0 99.39 0 0 0 0.0081 0.00000401 0 0.288 0 0.23 1 0 0 A 2 120 , 0 0.5 0.0058 0.0005 C KS S GS S The PID Parameters are: = 0.2234, 0.0187, = 37.9 pi d KKK t = 150 minute: 150 0 98.11 0 0 0 0.0081 0.00000401 0 0.36 0 0.23 1 0 0 A 2 150 , 0 0.5 0.0054 0.0006 C KS S GS S The PID Parameters are: = 0.2032, 0.0229, = 37.7 pi d KKK t = 182 minute: 182 0 97.21 0 0 0 0.0081 0.00000401 0 0.4368 0 0.23 1 0 0 A 2 182 , 0 0.5 0.0048 0.0007 C KSS GS S The PID Parameters are: = 0.1870, 0.0281, = 38.5 pid KK K 4.2. Individual PID Controllers Non-adaptive PID controllers use a fixed PID controller for the entire control period and rely on its robustness to maintain control performance. For each individual PID controller (with its transfer function found in the previ- ous subsection for t = 1, 20, 40, 60, 90, 120, 150 and 182 min) the simulation results on Gt and ut are shown below in Figur es 8-15. Under the individual PID controllers, the output Gt, the glucose level, did not really meet the design specification, and the glucose level is not near or at least in a small neighborhood of the glucose basal level. The overshoot of the system was too high and beyond the acceptable level. Also the settling time was not even close to where it should be as per the design requirement. And the steady state error was not satisfactory. A new method should be developed and implemented to meet all the design specifications. This method is explained in detail in the following section. 4.3. Switched PID Controllers The individual PID controllers could not lower the glu- cose level Gt of the patient to the neighborhood of the glucose basal level. Consequently, we introduce a new control-switching scheme that adapts controllers to meet design specifications. The switching control con- sists of the following items: One “time clock,” one “switch case” block, one “if action case” block, one “merge” block, and eight “off-on switches.” All these blocks are connected together to form the wiring dia- gram of the Control-Switching Scheme. The functions of the control switching scheme is de- tailed in Figure 16. The “Time Clock” is to provide the “Switch Case” block with time as a signal input to acti- vate it. The “Switch Case” block receives a single input from the clock, which it uses to form case conditions that de- termine which subsystem to execute. Each output port case condition is attached to a Switch Case Action sub- ![]() A. Hariri et al. / J. Biomedical Science and Engineering 4 (2011) 297-314 Copyright © 2011 SciRes. JBiSE 307 Figure 8. Gt and ut using 1 C G. Figure 9. Gt and ut using 20 C G. Figure 10. Gt and ut using 40 C G. ![]() A. Hariri et al. / J. Biomedical Science and Engineering 4 (2011) 297-314 Copyright © 2011 SciRes. JBiSE 308 Figure 11. Gt and ut using 60 C G. Figure 12. Gt and ut using 90 C G. Figure 13. Gt and ut using 120 C G. ![]() A. Hariri et al. / J. Biomedical Science and Engineering 4 (2011) 297-314 Copyright © 2011 SciRes. JBiSE 309 Figure 14. Gt and ut using 150 C G. Figure 15. Gt and ut using 182 C G. Figure 16. Control-Switching Scheme Simulation Diagram. ![]() A. Hariri et al. / J. Biomedical Science and Engineering 4 (2011) 297-314 Copyright © 2011 SciRes. JBiSE 310 system. The cases are evaluated top-down, starting with the top case. If a case value corresponds to the actual value of the input, its Switch Case Action subsystem is executed. The “Switch Case” model is divided into eight time interval zones as shown in Table 6. The “If Action Case” block consists of eight “If- Ac- tion Subsystem.” The “If-Action Case” implements Ac- tion Subsystems used in if-statement and switch control flow statements. Action Subsystems execute their pro- gramming in response to the conditional outputs of