Journal of Cancer Therapy
Vol.08 No.13(2017), Article ID:81031,9 pages
10.4236/jct.2017.813099

Measuring Radiotherapy Setup Errors in IMRT Treated Head and Neck Cancer Patients Requiring Bilateral Neck Irradiation, NCI-Egypt Experience

May G. Ashour1*, Tarek H. Shouman1, Ashraf H. Hassouna1, Reem M. Emad El Din1, Maha H. Mokhtar2, Mohamed A. Khalil2

1Radiation Oncology Department, National Cancer Institute, Cairo University, Cairo, Egypt

2Physics Unit, Radiation Oncology Department, National Cancer Institute, Cairo University, Cairo, Egypt

Copyright © 2017 by authors and Scientific Research Publishing Inc.

This work is licensed under the Creative Commons Attribution International License (CC BY 4.0).

http://creativecommons.org/licenses/by/4.0/

Received: November 9, 2017; Accepted: December 10, 2017; Published: December 13, 2017

ABSTRACT

Objective: We aim to quantify the magnitude of setup errors in intensity- modulated radiotherapy (IMRT) treated Head and Neck cancer patients and recommend appropriate PTV margin. Methods: 60 patients with head and neck cancer required bilateral neck irradiation were planned and treated by simultaneous integrated boost IMRT technique either treated radically or postoperative. Patients undergoing image-guided radiotherapy (IGRT) each with once weekly scheduled cone beam computed tomography (CBCT). The 3D displacements, systematic and random errors were calculated. The appropriate PTV expansion was determined using Van Herk’s formula. Results: Mean 3D displacement was 0.16 cm in the vertical direction, 0.14 cm in the horizontal direction and 0.16 cm in the longitudinal direction. Conclusion: Use of weekly CBCT allows the planning target volume (PTV) expansion to be reduced according to our setup. The appropriate clinical target volume (CTV)-PTV margin for our institute is 0.30 cm, 0.38 cm, and 0.33 cm in the horizontal, vertical, and longitudinal directions, respectively.

Keywords:

IMRT, CBCT, Setup Errors, Head and Neck

1. Introduction

In the last decade, there is a progress in the treatment of head and neck cancer (HNC). More aggressive treatment regimens, either delivery of radiotherapy with concomitant chemotherapy or altered fractionation schedules, improve tumour control and survival [1] , but at the expense of increased acute and late effects, which may have a more profound effect on function and quality of life (QOL). HNC arises in structurally complex and functionally important areas. Impairment of these areas from both disease and therapy can interfere with basic functions [2] .

This highlights the importance of improving existing radiotherapy techniques to reduce the dose in relevant structures as much as possible with the use of new computer assisted optimization methods, such as intensity-modulated radiotherapy (IMRT) treatment plans with highly conformal dose distributions can be obtained [3] .

IMRT is a highly conformal technique that allows the dose to be shaped around normal structures but full therapeutic doses to be delivered to the tumor and regions at high risk for disease [4] .

So it is important to take into consideration setup errors during patient positioning which can be defined as a discrepancy between the anatomy of the patient at the planning CT and during treatment. These errors can be divided into systematic errors (which are reproducible consistent errors, occurring in the same direction and magnitude) and random errors (which vary in direction and magnitude). The systematic errors cause a shift of the cumulative dose distribution [5] .

The rationale for the PTV margin is to minimize the effects of geometric (systematic errors) and residual (random errors) uncertainties [6] .

2. Patients and Methods

2.1. Selection

Sixty patients with Head and Neck squamous cell carcinoma who presented to the National Cancer Institute (NCI), Cairo University, from September 2015 till October 2016 with 3D imaging verification via CBCT were enrolled. Patients and tumour characteristics are shown in Table 1.

2.2. CT Simulation

All patients were immobilized using a thermoplastic mask with 5-point fixation and had contrasted CT scan from the vertex to mediastinum with 2.5 mm slice thickness with the patient placed on a neck rest that provided full neck extension.

