Applied Mathematics, 2011, 2, 1497-1506 doi:10.4236/am.2011.212212 Published Online December 2011 (http://www.SciRP.org/journal/am) Copyright © 2011 SciRes. AM Precision of a Parabolic Optimum Calculated from Noisy Biological Data, and Implications for Quantitative Optimization of Biventricular Pacemakers (Cardiac Resynchronization Therapy) Darrel P. Francis International Centre for Circulatory Health, National Heart and Lung Institute, Imperial College London, London, United Kingdom E-mail: d.francis@imperial.ac.uk Received August 26, 2011; revised November 9, 2011; accepted November 17, 2011 Abstract In patients with heart failure and disordered intracardiac conduction of activation, doctors implant a biven- tricular pacemaker (“cardiac resynchronization therapy”, CRT) to allow adjustment of the relative timings of activation of parts of the heart. The process of selecting the pacemaker timings that maximize cardiac func- tion is called “optimization”. Although optimization—more than any other clinical assessment—needs to be precise, it is not yet conventional to report the standard error of the optimum alongside its value in clinical practice, nor even in research, because no method is available to calculate precision from one optimization dataset. Moreover, as long as the determinants of precision remain unknown, they will remain unconsidered, preventing candidate haemodynamic variables from being screened for suitability for use in optimization. This manuscript derives algebraically a clinically-applicable method to calculate the precision of the opti- mum value of x arising from fitting noisy biological measurements of y (such as blood flow or pressure) ob- tained at a series of known values of x (such as atrioventricular or interventricular delay) to a quadratic curve. A formula for uncertainty in the optimum value of x is obtained, in terms of the amount of scatter (irrepro- ducibility) of y, the intensity of its curvature with respect to x, the width of the range and number of values of x tested, the number of replicate measurements made at each value of x, and the position of the optimum within the tested range. The ratio of scatter to curvature is found to be the overwhelming practical determi- nant of precision of the optimum. The new formulae have three uses. First, they are a basic science for any- one desiring time-efficient, reliable optimization protocols. Second, asking for the precision of every re- ported optimum may expose optimization methods whose precision is unacceptable. Third, evaluating preci- sion quantitatively will help clinicians decide whether an apparent change in optimum between successive visits is real and not just noise. Keywords: Cardiac, Pacemaker, Parabolic, Haemodynamics 1. Introduction Plato, Apology of Socrates, 38a Every year, ~100,000 cardiac resynchronization pace- makers are being implanted into patients with heart fail- ure, because they deliver substantial symptomatic and survival benefits [1-3] by altering intracardiac timings. After implantation, the process of determining which ti- mings to programme is described as “optimization” [4-6]. Responses of physiological variables to changes in pace- maker settings fit well to a parabola in the vicinity of the optimum [6,7]. Curve-fitting to calculate a clinical opti- mum has the advantage of permitting interpolation to settings which were not directly tested, and also avoids the problem that simply “picking the highest” leads to illusory optima, illusory increments in physiology from optimization, and illusory changes in optimum over time [8]. In clinical practice physiological measurements con-