an If-statement or Switch Case block. A schematic diagram of the “If Action Case” block is shown in Figure 17. The “Merge” block combines its inputs into a single output line whose value at any time is equal to the most re- cently computed output of its driving blocks. The num- ber of inputs can be specified by setting the block’s in- puts parameter. The “Off-On Switches” are used to turn the PID controllers OFF or ON. In general when the time clock is running, it feeds the “Switch Case” block with an input signal which in turn switches on the “If-Action Case” block as per the time interval that was specified in Table 6. Based on the status of the “If-Action Case”, a specific PID controller will be turned on and executed to control the output of the system. For zone 1, the time interval is between 0 - 1 minute, during this period of time the “Switch Case” is enabling input “In9” of the “If-Action Case”. When input “In9” is enabled, it will only execute the input “In1” to the output “Out1”. The input “In1” is connected to the first PID Controller. That means only the first PID Controller, (PID Controller 1), is working. At the end of the first minute, the “Switch Case” will switch to zone 2 which runs from the beginning of the minute number 2 and will last until the end of the minute number 20. During this period of time, the “Switch Case” is enabling input “In10” of the “If-Action Case”. When input “In10” is enabled, it will only execute the input “In2” to the output “Out2”. The input “In2” is connected to the second PID Controller. That means only the second PID Controller, (PID Controller 20), is working. The same procedure will be followed until the “Switch Case” switches be- tween the eight time zones that were specified in Table 6. In turn the PID Controllers will be executed based on the status of the “If-Action Case”. The Control-Switching Scheme Diagram shown in Figure 16 was simulated with all the PID controllers executed (connected to the circuit). The output Gt of the system is shown in Figure 18. It can be clearly seen that the PID controllers are able to bring the glucose level from 337 mg/dL to the basal level (99 mg/dL) within 40 minutes. But in about 70 minutes the value of the glucose starts going below the basal level, and it Table 6. Swithing Time Interval. Zone number Time Interval (minutes) 1 0 - 1 2 2 - 20 3 21 - 40 4 41 - 60 5 61 - 90 6 91 - 120 7 121 - 150 8 150 - 182 went further below the minimum value that the glucose can be at. In this case the person will be classified as a patient with the hypoglycemia (low sugar), and that is not acceptable. The Control-Switching Scheme Diagram was then simulated in which all the PID controllers are executed except the eighth PID Controller 182 C G. The graph of the output, Gt of the system is shown in Figure 19. It can be seen the same problem still exists. Again in this case the person will be classified as a patient with the hypoglycemia. The same procedure was repeated but with all the PID controllers executed except the PID Controllers 150 C G and 182 C G with the graph of the output shown in Figure 20; and excluding 120 C G, 150 C G, and 182 C G with the simulation result shown Figure 21. Again in both cases the person will be classified as a patient with the hypo- glycemia. When we exclude 90 C G, 120 C G, 150 C G, and 182 C G, the output Gt of the system, shown in Figure 22,. reaches the glucose basal level (99 mg/d/L) within 40 minutes, and it stays in that neighborhood. The input of the PID Controller system is shown in Figure 23. For verification, the same control strategy is evaluated on patient #2. Following the same modeling procedure that was performed for the diabetic patient #1, the model parameters are