2.3. Treatment Planning

Patients received 70 Gy/33 fx to the PTV1, 60 Gy/33 fx to PTV2 (high risk CTV), 54 Gy/33 fx to PTV3 (low risk CTV) in case of radical treatment. In the post- operative setting, two volumes were identified: CTV1 including the tumor bed and high-risk nodal areas and CTV2 including elective lymphatic areas. These volumes were irradiated to a total dose of 60 Gy/30 fx and 54 Gy/30 Fx,

Table 1. Patients’ characteristics.

respectively using the simultaneous integrated boost (SIB) technique. All patients were optimized using a step and shoot technique (Monaco Eleckta software planning system, Stockholm, Sweden), using Monte Carlo algorithm.

2.4. Daily Setup and Image Guidance

IGRT using either kilo-voltage cone beam CT (KV-CBCT) and/or electronic portal imaging device (EPID). Images were done weekly for all patients as shown in Figure 1.

Matching is done between the online image and the planning CT image in three directions; vertical, horizontal and longitudinal. The vertical direction accounts for the anterior-posterior error, the horizontal direction accounts for the change of position along the right to left axis while the longitudinal direction accounts for the change along the superior-inferior direction.

2.5. Data Collection

Study data were collected and managed using REDCap research electronic data capture tools V6.10.12 (Radiation Oncology Record, Vanderbelt University, Nashville, TN) and statistical analysis was done using SPSS v23 (IBM inc., Chicago, IL).

3. Results

Sixty patients were included, 18 patients (30%) had a primary cancer in the larynx, 16 patients (26.7%) in the nasopharynx, 11 patients (18.3%) in the oral cavity, 10 patients (16.7%) in the hypopharynx, 3 patients (5%) in the oropharynx and 2 patients (3.3%) sinonasal. 54 patients ended radiotherapy treatment and 6 didn’t continue treatment (2 patients developed Grade 3 mucositis and refused to continue, 1 patient developed diabetic coma and died, 2 patients

Figure 1. Matching between digitally reconstructed image (DRR) and image taken prior to treatment.

Table 2. Set up errors (in cm) using online verification.

Table 3. Difference in set up errors according to tumor location.

Table 4. Systematic and random error and recommended CTV-PTV margin in the present study (53 cases).

didn’t continue treatment due to social reasons, 1 patient had hypopharyngeal carcinoma rapidly deteriorating and didn’t continue treatment).

Most of the cases had stage III disease (28 patients, 46.7%) and stage IV disease (22 patients, 36.7%) while only 3 patients (5%) had stage I and 7 cases (11.7%) had stage II disease. Patient’s characteristics and treatment demographics are shown in Table 1.

All patients had online imaging using kilo-voltage cone beam CT (KV-CBCT) and/or electronic portal imaging device (EPID). Out of the 60 patients, only 53 patients had 3 consecutive weekly online verifications and the rest had EPID imaging due to machine break down. EPID was used only when the CBCT was out of function.

The mean of the detected error was 0.16 cm in the vertical direction, 0.14 in the horizontal direction and 0.16 in the longitudinal direction.

The overall mean error (M), systematic error (S) and random error (σ) were determined according to van Herk’s formalism and the PTV margins were calculated using the van Herk formula [5] . Our work showed PTV margin 3.0, 3.8, and 3.3 mm for x, y, and z respectively.

4. Discussion

This paper reports our clinical experience with IGRT in the treatment of H&N cancer, with the aim to define the overall accuracy of our setup procedures and to define the proper CTV-PTV margins to be adopted in our target delineation.

Assessments of set-up variations are essential on an institute-by-institute basis as the efficacy of radiotherapy treatment depends on patient’s setup, as they are dependent on factors such as immobilization devices and the clinical experience of the staff [7] .

Furthermore, the inadequate definition of PTV can lead to local failure due to geographical tumour misses, or to increase the dose to the normal tissues near the CTV.

The assessment of set up errors in the present study was done using CBCT and confirmed its effectiveness in reducing setup margins. The mean of setup errors was bigger in radical cases than in postoperative cases that may be due to shrinkage of target during treatment (Table 2). Previous studies by Li et al. [8] addressed setup uncertainties for H & N treatment sites and described a number of common factors that can contribute to setup errors for these sites. The factors consist of loosening of the fixation mask due to a reduced body contour following weight loss, or tumor shrinkage and tightening of the mask by the swelling of some part of the lesion.