D. P. FRANCIS 1498 tain noise which can sometimes be substantial in com- parison to the underlying signal, and which prevents the true underlying optimum being identified precisely even with curve fitting (Figure 1). The impact of such noise can be reduced by averaging multiple measurements, but this consumes resources such as time in a clinical envi- ronment or battery power if the measurements are con- ducted by the implanted device. It is therefore important to be able to calculate the precision of an optimum so that resource usage can be planned to be appropriate to achieve the clinically-required precision. As well as needing to know how many replicates are required, doctors planning an optimization protocol also need guidance regarding what range of settings to cover during testing, and how coarsely or finely. Widening the range or making finer-grained measurements (i.e. at more closely-spaced intervals) increase the cost of the optimization process, and so are only justified if there is a clinically-valuable increase in precision of the opti- mum. Finally, doctors caring for patients from day to day need to know the uncertainty of the optimum obtained this clinical optimization process. Without this know- ledge, it is impossible to interpret apparent differences in optimum within an individual patient between one as- sessment method and another, or over time [8,9] or after an operation or a heart attack. This paper derives a simple formula for the uncer- tainty of the parabolically-defined optimal setting of a pacemaker, and presents practical implications for pro- tocol design and for medical practice. 2. Description of Method 2.1. Physiological Measurements with Noise The pacemaker setting may be adjusted over a wide ran- ge of values. Within individual patients, clinical infor- mation provides a priori a constrained range which con- tains all biologically plausible locations of the optimum for that patient. During the optimization procedure, the clinician acquires a series of pairs of values of the pace- maker setting and the corresponding physiological mea- surement. The first value, the pacemaker setting, has negligible uncertainty because it is programmed digitally. However the second value, the measured physiological variable, is not perfectly reproducible and therefore has an element of uncertainty. Some choices of physiological variable, such as blood pressure, permit automatic acquisition, while others, such as Doppler velocity-time integral, typically require hu- man involvement for each measurement. In practice the protocol is often to make more than one replicate mea- surement at each setting, and then to summarize the data 80100 120 140 160 90 100 110 1 111 12 2 2 2 2 3 3 3 3 3 1 111 1 2 2 2 2 2 3 333 3 Optimization #1 Optimization #2 Optimization #3 Three optimization datas ets Three clinical optima Uncer tainty in location of optimum AV delay (ms) Physiological response (arbitrary units) 80100 120 140 160 90 100 110 1 111 12 2 2 2 2 3 3 3 3 3 1 111 1 2 2 2 2 2 3 333 3 Optimization #1 Optimization #2 Optimization #3 Three optimization datas ets Three clinical optima Uncer tainty in location of optimum AV delay (ms) Physiological response (arbitrary units) Figure 1. Sketch showing how noise in the measured data creates uncertainty in the optimum. Note: If there is enough time to conduct many optimizations, they can be analyzed separately to provide separ ate estimates of the optimum, so the uncertainty in the optimum can be observed. The upper panel sketches 3 optimization datasets in one simulated patient. All 3 datasets are plotted on the same graph and labeled “1” to “3”. The lower panel shows the 3 individual interpolated optima obtained by curve fitting with a parab- ola. This paper provides a method of calculating the uncer- tainty in a clinically-obtained optimum, without having to conduct several indepe nde nt optimizations. for each setting by the mean of the raw measurements at that setting. Let the optimization protocol try S different values xi of the pacemaker setting, evenly spaced at intervals of x. Let the averaged physiological measurements represent- ing each setting be denoted yi, each of which is the mean of R replicate raw measurements. The observed data yi during clinical testing are composed of an underlying value yund,i and an error component εR. ,und R ii yy Let us consider the error component to be independent Copyright © 2011 SciRes. AM