identified as P1 = 0, P2 = 0.42/100, P3 = 2.56/1000000, I0 = 209, G0 = 297, = 3.72/1000, h = 154, n = 0.22, a = 2, Gb = 100, Ib = 8, [10]. Without control, the above data was implemented in model simulation. The output of the simulation diagram is shown in Figure 24, which shows that without proper control the glucose level does not come down to the basal level after injecting an amount of 297 mg/dL of glucose inside a diabetic patient. The level of the glucose inside a diabetic patient decreases for the first 120 min- utes and starts increasing afterward and reaches the value of about 270 mg/dL after 3 hours from the time the glucose was injected. The same control-switching scheme that was per- formed for diabetic patient #1 is repeated for diabetic patient #2. The values of the parameters for the first four PID controllers (at t =1, 20, 40 and 60 minutes) are ![]() A. Hariri et al. / J. Biomedical Science and Engineering 4 (2011) 297-314 Copyright © 2011 SciRes. JBiSE 311 8 Out8 7 Out7 6 Out6 5 Out5 4 Out4 3 Out3 2 Out2 1 Out1 PID PI D Controller 90 PID PI D Controller 60 PID PI D Controller 40 PID PI D Controller 20 PID PID Controller 180 PID PID Controller 150 PID PID Controller 120 PID PI D Controller 1 1 In1 8 Out8 7 Out7 6 Out6 5 Out5 4 Out4 3 Out3 2 Out2 1 Out1 case: { } In1Out1 If Actio n Subsystem 90 case: { } In1Out1 If Actio n Subsystem 60 case: { } In1Out1 If Actio n Subsystem 40 case: { } In1Out1 If Actio n Subsystem 20 case: { } In1Out1 If Actio n Subs y stem 180 case: { } In1Out1 If Actio n Subsystem 150 case: { } In1Out1 If Actio n Subsystem 120 case: { } In1Out1 If Actio n Subsystem 1 16 In1 6 15 In1 5 14 In1 4 13 In1 3 12 In1 2 11 In1 1 10 In1 0 9 In9 8 In8 7 In7 6 In6 5 In5 4 In4 3 In3 2 In2 1 In1 Figure 17. “If Action Case” system and PID controller switching function module. summarized in Table 7. The control-switching scheme was simulated for dia- betic patient #2 by using only the first four PID control- lers. The output Gt of the system is shown in Figure 25. The output Gt reaches the glucose basal level (100 mg/d/L) within 60 minutes and it stays in that neighborhood. The graph of the input of the PID Con- troller system is shown in Figure 26. Based on the simulation results, although adaptive con- trol can potentially improve control performance, it is ![]() A. Hariri et al. / J. Biomedical Science and Engineering 4 (2011) 297-314 Copyright © 2011 SciRes. JBiSE 312 Figure 18. Plot of G(t) when all PID Controllers are executed. Figure 19. Plot of G(t) when all PID Controllers except con- troller 182 C G are executed. Figure 20. Plot of G(t) when all PID Controllers except con- troller 150 C G and 182 C G are executed. Figure 21. Plot of G(t) when all PID Controllers except controller 120 C G, 150 C G and 182 C G are executed. Figure 22. Plot of G(t) when all PID Controllers except controller 90 C G, 120 C G, 150 C Gand 182 C G are executed. Figure 23. Plot of u(t) when all PID Controllers except controller 90 C G, 120 C G, 150 C Gand 182 C G are executed. ![]() A. Hariri et al. / J. Biomedical Science and Engineering 4 (2011) 297-314 Copyright © 2011 SciRes. JBiSE 313 Figure 24. Output of the Simulation Diagram for diabetic pa- tient #2. Figure 25. Plot of G(t) when PID Controllers 1, 20, 40 and 60 are executed. Figure 26. Plot of u(t) when PID Controllers 1, 20, 40 and 60 are executed. Table 7. PID Controller for diabetic patient #2. PID Controllers Parameters PID at t = 1 min PID at t = 20 min PID at t = 40 min PID at t = 60 min Kp 0.0412 0.1069 0.0868 0.0768 Ki 2.7402×10-4 0.0047 0.0086 0.014 Kd 10.1 30.6 33.1 33.6 sometimes unnecessary, or even harmful when swit- ching overly frequently. Our results show that when the switching scheme is limited to the first four PID con- trollers, the performance is in fact enhanced. This may be related to the fact that some PID controllers are more robust with respect to the model variations. On the other hand, in comparison to individual controllers, the con- trol-switching scheme achieves design specification while all individual controllers fail to deliver the re- quired performance. 