J. Biomedical Science and Engineering, 2011, 4, 282-288 JBiSE
doi:10.4236/jbise.2011.44038 Published Online April 2011 (http://www.SciRP.org/journal/jbise/).
Published Online April 2011 in SciRes. http://www.scirp.org/journal/JBiSE
Carpal tunnel syndrome diagnosis: validation of a clinic-based
nerve conduction measurement device
Timothy P. Green1, Mika Kallio2, Malcolm R. A. Clarke1, Pankaj Pathak1, Ve i jo Lesonen3,
Uolevi Tolon en2
1Department of Orthopaedics, Leicester General Hospital, Leicester, England;
2Department of Clinical Neurophysiology, Oulu University Hospital, Oulu, Finland;
3Mediracer Ltd, Oulu, Finland.
Email: Mika.Kallio@oulu.fi
Received 1 February 2011; revised 3 March 2011; accepted 8 March 2011.
ABSTRACT
Background: Carpal Tunnel Syndrome (CTS) is the
commonest upper limb nerve entrapment synd rome
seen in practice. In many centres, nerve conduction
studies (NCS) have been adopted as a routine part
of the diagnostic process. In the United Kingdom,
the time taken to access diagnostic tests has been
likened to a “hidden waiting list”, lengthening the
time taken for a patient to access treatment. In the
current healthcare climate with a centrally driven
aim to reduce patient waiting time to a max imum of
eighteen weeks, including tests, such waiting is even
more unacceptable. Aim. This study was performed
in order to evaluate a simple handheld device for
quantifying median nerve lesions in CTS. Design of
study: A prospective blinded cohort study. Setting:
Leicester General Hospital, Carpal Tunnel Service
Method: Participants were recruited from the nor-
mal referral stream. If the clinical findings were
consistent with a diagnosis of CTS, they were for-
mally consented to the study in which results from
the new handheld device were compared with tra-
ditional NCS. Final test group consisted of 63 par-
ticipants. Results: For the new device the correct
positive detection rate for abnormal nerve conduc-
tion was 91% (74/81 hands). Of the seven abnormal
results not picked up by the new device, four were
in asymptomatic hands (positive per cent agree-
ment in symptomatic hands 95%). There were no
false positives with the new system. (Negative per
cent agreement 100%) Conclusion: We conclude tha t
this new device demonstrates a high degree of con-
cordance with currently available traditional NCS.
The study suggested ways in which the accuracy
could be further improved.
Keywords: Carpal Tunnel Syndrome; Nerve Conduction
Studies; Diagnosis; Portable
1. INTRODUCTION
Carpal Tunnel Syndrome (CTS) is the most common
upper limb nerve entrapment syndrome [1,2]. Diagnosis
is largely based on symptoms and clinical signs, but
these are of variable sensitivity and specificity. In many
centres, nerve conduction studies (NCS) have been adopted
as a routine part of the diagnostic process despite evi-
dence that up to 16% of patients may have normal stud-
ies in the presence of clinically proven CTS [3,4]. In-
sisting on NCS for every patient satisfies the clinicians’
need for objective evidence but adds a significant extra
expense and delay to the diagnosis and treatment of this
very common condition. It may also introduce a false
negative, failing to diagnose CTS when it is present [1].
In the United Kingdom, the time taken to access di-
agnostic tests is several weeks. With the current centrally
driven aim to reduce patient treatment time, including
tests, to a maximum of eighteen weeks, delays of months
are unacceptable.
A new, portable device that can simply and cheaply
diagnose a median nerve lesion and quantify its severity
in the clinic may have a role in establishing the diagnosis.
In conjunction with clinical findings this could reduce
both treatment times and the number of hospital visits. A
good practice guide produced by the Department of
Health [5] suggests that this device is “worthy of further
assessment”.
This paper describes a study designed to compare re-
sults from the new device with traditional NCS.
1.1. Background
Our Carpal Tunnel Service was set up in 1999 in order
T. P. Green et al. / J. Biomedical Science and Engineering 4 (2011) 282-288
Copyright © 2011 SciRes. JBiSE
283
to improve the service we provided to those affected by
this condition in our local community. It has proved to
be very successful, reducing overall treatment time
from 106 weeks to 6 weeks. Over six and a half thou-
sand people have been treated to date. The achieve-
ments of the service have been presented at local and
national meetings and have attracted recognition from
the Department of Health.