D. P. FRANCIS1499 of yund,i and normally distributed, with its standard devia- tion for a single raw measurement being 1 and, by the central limit theorem, the standard deviation for the error εR of the average of R replicates being 1R R . 2.2. The Optimization Process The best-fit parabola, Axi 2 + Bxi + C, to the observed data is defined by the standard least-squares method, to minimize the squared deviation between the observations yi and the parabola. The clinical optimum from that fitted parabola is –B/2A, which will differ from the true opti- mum OptTrue by an amount whose standard deviation can be calculated as follows. The best-fit parabola is defined by having A, B and C values that minimize the squared error 2 2 ii i AxBxC y . This is achieved by setting all the partial derivatives of F with respect to A, B and C to zero: 22 2 2 20 20 20 ii ii ii i ii i FAAxBxCyx FBAxBxCyx i F CAxBxCy 2 ii ii i Therefore 432 32 21 iii iii ii xBxCx xy xBxCxxy xBxCy These can be solved for A, B and C as follows: 22 2 42 42 22 42 , iii i ii iiiiiii iii Sxyx y A Sx x 2 yxyx BC xSx x xy 3. Results 3.1. The Clinical Optimum The clinician will choose as optimal the pacemaker set- ting which corresponds to the middle of the fitted parab- ola, i.e. ˆ2 opt B x . In terms of the clinical data xi and yi, this clinical optimum is 2 24 22 2 ˆ22 ii opt iii i xSx xSx xyxy For simplicity, and without loss of generality, a coor- dinate system for x can be chosen that makes 0 the centre of the range of settings tested during optimization. This symmetrical arrangement of xi values provides the fol- lowing convenient identities: 22 344 2 0,11 12, 0,1 371 240 ii ii xxxSSS xxxSSSS Applying these substitutions permits the clinical opti- mum to be described as follows: 22 22 2 24 ˆ 51 60 ii opt ii Sxxy xSxy x i y . (1) If yi is augmented by any constant k, the numerator is unchanged because iiiii ii ykxykx xy . The denomi- nator is also unchanged because it is augmented by 22 2 51 60 i Sxkkx which is 22 22 60 1 51 12 kxSS kS Sx 0 . Therefore without loss of generality the observed val- ues yi may be defined in terms of the underlying quad- ratic curvature coefficient Aund, the true optimum xopt, and the noise component εR as follows: 2 ,iundioptR i yA xx 2 , 22 , 2 22 , 2 1 12 iundioptRi undioptioptR i undundoptR i yAxx AxxxSx SS Ax ASx 2 , 322 , 2 2 , 2 1 6 iiiundioptRi undiopt iopt iiRi optundiR i xyxAxx Axxxxxx SS xA xx 2 22 , 43222 , 22 2 422 2 , 2 13 71 240 12 iiiundioptR i undiopt ioptiiRi und opt iRi xyx Axx Axxxxx x SS SSS Ax xx x 3.2. Expression for the Clinical Optimum i . Applying these within Equation (1) gives: Copyright © 2011 SciRes. AM
D. P. FRANCIS Copyright © 2011 SciRes. AM 1500 2 22 2 , 222 22 224222 , , 2 222 , ˆ 1 24 6 1137 51 60 12240 12 1 24 212 opt optundiRi undundoptR iundoptiR i optundiRi x SS SxxAx x SSSS SSS SxAx ASxAxxxx SS SxxAxx 2 1 222 22 242 ,, 22 422 , 222222 44 2 ,, 1137 5 16060 12 240 14 24 3 51 11375160 12 4 undR iundiR i undoptiR i Rii Ri und und SSSS S S xAxAxx SS S Ax xSxx SSSSSSS x xx x AA 222 2 4 , 222 2 4 2 ,, , 22 2 ,, 2242 2 1424 3 ˆ 1451 60 3 6 1 15 180 141 optiR i und opt Rii Ri und und optiRi und 4 ii und und SS SSx xx x A xSS SSx xx AA xx AxSS x A xSSA xSSS Ri 3.3. Imprecision of the Clinical Optimum It can be seen from this that the clinically observed opti- mum ˆopt differs from the underlying optimum opt because of two types of error, which affect the numerator and denominator respectively. The component in the numerator is a simple additive error. The impact of the denominator error, however, scales with opt . In clinical situations where the errors are large in com- parison with the degree of curvature that is manifested, the entire denominator of this formula falls near (or be- yond) zero and behavior of the expression becomes strongly nonlinear. In such a situation there is almost no useful information about the underlying optimum avai- lable from the acquired data. Clinically this can be con- sidered likely whenever the information content (or in- traclass correlation coefficient) is low [8]. In most situations of well-designed optimization pro- tocols, however, the error is not large compared with the curvature of the signal, which makes it possible to pro- vide a closed-form expression for the imprecision of the optimum. To do this, the variance of the numerator is first shown to be , 22 1 2 222 22 222 1 2 6 1 63 11 R iRi und S S und und i Var x AxSS i AxSSA xSS Likewise the variance of the denominator is , 22 2 , 42 2 22 , 22 2 2 24 22 15 4 180 14 151 12 14 180 14 R Ri und iRi und Ri und und Var AxSS x AxSS S Si Var AxSS S AxSS S