In consideration of these factors, the fixation mask was remade in our institution if considerable discrepancies occurred, and rescanning and replanning were performed to reduce setup errors for these sites.

We found that the mean setup error in the horizontal direction was high in hypopharyngeal cases, and in the vertical direction for oropharyngeal cases and

in the longitudinal direction for the sinonasal cases (Table 3).

According to some authors, in HN tumors organ motion may be neglected. In fact, Hamlet et al. reported that in larynx tumors breathing and swallowing have no significant impacts on dose distribution [9] .

Also, Van Asselen et al. reported that there was no need to adjust the margins around the CTV to take into account the larynx displacements occurred due to swallowing [10] .

Our work showed PTV margin of 3.0, 3.8, and 3.3 mm for X, Y, and Z, respectively (Table 4), while these margins were 4.9, 6.4, and 5.8 mm in Mesías et al. [11] , 5.7, 5.3, and 6.2 mm in Saha et al. [12] , 2.3, 1.9, and 2.0 mm in Liu et al. [13] , and 3.0, 1.3, and 2.6 mm in Su et al. [14] .

The minimum reported PTV margin is 1.9 mm. The PTV margin that is suggested in our present work falls within the reported margins. Nevertheless, each margin is institution specific and small differences are to be expected even though the methodology is the same. However, in the case of close proximity of OAR and high-dose regions, or when re-irradiation has to be performed, daily image guidance a reduced PTV margin should be strongly considered.

These consequences are reflected in some dosimetric studies which reported that PTV margin need to be larger to account for setup errors in case that daily IGRT is not used [15] [16] and that the improved setup accuracy achieved using IGRT can protect healthy tissues to a great extent [17] .

5. Conclusion

Set-up variations should be obtained in each department to calculate institute- specific margins, as the planning margins reported in the literature are only indicative and range from 3 - 5 mm. Verification protocol of bony and soft tissue anatomical references information provided by CBCT improves the set-up accuracy and allows full radiation dose delivery to the target with decreasing the radiation dose to normal critical structures. According to our results, the appropriate CTV-PTV margin for our institute is 0.30 cm, 0.38 cm, and 0.33 cm in the horizontal, vertical, and longitudinal directions, respectively.

Conflicts of Interest

None of the authors has any conflict of interest to declare.

Cite this paper

Ashour, M.G., Shouman, T.H., Hassouna, A.H., El Din, R.M.E., Mokhtar, M.H. and Khalil, M.A. (2017) Measuring Radiotherapy Setup Errors in IMRT Treated Head and Neck Cancer Patients Requiring Bilateral Neck Irradiation, NCI-Egypt Experience. Journal of Cancer Therapy, 8, 1160-1168. https://doi.org/10.4236/jct.2017.813099

References

  1. 1. De Ruyck, K., Duprez, F., Werbrouck, J. and Young, W. (2013) A Predictive Model for Dysphagia Following IMRT for Head and Neck Cancer: Introduction of the EMLasso Technique. Radiotherapy Oncology, 107, 295-299. https://doi.org/10.1016/j.radonc.2013.03.021

  2. 2. Murphy, B. (2009) Late Treatment Effects: Reframing the Questions. Lancet Oncology, 10, 530-531. https://doi.org/10.1016/S1470-2045(09)70118-4

  3. 3. Van Der Laan, H., Christianen, M., Bijl, H., Shilstra, C. and Langendijk, J.A. (2012) The Potential Benefit of Swallowing Sparing Intensity-Modulated Radiotherapy to Reduce Swallowing Dysfunction: An in Silico Planning Comparative Study. Radiotherapy Oncology, 103, 76-81. https://doi.org/10.1016/j.radonc.2011.11.001

  4. 4. Werbrouck, J., De Ruyck, K., Duprez, F., Veldeman, L., Claes, K., et al. (2009) Acute Normal Tissue Reactions in Head and Neck Cancer Patients Treated with IMRT: Influence of Dose and Association with Genetic Polymorphisms in DNA DSB Repair Genes. International Journal of Radiation Oncology Biology Physics, 73, 1187-1195. https://doi.org/10.1016/j.ijrobp.2008.08.073

  5. 5. Van Herk, M. (2004) Errors and Margins in Radiotherapy. Seminar Radiation Oncology, 14, 52-64. https://doi.org/10.1053/j.semradonc.2003.10.003

  6. 6. ICRU (1993) Prescribing, Recording and Reporting Photon Beam Therapy ICRU Report No 50. ICRU, Bethesda, MD.