In Oulu, Finland, a group of Clinical Neurophysiolo-
gists developed a hand-held device for diagnosing the
median nerve lesion in CTS based on the work of Uncini
et al. [6] A clinical trial to prove its effectiveness was
performed [7], but was slow to recruit patients in a
country with a population of only five million people.
Finpro, an agency set up to support Finnish innovation
contacted the Department of Health in the UK to locate
clinical partners that could help validate the system fur-
ther. This study is the result of a collaborative work be-
tween our unit and Oulu.
2. METHODS
The study was based in secondary care in a single hos-
pital that is the normal base for the Carpal Tunnel Ser-
vice. Approval was sought from and granted by the local
Ethics Committee.
The study set out to compare results from the new de-
vice with traditional nerve conduction studies. The new
device is a portable, hand-held instrument designed to be
used by non-specialists and so was operated by local
clinicians after a brief familiarisation period. Experi-
enced clinical neurophysiologists from Finland carried
out the traditional nerve conduction studies for com-
parison.
The null hypothesis was that there would be no sig-
nificant difference between the two systems.
In order to overcome the practicalities in getting the
two research teams together, this prospective cohort
study was set up to be completed over three days. Con-
centrating the measurements into a short time avoided
some of the variables in testing, such as ambient tem-
perature.
All participants had both hands tested by each method
regardless of whether or not both hands were sympto-
matic. During testing, the British and Finnish groups
were blinded to the results obtained by the other group.
The study was submitted to the local ethics committee
(University Hospitals of Leicester) January 2006 and
passed by them on first review.
2.1. Participants
Participants were recruited from the normal referral
stream to our Carpal Tunnel Service. They were referred
by their General Practitioners or from other Specialists
within secondary care. The referring doctors all thought
their patients exhibited symptoms suggestive of carpal
tunnel syndrome. (i.e. tingling and numbness affecting
the radial side of the hand). There were no other exclu-
sion criteria.
All potential participants (n = 120) were sent a letter
inviting them to take part in the study. All those that ac-
cepted (n = 104) were clinically assessed. A symptom
form was completed which included the distribution of
any sensory disturbance. If the clinical findings were
consistent with a diagnosis of CTS, they were formally
recruited to the study, consented and allocated onto one
of the three study days. Any participant whose clinical
assessment suggested an alternative diagnosis, a more
comprehensive problem than a mere suspicion of CTS,
or who declined to continue in the study was referred on
as appropriate or transferred into the standard Carpal
Tunnel Service.
The final number of participants recruited into the
study was 65 of whom 49 were female and 16 were male.
The age range was from 23 to 83 (mean 52.3) years. Two
of the recruited cohort had to cancel on the day of their
study because of personal reasons, leaving a test group
of 63 (48 females, 15 males, aged 23 - 81, mean 52.8)
equating to a total of 126 tested hands. Clinical symp-
toms of CTS were unilateral in 34 participants and bilat-
eral in 29.
2.2. Interventions
Each participant had base line measurements of weight
and height recorded before being allocated to one of the
Clinical Neurophysiologists so that a traditional set of
neurophysiological measurements could be carried out.
Orthodromic sensory conduction latencies were meas-
ured by stimulating index and ring fingers with felt pad
electrodes and recording at wrist 1 cm above the distal
wrist groove. This is the method used also by the new
device. Additionally, the traditional measurement in-
cluded mixed conduction velocities of the median and
ulnar palm to wrist segment of 8 cm. Finally, median and
ulnar distal motor latencies were examined with 7 cm
interelectrode distance. (Measurements were made using
Keypoint® 4 and Keypoint® Portable, Medtronic, Skov-
lunde, Denmark). The participants were then randomly
allocated to one of the local clinicians so as to have
nerve conduction studies carried out using the new de-
vice. (Mediracer®, Mediracer International, Oulu, Finland).
The new device uses two self-adhesive, disposable hy-
drogel coated electrodes. One is a stimulating electrode
that wraps around the proximal phalange of the finger
being tested and the second is a recording electrode that
is placed over the wrist 1 cm above the distal wrist
groove (Figure 1).