D. P. FRANCIS1501 By the binomial expansion, 1/(1−z) = 1 + z + z2 + z3 which can be approximated with 1 + z as long as the magnitude of z is well below 1. Thus as long as we can make the following assumption: 24 222 14180 und AxSS S (2), and as long as the numerator and denominator errors are not large, a linear approximation can be used for the va- riance of ˆopt as follows: 2 22 2 2 24 22 22 2 22 22 3 ˆ 1 180 14 34180 14 R 2 R opt und opt und opt und Var xAxSS xAxSS S Sx Ax SS S The standard error of the clinical optimum, being the square root of this, is therefore 2 2 2 60 31 4 ˆ 1 R opt opt und x x S SE xAx SS . 3.4. Contributory Factors to Imprecision of the Clinical Optimum In general the optimization protocol may have multiple replicates, i.e. R 1, in order to reduce the effect of noise. 1R so that in terms of the fundamental bio- logical characteristics, the standard error of the clinical optimum is R 1 2 2 2 ˆ 31 60 14 1 opt opt und SE x x x S RS xS (3) This formulation highlights the 4 contributory factors to imprecision of the optimum clearly. First, the impreci- sion of the clinical optimum falls with the square root of , the number of individual physiological measure- ments made. Since making additional measurements ei- ther consumes scarce time in a clinical environment, or battery power if conducted automatically by the pace- maker, having knowledge of this tradeoff between num- ber of measurements and imprecision may be helpful. SR Second, by observing that the second term is almost the reciprocal of the width of the range of settings tested, , it can be seen that the wider the range tested, the more precise the optimum. In practice this is limited by loss of fit to the parabola for settings far from the op- timum. However, within the range over which behaviour is parabolic, this analysis suggests it is desirable to cover a wide range rather than to focus exclusively on the very close vicinity of the optimum. 1S Third, biology sets an lower limit on 1 und . While measurement error can be reduced by choosing meas- urement techniques with smaller instrument noise, even- tually almost all the variability between replicates is genuine biological variation between heart beats, which places a lower boundary on 1 . Meanwhile Aund is a manifestation of dependence of physiology upon changes in pacemaker setting, and is determined by the patient’s own biological characteristics. For ideal measurement modalities, where the equipment contributes no noise, all the variability is biological, and may well be similar across different modalities. For example, measures of pressure and flow, although fundamentally different and having distinct units, may change proportionally with the same constant of proportionality in response to both sig- nal (change in pacemaker setting) and noise (spontane- ous biological variability). Thus 1 und may have a bio- logically-imposed lower limit within an individual pa- tient which cannot be improved upon by the clinician. The final term can be neglected if the optimum is very close to the centre of the tested range, i.e. 1 2 opt S x . However, it rises as opt rises. For rapid interpretation by non-mathematicians it may be useful to develop a dimensionless variable E, represent- ing how far the true optimum is away from the centre of the tested range, running from 0 when opt = 0, to 1 when 1 2 opt S x . Reworking Equation (3) to use this, and introducing W = 1Sx , the width of the range of settings tested, gives: 1 2 2 2 2 1 31 1 ˆ1154 1 opt und S S SE xE WA S RS S (4) This formula is only valid when the condition de- scribed above in (2) is satisfied. That condition, the bare minimum number of raw measurements needed before the standard error of the optimum can be validly esti- mated, can be approximated in this simplified way: 1 2 2 180 und RS WA x (5) Copyright © 2011 SciRes. AM