5. CONCLUSIONS This study reveals that typical PID controllers may not be sufficient to deliver satisfactory control performance in glucose level control problems. This is mainly due to the nonlinear nature of patient dynamic models and limited robustness of the PID controllers. An adaptive control that switches controllers based on operating conditions can potential enhance control performance. However, the switching control scheme must be carefully designed to ensure that control specifications be met. Our results show that overly frequent switching of controllers may have detrimental effects on control performance. We show that by reducing the number of PID controllers in the switch- ing scheme, not only control complexity is reduced, but performance is actually enhanced. REFERENCES [1] Karam, J.H., Grodsky, G.M. and Forsham, P.H. (1963) Excessive insulin response to glucose in obese subjects as measured by immunochemical assay. Diabetes, 12, 196-204. [2] Ginsberg, H., Olefsky, J.M. and Reaven, G.M. (1974) Further evidence that insulin resistance exists in patients with chemical diabetes. Diabetes, 23, 674-678. [3] Reaven, G.M. and Olefsky, J.M. (1977) Relationship between heterogeneity of insulin responses and insulin resistance in normal subjects and patients with chemical diabetes. Diabetologia, 13, 201-206. doi:10.1007/BF01219700 [4] Lerner, R.L. and Porte, D. (1972) Acute and steady state insulin responses to glucose in nonobese, diabetic sub- jects. Journal of Clinical Investigation, 51, 1624-1631. doi:10.1172/JCI106963. [5] Reaven, G.M. (1980) Insulin-independent diabetes mellitus: Metabolic characteristics. Metabolism Clinical and Ex- ![]() A. Hariri et al. / J. Biomedical Science and Engineering 4 (2011) 297-314 Copyright © 2011 SciRes. JBiSE 314 perimental, 29, 445-454. doi:10.1016/0026-0495(80)90170-5 [6] Bergman, R.N. and Cobelli, C. (1980) Minimal modeling, partition analysis, and the estimation of insulin sensitivity. Federation Proceedings, 39, 110-115. [7] Shen, S-W., Reaven, G.M. and Farquhar, J.W. (1970) Comparison of impedance to insulin-mediated glucose uptake in normal and diabetic subjects. Journal of Clini- cal Investigation, 49, 2151-2160. doi:10.1172/JCI106433. [8] Bergman, R.N., Ider, Y.Z., Bowden, C.R. and Cobelli, C. (1979) Quantitative estimation of insulin sensitivity. Ameri- can Journal of Physiology, 236, E667-E677. [9] Pacini, G. and Bergman, R.N. (1986) MINMOD, A computer program to calculate insulin and pancreatic re- sponsivity from the frequently sampled intravenous glu- cose tolerance test. Computer Methods and Programs in Biomedicine, 23, 113-122 [10] Bergman, R.N., Phillips, L.S. and Cobelli, C. (1981) Physiologic evaluation of factors controlling glucose tol- erance in man, measurement of insulin sensitivity and -cell glucose sensitivity from the response to intrave- nous glucose. Journal of Clinical Investigation, 68, 1456 -1467. doi:10.1172/JCI110398 [11] Bergman, R.N. and Urquhart, J. (1971) The pilot gland approach to the study of insulin secretory dynamics. Re- cent Progress in Hormone Research, 27, 583-605. [12] Nomura, M., Shichiri, M., Kawamori, R., Yamasaki, Y., Iwama, N. and Abe, H. (1984) A mathematical insulin- secretion model and its validation in isolated rat pancre- atic islets perifusion. Computers and Biomedical Re- search, 17, 570-579. doi:10.1016/0010-4809(84)90021-1 [13] Buchanan, T.A., Metzger, B.E., Freinkel, N. and Berg- man, R.N. (1990) Insulin sensitivity and B-cell respon- siveness to glucose during late pregnancy in lean and moderately obese women with normal glucose tolerance or mild gestational diabetes. American Journal of Ob- stetrics and Gynecology, 162, 1008-1014. [14] Lynch, S.M. and Bequette, B.W. (2001) Estimation-based model predictive control of blood glucose in type i dia- betics: A simulation study. IEEE Transations on Bio- medical Engineering conferences, 79-80. [15] Fisher, M.E. (1991) A semi-closed loop algorithm for the control of blood glucose levels in diabetes. IEEE Transa- tions on Biomedical Engineering conferences, 57-61. [16] Furler, S.M., Kraegen, E.W., Bell, D.J., Smallwood, R.H., and Chisolm, D.J. (1985) Blood glucose control by in- termittent loop closure in the basal mode: Computer simulation studies with a diabetic model. Diabetes Care, 8, 553-561. [17] Ibbini, M.S., Masadeh, M.A. and Amer, M.M.B. (2004) A semiclosed-loop optimal control system for blood glu- cose level in diabetics. Journal of Medical Engineering & Technology, 28, 189-196. doi:10.1080/03091900410001662332. [18] Furler, S.M., Kraegen, E.W., Smallwood, R.H. and Chisholm, D.J. (1985) Blood glucose control by inter- mittent loop closure in the basal mode: computer simula- tion studies with a diabetic model. Diabetes Care, 8, 553- 561. doi:10.2337/diacare.8.6.553 [19] Insel, P.A., Liljenquist, J.E., Tobin, J.D., Sherwin, R.S., Watkins, P., Andres, R. and Berman, M. (1975) Insulin control of glucose metabolism in man. Journal of Clini- cal Investigation, 55, 1057-1066. doi:10.1172/JCI108006 [20] Grodsky, G.M. (1972) A threshold distribution hypothesis for packet storage of insulin and its mathematical mod- eling. Journal of Clinical Investigation, 51, 2047-2059. doi:10.1172/JCI107011 [21] Sherwin, R.S., Kramer, K.J., Tobin, J.D., Insel, P.A., Liljenquist, J.E., Berman, M. and Andres, R. (1974) A model of the kinetics of insulin in man. Journal of Clini- cal Investigation, 53, 1481-1492. [22] Turner, R.C., Holman, R.R., Mathews, D., Hockaday, T.D.R. and Peto, J. (1979) Insulin deficiency and insulin resistance interaction in diabetes: Estimation of their relative contribution by feedback analysis from basal in- sulin and glucose concentrations. Metabolism Clinical and Experimental, 28, 1086-1096. doi:10.1016/0026-0495(79)90146-X [23] Lerner, R.L. and Porte, D. (1971) Relationships between intravenous glucose loads, insulin responses and glucose disappearance rate. Journal of Clinical Endocrinology & Metabolism, 33, 409-417 [24] Bolie, V.W. (1961) Coefficients of normal blood glucose regulation. Journal of Applied Physiology, 16, 783-788. [25] Guyton, A.C. and Hall, J.E. (1996) Text book of medical physiology, 9th Edition, Saunders. [26] The American Diabetes Association is a US leading non- profit health organization providing diabetes research, information and advocacy since 1940. [27] Gaetano, A.D. and Arino, O. (2000) Mathematical mod- eling of the intravenous glucose tolerance test. Journal of Mathematical Biology, 40, 136-168. doi:10.1007/s002850050007 [28] Marquardt, D.W. (1963) An algorithm for least-squares estimation of non-linear parameters. Journal of the Soci- ety for Industrial and Applied Mathematics, 11, 431-441. doi:10.1137/0111030 [29] Goodwin, C. and Payne, R.L. (1977) Dynamic System Identification. Academic Press, Inc., New York. [30] Lourakis, M. (2005) A brief description of the leven- berg-marquardt algorithm implemented by levmar. Insti- tute of Computer Science, Foundation for Research and Technology. [31] De Groot, M.H. and Schervish, M.J. (2002) Probability and statistics. 3rd Edition, MA: Addison-Wesley, Boston. |