T. P. Green et al. / J. Biomedical Science and Engineering 4 (2011) 282-288
Copyright © 2011 SciRes. JBiSE
284
Figure 1. The new device (Mediracer®, Mediracer Interna-
tional, Oulu, Finland) uses two self-adhesive, disposable hy-
drogel coated electrodes. One is a stimulating electrode that
wraps around the finger being tested and the second is a re-
cording electrode that is placed over the wrist.
Tests are made on the ring finger which is innervated
by both median and ulnar nerves plus the index finger
where there is only median nerve innervation. The hand-
piece is attached to the electrodes and has an automatic
menu for the operator. The test begins once background
muscle activity drops below threshold.
A stimulus is gradually increased until the participant
becomes just aware of it. The stimulus is then automati-
cally increased by 250% and repeated sixty-four times
over 35 seconds. Overall test time is in the order of five
minutes. The handset shows how many stimuli were
successfully received, allowing the test to be repeated if
necessary. Data is downloaded by a Bluetooth™ link to
a computer, where the software calculates and displays a
nerve conduction latency/amplitude graph (Figure 2).
Figure 2. Double peak shown after ring finger stimulation
(blue line). Interpeak distance 1.0 ms is consistent with
mild median nerve damage (peak latencies rounded).
Forefinger response (red line) is also significantly longer
(0.9 ms) when compared to the first peak from ring finger.
All the graphs were visually checked off line and peak
latency cursors reset if needed by a specialist in clinical
neurophysiology who was not aware of any other patient
data. The peak latencies needed adjustments in 13 pa-
tients.
Following the two different NCS tests, the participants
were seen by one of two other local clinicians to review
the results of their tests and decide their further man-
agement.
2.3. Main Outcome Measures
Each participant was assessed to determine whether their
symptoms were suggestive of CTS. The presence of
symptoms in either hand was recorded, as was any pre-
vious surgery. Hand dominance and occupation were
noted. The Boston Hand Score [8] for CTS was used to
quantify symptom severity (SSS) and functional severity
(FSS).
The neurological severity of CTS was classified in a
simplified manner. In the present study neurophysi-
ological severity classification was obtained from tradi-
tional measurements by using Padua et al.’s [4] five ab-
normal NCS classes which have been reduced to three-
mild (includes minimal), moderate and severe (includes
extreme). The reasons for this are discussed later. In our
series there were nine minimal lesions included to the
mild class and one extreme case included to the severe
group. In the new device study the same three severity
classes were scaled using reference values obtained from
the earlier study [7]. These were determined fitting the
new device data to the Padua et al.’s [4] scales of the
traditional device measurements. For the new device,
mild and moderate severity can be assessed using only
measurements from the ring finger, as separate responses
from median and ulnar nerves are easily distinguished as
seen in Figure 2. In severe cases however, no median
nerve signal may be seen. In this case, adding the result
from stimulation of the forefinger will confirm the ab-
sence of a median signal and avoid missing a severe case.
(Ta b l e 1 ) The absolute values for the new device study
latencies are taken from the previous study [7].
In order to test the agreement of the similarly meas-
ured nerve responses between the Mediracer and Key-
point devices, scatter plots with linear regression line fit
and Bland-Altman plots [9] for the latency differences
between ring finger median (4PM) and ulnar nerve (4PU)
responses and ring finger (4PU) and index finger (2P)
responses were calculated. This was analysed using R
statistics and OriginPro 8 software. Pearson’s and intra-
class correlation coefficients were also calculated using
the SPSS 17.0 software.
3. RESULTS
There were 92 symptomatic hands in 63 participants, of
T. P. Green et al. / J. Biomedical Science and Engineering 4 (2011) 282-288
Copyright © 2011 SciRes. JBiSE
285
Table 1. Reference values for latency differences using the
new device.
Latency differences between ring finger peaks
Normal Mild Moderate
Severe
0.7 ms 0.8 - 1.1 ms 1.2 ms
A single ulnar
derived peak.