D. P. FRANCIS 1502 If exceeds the above formula by (for example) more than 5-fold, the uncertainty of the optimum is iden- tified reliably using Equation (4). If exceeds by only a small margin, Equation (4) will underestimate the observed scatter between repeat optimizations. Finally, if R S does not even exceed the right hand side, the dataset is so poor in information content that it should be dis- carded, and the protocol redesigned. RS RS In practice the doctor does not know the underlying values of 1 and und but instead assesses them from clinical measurements: 1 ˆ and ˆ respectively. Figure 2 summarizes all the relevant variables in a format convenient for visual appreciation. It shows the steps necessary to gauge the uncertainty of the optimum. The expressions can be applied to any unit of “pace- maker setting”, and any unit of “physiological response”. Typically pacemaker settings (atrioventricular and inter- ventricular delay) are expressed in either milliseconds or seconds. Physiological response may be based on flow, for which suitable units might be true flow rates (e.g. ml/min, L/min) or expressed per beat (ml/beat), or as a peak instantaneous flow rate, or as an average velocity (cm/min or m/min) or peak velocity, or velocity-time integral (cm/beat). Alternatively physiological response may be pressure (e.g. mmHg) as a systolic, mean, or pulse pressure, or a value derived from pressure such as intraventricular peak first derivative of pressure (dp/dtmax). In principle uncalibrated physiological response vari- ables and even those of unknown physical unit may be used, as long as it is reasonable to believe they vary ap- proximately linearly with cardiac performance. Any physical units may be used to express scatter 1 and width W but, once they are decided, the units that must be used to express curvature und are [scatter]/ [width]2. 3.5. Simplified Expression for Rapid Appreciation In practice the number of settings tested is usually fairly large, e.g. S 6, and E does not exceed 1. Thus to explain the implications of the formula to a non-mathe- matician protocol designer, it may be sufficient to use the following simplified approximation. First check that 1 2 2 ˆ 180 ˆ RS WA , otherwise redesign the experiment, with more replicates or with a variable which has a sma- ller 1 ˆ ˆ ratio. Second: 12 3 ˆ~1 opt und SE xE A WRS 15 (6) The left factor contains 3 variables known precisely from the experimental design. The right factor contains 3 variables that must be estimated from observations in the patient. 3.6. Examples of Application to Existing Protocols 3.6.1. Example 1. Atrioventricular Delay Optimization In a published study [10] of clinical optimization in 15 patients, there were S = 6 atrioventricular delay settings tested, and R = 6 replicate measurements. The width of the spectrum of settings covered was W = 0.200 − 0.040 s = 0.160 s. The observed scatter between individual re- plicate measurements was 1 ˆ = 3.9 mmHg. The ob- served curvature was ˆ = 1194 mmHg·s–2. With these observed data, 1 2 2 ˆ ˆ WA 180 was 2.9, which RS com- fortably exceeded. For an optimum lying near the middle of the tested range, E ≈ 0, so 3153.9 ˆ0.200 1194 66 35 opt SE x ≈ 0.005 s. For optima half-way to the edge of the tested spectrum, ˆopt SE x is almost double, at 0.010 s. For optima at the edge of the tested spectrum, is almost 4-fold higher than at the middle of the range, i.e. ≈ 0.019 s. ˆopt SE x Pacemakers only permit quantized values to be pro- grammed, for example in steps of 0.010 s. For optima lying near the centre of the tested range in that study, the optimization procedure can be seen to be sufficient to identify the optimum for clinical purposes. However, for optima at the edge of the tested range, the protocol would need to be adjusted to maintain that level of optimization precision. The most generically effective step to achieve this would be to conduct more replicate measurements. 3.6.2. Example 2. Interventricular Delay Optimization In the same study, [10] optimization of interventricular delay was also examined. There were again S = 6 settings tested, and R = 6 replicate measurements. The width of the spectrum of settings covered was W = 0.120 s. Ob- served measurement scatter 1 mmHg. Curva- ture was measured to be Aund = 67 mmHg·s–2. The bare minimum number of measurements needed ˆ3.9 1 2 2 ˆ 180 ˆ WA is ~2900, which RS falls far below. This signals that application of Equation (4) will likely sub- stantially underestimate the true standard error of the optimum. Despite therefore being only a crude lower estimate, Copyright © 2011 SciRes. AM