  7. 7. Cacicedo, J., Perez, J., Ortiz de Zarate, R., Del Hoyo, O., Casquero, F., et al. (2015) A Prospective Analysis of Inter- and Intrafractional Errors to Calculate CTV to PTV Margins in Head and Neck Patients. Clinical and Translational Oncology, 17, 113-120. https://doi.org/10.1007/s12094-014-1200-z

  8. 8. Li, H., Zhu, X.R., Zhang, L., Dong, L., Tung, S., et al. (2008) Comparison of 2D Radiographic Images and 3D Cone Beam Computed Tomography for Positioning Head-and-Neck Radiotherapy Patients. International Journal of Radiation Oncology Biology Physics, 71, 916-925. https://doi.org/10.1016/j.ijrobp.2008.01.008

  9. 9. Hamlet, S., Ezzel, G. and Aref, A. (1994) Larynx Motion Associated with Swallowing during Radiation Therapy. International Journal of Radiation Oncology Biology Physics, 28, 467-470. https://doi.org/10.1016/0360-3016(94)90073-6

  10. 10. Van Asselen, B., Raaijmakers, C.P., Lagendijk, J.J. and Terhaard, C.H. (2003) Intrafraction Motions of the Larynx during Radiation Therapy. International Journal of Radiation Oncology Biology Physics, 56, 384-390. https://doi.org/10.1016/S0360-3016(02)04572-8

  11. 11. Mesias, M.C., Boda-Heggemann, J., Thoelking, J., Lohr, F., Wenz, F., et al. (2016) Quantification and Assessment of Interfraction Setup Errors Based on Cone Beam CT and Determination of Safety Margins for Radiotherapy. PloS One, 11, e0150326. https://doi.org/10.1371/journal.pone.0150326

  12. 12. Saha, A., Mallick, I., Das, P., Shrimali, R.K., Achari, R., et al. (2016) Evaluating the Need for Daily Image Guidance in Head and Neck Cancers Treated with Helical Tomotherapy: A Retrospective Analysis of a Large Number of Daily Imaging-Based Corrections. Clinical Oncology, 28, 178-184. https://doi.org/10.1016/j.clon.2015.11.014

  13. 13. Liu, G., Zhang, S., Ma, Y., Wang, Q., Chen, X., et al. (2016) Effects of Error on Dose of Target Region and Organs at Risk in Treating Nasopharynx Cancer with Intensity Modulated Radiation Therapy. Pakistan Journal of Medical Sciences, 32, 95.

  14. 14. Su, J., Chen, W., Yang, H., Hong, J., Zhang, Z., et al. (2015) Different Setup Errors Assessed by Weekly Cone-Beam Computed Tomography on Different Registration in Nasopharyngeal Carcinoma Treated with Intensity-Modulated Radiation Therapy. Oncology Targets Therapy, 8, 2545.

  15. 15. Yin, W.J., Sun, Y., Chi, F., Fang, J.L., Guo, R., et al. (2013) Evaluation of Inter-Fraction and Intra-Fraction Errors during Volumetric Modulated Arc Therapy in Nasopharyngeal Carcinoma Patients. Radiation Oncology, 8, 78. https://doi.org/10.1186/1748-717X-8-78

  16. 16. Djordjevic, M., Sjoholm, E., Tullgren, O. and Sorcini, B. (2014) Assessment of Residual Setup Errors for Anatomical Sub-Structures in Image-Guided Head-and-Neck Cancer Radiotherapy. Acta Oncologica, 53, 646-653. https://doi.org/10.3109/0284186X.2013.862593

  17. 17. Nguyen, N.P., Ceizyk, M., Vos, P., Vinh-Hung, V., Davis, R., et al. (2010) Effectiveness of Image-Guided Radiotherapy for Laryngeal Sparing in Head and Neck Cancer. Oral Oncology, 46, 283-286. https://doi.org/10.1016/j.oraloncology.2010.01.010