No apparent
median response
Latency differences between ring and forefinger peaks
Normal Mild Moderate
Severe
0.5 ms 0.6 - 0.9 ms 1.0 ms
Confirmation of
no median peak
from forefinger
Limits for carpal tunnel severity classes using the new device. If both fin-
gers are stimulated, as they were in this study, it will be seen that different
classifications might be arrived at for the same hand. In such a case the
more severe classification was accepted.
whom 21 were right, 13 were left and 29 were bilateral.
Ten participants had had previous carpal tunnel surgery
on one hand up to sixteen years before, but none of these
operated hands were still symptomatic.
For symptomatic hands the Boston Hand Score showed
a mean symptom severity score of 32.6 and a functional
severity score of 19.0 (Normal or asymptomatic scores
are 11 SSS and 8 FSS)
The neurophysiological testing produced a table of
results for both testing systems (Table 2).
There were seven hands where differences were seen
between the two systems. Four of these seven were in
asymptomatic hands. (Nos. 7, 10, 17 and 22). Two of
these had previously undergone carpal tunnel decom-
pressions. Traditional NCS picked up mild residual
Table 2. The severity of CTS in clinically symptomatic and
non-symptomatic hands using traditional and new device nerve
conduction velocity measurements.
Traditional NCS New Device
Normal 31 33
Abnormal 61 58
Mild 9 12
Moderate 43 31
Severe 9 15
Symptomatic
hands (92)
Non-diagnostic 0 1
Normal 14 18
Abnormal 20 16
Mild 11 11
Moderate 8 4
Non-symptomatic
hands (34)
Severe 1 1
Totals 126 126
sensory nerve conduction abnormalities in all four and a
mild motor nerve conduction abnormality in one. In
comparison, the new device recorded these asympto-
matic hands as normal.
These differences using the new device may be ex-
plained by a dilution effect from the longer distance the
finger stimulation method uses. In traditional NCS the
nerve is measured in addition to finger-to-wrist segment
over the palm-to-wrist segment. This is also the reason
we used a different classification in this study.
Non-significant mild deficits could be harder to detect
with the new device using only finger-to-wrist segment.
Another, albeit rare, explanation is the presence of a
communicating branch from the median side of the ring
finger in the palm to the ulnar nerve. Finally, low stimu-
lus intensity (2.5 times sensory threshold) is not always
supramaximal and might not evoke a response in all ax-
ons. The remaining three hands were symptomatic (Nos.
35, 47 and 50). They had mean symptom severity scores
of 32.3 and mean functional severity scores of 20.0. This
is not statistically significant from the rest of the symp-
tomatic group. Nos. 35 and 50 had however, only very
mild sensory conduction abnormalities. No. 35 also had
a mild motor conduction abnormality. Both cases were
negative using the new device and again, the reasons
suggested above may explain the different results. The
final case (No. 47) had a latency difference classified as
severe on traditional testing whereas there was no re-
sponse in either ring finger stimulation or forefinger
stimulation using the new device. Missing median and
ulnar nerve responses may be due to polyneuropathy or
because of simultaneous median and ulnar nerve lesion
and these patients should always be re-directed to tradi-
tional NCS.
When testing only symptomatic hands that were ab-
normal on traditional NCS, the new device correctly
found abnormalities in 58 out of 61 hands (positive per-
cent agreement 95%). No hands showed a false positive
result with the new device (negative per cent agreement
100%). In this study, the new device classified the sever-
ity of CTS of most of the patients into the same classes
as the traditional NCS (Table 2). There was a trend in
that the new device recorded more severe cases than the
traditional instrumentation. This was probably due to the
stimulus intensity level used in this series, i.e. two and a
half times a participant’s sensory threshold. This will not
always be supra maximal, i.e. does not necessarily
stimulate all nerve fibres. This could be overcome by
choosing higher initial stimulus intensities.