D. P. FRANCIS Copyright © 2011 SciRes. AM 1503 80100 120140 160 90 100 110 1 111 12 2 2 2 2 3 3 3 3 3 AV delay (ms) Physiological respo nse (mmHg) Curvature Coefficient of AV delay 2 in curve fit A = 0.0083 mmHg ms -2 1 Measurement scatter Extremeness Standard deviation of multiple replicate measurements at one setting How far out the optimum is from the middle of the tested range E ≈0 Extremeness 1 1 = 2.5 mm H g 80100 120 140 160 Width Settings Replicates S = 5 W = 120 ms R = 3 3 variables set by protocol design3 variables measured experimentally 0 3 variables set by protocol design ms9.0 4 1 151 ˆ ˆ 1 113 ˆ2 2 2 2 1 E S S A S S W SR xSE opt 2 2ˆ ˆ 180 1 AW SR Step 1. To exclude grossly inade quate optimizations, che ck that In this case, 15 >> 0.08, which is very satisfactory. Step 2. Calculate standard error of optimum 80100 120140 160 90 100 110 1 111 12 2 2 2 2 3 3 3 3 3 AV delay (ms) Physiological respo nse (mmHg) Coefficient of AV delay 2 in curve fit A = 0.0083 mmHg ms -2 Curvature 1 Measurement scatter Extremeness Standard deviation of multiple replicate measurements at one setting How far out the optimum is from the middle of the tested range E ≈0 Extremeness 1 1 = 2.5 mm H g 3 variables measured experimentally 0 80100 120 140 160 Width Settings Replicates S = 5 W = 120 ms R = 3 2 2ˆ ˆ 180 1 AW SR Step 1. To exclude grossly inade quate optimizations, che ck that In this case, 15 >> 0.08, which is very satisfactory. Step 2. Calculate standard error of optimum ms9.0 4 1 151 ˆ ˆ 1 113 ˆ2 2 2 2 1 E S S A S S W SR xSE opt Figure 2. Illustration of the 6 variables that affect precision of the optimum. Note: Three are set by the design of the protocol and are therefore under the control of the clinician (within reason) and three are measured experimentally and depend on biology and on the physiological variable chosen for monitoring. Step 1 highlights the necessity for performing enough repli- cate measurements before attempting to calculate a standard error. for an optimum lying in the middle of the tested range, Equation (4) provides > ~0.118 s. For optima at the edge of the tested spectrum, is 4-fold higher, i.e. > ~0.422 ms. These lower limits on the stan- dard error are so large that the 95% confidence intervals (1.96 standard errors) dwarf the entire spectrum of clinically plausible settings. ˆopt SE x ˆopt SE x This analysis shows that interventricular delay opti- mization conducted in this way in such patients is unre- liable. Interventricular delay optimization protocols recommending a few (or even just one) measurement at each setting will never work. Biological variability has an ineradicable lower limit which can only be overcome by repetitions so numerous as to be unrealistic for man- ual methods such as echocardiography, and challenging even for automatable haemodynamic approaches. The curious silence of the world literature on the ele- mentary question of blinded test-retest reproducibility of interventricular delay optimization is a dog that didn’t bark [11]. 3.7. Examples of Application in Protocol Design A doctor designing an optimization protocol can use this formula to ensure that the protocol delivers a clinically- satisfactory degree of precision. Suppose the protocol needs to deliver optimization within 0.010 s on 95% of occasions, i.e. ˆopt SE x needs to be 0.005 s. Rewriting Equation (4), we see that the number of replicate meas- urements required is: 1 2 22 2 32 11 1 31 4 ˆund opt SS RE A SS S WSEx 15 (7). A simplified form of the relationship, for easy appre- ciation by non-mathematicians, is shown below: 1 2 2 31 ~1 ˆund opt RE SA WSEx 15 Three of the variables are handled easily. The protocol designer can choose how many settings to test (S, e.g. 6), and the width covered (W, e.g. 0.160 s) and can arrange from physiological knowledge that the optimum will lie in the middle 50% of the spectrum of tested settings, i.e. E< 1/2. The value of 1 ˆ ˆ must be assessed experimentally