Figures 3(a)-(c) present the agreement of ring and
fore finger latency differences (4PM-4PU and 2P-4PU)
between the two systems. In Figure 3(b), there is one
outlier on the ring finger 4PM-4PU latency differences,
T. P. Green et al. / J. Biomedical Science and Engineering 4 (2011) 282-288
Copyright © 2011 SciRes. JBiSE
286
(a)
(b)
(c)
Figures 3. The graphs displays agreement levels be-
tween Keypoint and Mediracer fore and ring finger la-
tency differences. (a) Difference between fore (2P) and
ring finger (4PU) latencies between Keypoint and Medi-
racer. (b) Difference between ring finger double peak
(4PM-4PU) latencies showing one outlier and with out-
lier removed in. (c) Horizontal lines showing values are
drawn at the mean difference, and at the upper and lower
limits of agreement.
that shifts the upper agreement to a non-acceptable level.
In a 79 year old lady (No. 4) standard nerve conduction
measurements gave a double peak with 2.7 ms interpeak
interval and Mediracer measurement gave 0.7 ms inter-
peak conduction time. If this outlier is removed the
agreement levels are very acceptable as shown in Figure
3(c).
Figure 4 presents scatter plot of 4PM-4PU latency
differences in the two systems and linear regression fit
with and without the outlier, showing no important sys-
tematic difference. It is possible that this outlier was due
to an error in the traditional NCS measurement. When
using a felt pad stimulating electrode and rather low
stimulus intensity it is possible to evoke cathode stimu-
lation on the ulnar side and anode stimulation on median
side of the ring finger. This possibility is supported by
the fact that in index finger stimulation both the tradi-
tional and the new system produced similar latencies
(3.1 vs. 3.3 ms).
Pearson’s and intraclass correlation coefficients are
presented in Table 3. There was high correlation be-
tween the devices in the ring finger double peak and ring
and forefinger peak latency differences. For the absolute
latencies the correlation was not good as the new device
missed some median nerve responses that traditional
device did not. However, in spite of that latency dis-
crepancy, these few cases were neurophysiologically
classified as clearly abnormal in both methods: in the
traditional measurement moderately abnormal and in the
new device study severely abnormal.
Figure 4. Scatterplot of 4PM-4PU latency differences in the
two systems and linear regression fit with (solid line) and
without the outlier (dashed line). The estimate slope is 0.54,
intercept 0.49 and respective 95% confidence intervals are
(0.42, 0.66) and (0.34, 0.63). Without the outlier regression
slope and intercept estimates are 0.77 and 0.25, respective
95% confidence intervals (0.68, 0.87) and (0.13, 0.36).
T. P. Green et al. / J. Biomedical Science and Engineering 4 (2011) 282-288
Copyright © 2011 SciRes. JBiSE
287
Table 3. Interexaminer summary statistics and reliability results for mediracer and keypoint nerve conduction studies.
Traditional device New device Pearson Paired t-test Intraclass
Mean (SD) Range Mean (SD) Range correlation (P-value) correlation
Ring finger stimulation
4PU latency 2.5 (0.28) 2.0 - 3.1 2.6 (0.43) NR - 3.3 0.20 0.019 0.26
4PM latency 4.2 (1.1) NR - 8.0 3.9 (0.8) NR - 5.9 0.93 0.201 0.96
4PM-U latency 1.7 (1.1) 0.6 - 5.8 1.3 (0.7) 0.2 - 2.9 0.96 0.016 0.97
Forefinger stimulation
2P latency 3.8 (0.99) 2.5 - 5.7 2.9 (0.22) NR - 5.7 0.27 0.004 0.29
2P-4PU latency 1.3 (1.02) –0.2 - 3.5 1.1 (0.89) –0.2 - 3.30.96 0.004 0.98
NR = No response, latencies in ms.
4. DISCUSSION
The number of patients being referred with a diagnosis of
carpal tunnel syndrome is rising in the United Kingdom
and abroad [1,2,10]. It accounts for almost 10% of all
referrals to our elective Orthopaedic unit.
There seems little doubt that surgical treatment is effec-
tive, but there is still debate as to which patients need sur-
gery. A tendency towards milder cases presenting earlier
makes clinical diagnosis less reliable [2,10]. Any attempt
to monitor a clinically mild presentation of CTS by re-
peated tests has normally been precluded by the difficulty
and expense of accessing such tests. Even single diagnos-
tic NCS examinations may be subject to long delays.
Clinical protocols have been developed to reduce the
need for NCS. This is a pragmatic solution but may not
represent the best or most accurate assessment of the pa-
tient who presents with symptoms suggestive of CTS.