D. P. FRANCIS 1504 for each proposed optimization variable. This requires only a few minutes per subject. Table 1 shows how this ratio, and other variables, af- fect the number of replicates needed, R, to achieve the desired optimization precision. Although mathematically the tested range W, the re- quired precision , and the scatter-to-curvature ratio ˆopt SE x 1 The many haemodynamic variables that are affected by AV delay appear to be initially change proportionally [12] which suggests that Aund may be approximately the same proportion of the baseline value of each variable. The protocol designer desiring small 1 ˆ ˆ should therefore favour the variable whose scatter 1 ˆ is the smallest proportion of its baseline value. ˆ ˆ each have the same impact on the number of replicates required, in clinical practice 1 ˆ ˆ4. Limitations has by far the greatest range of possible values, and the design process should therefore focus on ensuring that this is small, if the optimization protocol is to be practical. Although it is convenient to describe optimization re- sponses with a parabola, the true underlying shape may Table 1. Impact of physiology (scatter, curvature and extremeness) and protocol design (number of tested settings, width of spectrum) on the number of replicates needed for optimization of a desired level of precision. Distribution of tested settings and physiological characteristics Number of replicates per setting needed to achieve this Number of tested settings Width of spectrum Desired precision ExtremenessScatter Curvature Scatter/ Curvature ratio S W ˆopt SE x E 1 ˆ ˆ 1 ˆ ˆ R s s mmHg mmHg·s–2 s 2 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 4 6 8 10 12 0.16 0.16 0.16 0.16 0.16 0.16 0.16 0.16 0.16 0.16 0.16 0.16 0.16 0.16 0.16 0.16 0.16 0.16 0.16 0.08 0.12 0.16 0.20 0.16 0.16 0.16 0.16 0.16 0.010 0.005 0.005 0.005 0.005 0.005 0.005 0.005 0.005 0.005 0.005 0.005 0.005 0.005 0.005 0.002 0.005 0.010 0.020 0.005 0.005 0.005 0.005 0.005 0.005 0.005 0.005 0.005 0.5 0.5 0.5 0.5 0.5 0.5 0.5 0.5 0.5 0.5 0 0.25 0.5 0.75 1 0.5 0.5 0.5 0.5 0.5 0.5 0.5 0.5 0.5 0.5 0.5 0.5 0.5 3.9 3 3 3 3 3 1 3 10 30 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 1194 30 100 300 1000 3000 1000 1000 1000 1000 1000 1000 1000 1000 1000 1000 1000 1000 1000 1000 1000 1000 1000 1000 1000 1000 1000 1000 0.00327 0.1 0.03 0.01 0.003 0.001 0.001 0.003 0.01 0.03 0.003 0.003 0.003 0.003 0.003 0.003 0.003 0.003 0.003 0.003 0.003 0.003 0.003 0.003 0.003 0.003 0.003 0.003 6 21,929 1974 219 20 2 2 20 219 1974 5 9 20 38 64 123 20 5 1 79 35 20 13 24 20 17 14 13 In practice the number of tested settings, width of spectrum, desired precision, and extremeness, all have a limited range of realistic values (as shown). In con- trast, curvature and scatter have a wide range of possible values and, although rarely formally quantified in research reports, have enormous impact on the precision of the optimization process. This table can be used for physiological response measures that have any measurement unit (not only mmHg) since it is the relative size of scatter versus curvature, rather than their absolute values, that is important. For example, if “mmHg” was replaced by “% of value at refer- ence setting”, the table could without any other alteration be used for any variable such as echo-Doppler velocity-time integral, uncalibrated pressure signal, bioimpedance-derived stroke volume, or any marker of pulsatility in the peripheral circulation. Copyright © 2011 SciRes. AM
D. P. FRANCIS Copyright © 2011 SciRes. AM 1505 be more complex. For example, where intrinsic (or even just fusion) conduction becomes active as AV delay is lengthened, the data points may deviate upwards from the parabolic trend, and so fitting to a parabola may bias the fitted AV delay optimum toward higher values. Nev- ertheless, the principles in this manuscript would still hold true. Moreover, in the close vicinity of the optimum, which is relevant to precise optimization, curvature may be closer to parabolic. Merely applying formulae will not make optima more precise. Only selecting a physiological variable of suita- bly low 1 ˆ ˆ ratio, and taking time to conduct enough measurements (shown in Equation (7)), can im- prove precision. Lack of interest in scatter and curvature in a doctor conducting optimization is as uninspiring as lack of interest in wings and engines in an aircraft pilot. 