Introducing a new pathway in which the confidence en-
gendered by NCS can be added to clinical protocols, while
speeding and simplifying the process would be an ideal
solution. Moreover, since up to 25% of patients present at a
sufficiently late stage such that their chances of successful
treatment have already been prejudiced, early diagnosis is
important [2].
Where clinical signs are positive but NCS are negative,
it could be argued that easy access to repeat testing is im-
perative if a decision not to operate is made. Not all the
parameters of standard NCS are required or even under-
stood by many clinicians. Alterations of nerve conduction
speeds can be subtle and affected by many conditions, but
in carpal tunnel syndrome with good clinical signs, a sim-
ple yes/no answer as to whether there is a sensory median
nerve conduction delay between finger and wrist may act
as useful confirmation. Quantification of the severity of
such a conduction delay can help guide treatment options.
Pure motor involvement of the median nerve is rare –0.3%
- 1% [7,11].
The new device under test uses peak latency difference
between the median and ulnar nerves as the main meas-
urement parameter. This was chosen after an earlier ob-
servation showed that latency difference is more sensitive
than SCV in the diagnosis of median nerve lesion in CTS
[12]. Moreover, the measurement of amplitudes seems not
to increase the number of abnormalities detected by using
only latency measurement [13]. For this reason and for
making the method as simple as possible to use, the stan-
dard anatomical landmarks and not the standard distances
were chosen for use in the new device NCS.
In 9% of all hands, in 5% of symptomatic hands, the new
device missed the abnormality detected by the traditional
examination. Six of these seven hands showed a mild or
very mild abnormality and one a severe abnormality. In
this last case the new device did not find any responses
either in median nerve or ulnar nerve measurements and a
pure median nerve lesion diagnosis could not be made.
This loss of both signals is easy to recognize and suggests
referral for traditional NCS. CTS with a mild or absent
median nerve lesion is usually treated conservatively, and
so false negative results will not lead to incorrect treatment.
In the severity classification of CTS the new device
slightly over exaggerated severe cases. This was possibly
due to submaximal stimulation intensities in some meas-
urements.
The small number of incorrect results from the new de-
vice study suggested some ways to improve it. The risk of
using too weak a stimulus current has been corrected in the
latest version. The responses are now always visually
checked in real time on the computer screen during the
stimulation process in order to ensure supramaximal
stimulus. As well as that, in order to confirm the ring finger
ulnar nerve response measured in this study, the little finger
response should also always be measured. The two digital
nerves in this finger give a greater axonal volume and thus
a more accurate and sensitive ulnar response compared
with that in the ring finger. This way, possibly more low
responses could be recorded by the new device, reducing
the number of the lost responses and thereby also the slight
T. P. Green et al. / J. Biomedical Science and Engineering 4 (2011) 282-288
Copyright © 2011 SciRes. JBiSE
288
over exaggeration of severe lesions in the present study.
Comparison of the two methods of NCS testing shows a
high degree of agreement between the systems. When
coupled with clinical tests and a symptom and functional
severity score, the diagnosis and quantification of severity
of CTS can be supported with a high degree of confidence
using the new device However, when only sensory nerve
conduction (SNC) measurements are made, and a limited
number of nerves are studied, the present new method
should only apply to patients with a CTS as in the present
study or in a high suspicion of that. In these cases the
probable median nerve lesion and its severity can be de-
fined, supplementing the clinical diagnosis of CTS and
thus helping the choice of conservative or operative ther-
apy. In cases with clinical diagnostic dilemma, traditional
neurophysiology should be primarily consulted.
The present study showed that automatic cursor place-
ments of the new device study missed in several cases. For
this reason the cursor setting must always be checked
visually as would also happen in traditional NCS. When
using the new device in primary or non-specialist practice
the interpretation of the data can be obtained via the
internet from a specialist in clinical neurophysiology and
this would be our recommendation.
The simplicity of the hand-held system and the possi-
bility of it being used by a non-specialist suggest the defi-
nite possibility of having a “one-stop” clinic for the vali-
dated diagnosis of the condition of carpal tunnel syndrome.
The ability to refer results to a specialist in clinical neu-
rophysiology via a web-based system for a report increases
the usefulness of the new system. The new system has
obvious potential to make it easier to obtain repeated
measurements in order to assess progress of the disease or
response of treatment.
5. ACKNOWLEDGEMENTS
The authors would like to acknowledge the contributions of Mr ML
Newey, Mr CJ Kershaw, Mrs A Clarke, Mrs P Rouse and Mrs P Red-
wood who all took part in carrying out the study. Mrs Hanna Heikki-
nen MSc contributed to the statistical part of the study.
REFERENCES
[1] Atroshi, I., Gummeson, C., Johnsson, R., Ornstein, E., Ran-
stam, J. and Rosen, I. (1999) Prevalence of carpal tunnel
syndrome in a general population. Journal of the Ameri-
can Medical Association, 282, 153-158.
[2] Bland, J.D.P. and Rudolfer, S.M. (2003) Clinical surveil-
lance of carpal tunnel syndrome in two areas of the
United Kingdom, 1991-2001. Journal of Neurol Neuro-
surg Psychiatry, 74, 1674-1679.
doi:10.1136/jnnp.74.12.1674
[3] American Academy of Neurology and the American
Academy of Physical Medicine and Rehabilitation. Prac-
tice parameter (2002) Electrodiagnostic studies in carpal
tunnel syndrome. Report of the American Association of
Electrodiagnostic Medicine. Neurology, 58, 1589-1592.
[4] Padua, L., LoMonaco, M., Gregori, B., Valente, E.M.,
Padua, R. and Tonali, P. (1997) Neurophysiological clas-
sification and sensitivity in 500 carpal tunnel syndrome
hands. Acta Neurologica Scandinavica, 96, 211-217.
doi:10.1111/j.1600-0404.1997.tb00271.x
[5] Department of Health (2007) Transforming Clinical Neu-
rophysiology Diagnostic Services to deliver 18 Weeks. A
Good Practice Guide.
[6] Uncini, A., Di Muzio, A., Awad, J., Manente, G., Tafuro,
M. and Gambi, D. (1993) Sensitivity of three median-
to-ulnar comparative tests in diagnosis of mild carpal
tunnel syndrome. Muscle Nerve, 16, 1366-1373.
doi:10.1002/mus.880161215
[7] Tolonen, U., Kallio, M., Ryhänen, J., Raatikainen, T.,
Honkala, V. and Lesonen, V. (2007) A handheld nerve
conduction measuring device in carpal tunnel syndrome.
Acta Neurologica Scandinavica, 115, 390-397.
[8] Levine, D.W., Simmons, B.P., Koris, M.J., et al. (1993) A
self-administered questionnaire for the assessment of se-
verity of symptoms and functional status in carpal tunnel
syndrome. Journal of Bone and Joint Surgery, 75, 1585-
1592.
[9] Bland, J.M. and Altman, D.G. (1986) Statistical methods
for assessing agreement between two methods of clinical
measurement. Lancet, 1, 307-310.
[10] Hobby, J.L. and Dias, J.J. (2006) A review of hand sur-
gery provision in England. Journal of Hand Surgery, 2,
230-235.
[11] Repaci, M., Torrieri, F., Di Blasio, F. and Uncini, A.
(1999) Exclusive electrophysiological motor involve-
ment in carpal tunnel syndrome. Clinical Neurophysiol-
ogy, 110, 1471-1474.
doi:10.1016/S1388-2457(99)00071-1
[12] Chang, M.H., Liu, L.H., Lee, Y.C., Wei, S.J., Chiang,
H.L. and Hsieh, P.F. (2006) Comparison of sensitivity of
transcarpal median motor conduction velocity and con-
ventional conduction techniques in electro diagnosis of
carpal tunnel syndrome. Clinical Neurophysiology, 117,
984-991. doi:10.1016/j.clinph.2006.01.015
[13] Eisen, A., Schulzer, M., Pant, B., MacNeil, M., Stewart,
H., Trueman, S. and Mak, E. (1993) Receiver operating
characteristic curve analysis in the prediction of carpal
tunnel syndrome: A model for reporting electrophysio-
logical data. Muscle Nerve, 16, 787-796.
doi:10.1002/mus.880160715