5. Conclusions The practical implications may be summarized from in- spection of each factor in Equation (6) in turn. To obtain precise clinical optima: most importantly, either commit enough resources to obtain enough measurements, or do not embark on optimization; cover a wide range of pacemaker settings while re- maining within the parabolic region of the response curve; choose a physiological variable with as narrow a random variability (in relation to its sensitivity to pacemaker setting change) as possible; and if possible, design the spectrum of settings tested so that the true optimum will lie near its middle rather than at an extreme. Clinicians treating patients and scientists designing and conducting studies have not had simple, quick me- thods to establish the uncertainty in planned or actual optimization procedures. As a result, patients may be undergoing apparent optimization procedures that are in fact worsening the programming of their pacemaker. Moreover without methods for recognizing unreliable optimizations, clinicians finding the optimum appearing to change every 1 or 2 years may feel compelled to carry out worthless optimization procedures more frequently, [8] which wastes clinical resources and may be harmful to patients. The results in this paper permit easy quantification of uncertainty of the optimum of a cardiac resynchroniza- tion therapy pacemaker. They help patients gain the best physiological benefit, help doctors design protocols that deliver efficient and reliable optimization, and assist in distinguishing genuine change in patient physiology over time, versus random noise. An unexamined optimization is not worth doing. 6. Acknowledgements The author is supported by a Senior Research Fellowship (FS/10/038) from the British Heart Foundation. 7. References [1] W. T. Abraham, W. G. Fisher and A. L. Smith, “Cardiac Resynchronization in Chronic Heart Failure,” New Eng- land Journal of Medicine, Vol. 346, 2002, pp. 1845-1853. doi:10.1056/NEJMoa013168 [2] A. Kyriacou, P. A. Pabari, and D. P. Francis, “Cardiac Resynchronization Therapy Is Certainly Cardiac Therapy, But How Much Resynchronization and How Much Atrioventricular Delay Optimization?” Heart Failure Re- views, 2011, Article in Press. doi:10.1007/s10741-011-9271-1 [3] J. G. Cleland, J. C. Daubert, E. Erdmann, N. Freemantle, D. Gras, L. Kappenberger and L. Tavazzi, “Cardiac Re- synchronization-Heart Failure (CARE-HF) Study Inves- tigators. The Effect of Cardiac Resynchronization on Morbidity and Mortality in Heart Failure,” New England Journal of Medicine, Vol. 352, No. 15, 2005, pp. 1539- 1549. doi:10.1056/NEJMoa050496 [4] A. Auricchio, C. Stellbrink, S. Sack, M. Block, J. Vogt, P. Bakker, C. Huth, F. Schöndube, U. Wolfhard, D. Böcker, O. Krahnefeld and H. Kirkels, “Pacing Therapies In Congestive Heart Failure (PATH-CHF) Study Group. Long-Term Clinical Effect of Hemodynamically Opti- mized Cardiac Resynchronization Therapy in Patients With Heart Failure and Ven-Tricular Conduction Delay,” Journal of the American College of Cardiology, Vol. 39, No. 12, 2002, pp. 2026-2033. doi:10.1016/S0735-1097(02)01895-8 [5] M. D. Bogaard, P. A. Doevendans, G. E. Leenders, P. Loh, R. N. Hauer, H. van Wessel and M. Meine, “Can Optimization of Pacing Settings Compensate for a Non- Optimal Left Ventricular Pacing Site?” Europace, Vol. 12, No. 9, 2010, pp. 1262-1269. doi:10.1093/europace/euq167 [6] I. E. van Geldorp, T. Delhaas, B. Hermans, K. Vernooy, B. Broers, J. Klimusina, F. Regoli, F. F. Faletra, T. Moc- cetti, B. Gerritse, R. Cornelussen, J. Settels, H. J. Crijns, A. Auricchio and F. W. Prinzen, “Comparison of a Non- Invasive Arterial Pulse Contour Technique and Echo Doppler Aorta Velocity-Time Integral on Stroke Volume Changes in Optimization of Cardiac Resynchronization Therapy,” Europace, Vol. 13, No. 1, 2011, pp. 87-95. doi:10.1093/europace/euq348 [7] Z. I. Whinnett, J. E. Davies, G. Nott, K. Willson, C. H. Manisty, N. S. Peters, P. Kanagaratnam, D. W. Davies, A. D. Hughes, J. Mayet and D. P. Francis, “Efficiency, Re- producibility and Agreement of Five Different Hemo- dynamic Measures for Optimization of Cardiac Resyn- chronization Therapy,” International Journal of Cardio- logy, Vol. 129, No. 2, 2008, pp. 216-226.
D. P. FRANCIS 1506 doi:10.1016/j.ijcard.2007.08.004 [8] P. A. Pabari, K. Willson, B. Stegemann, I. E. van Geldorp, A. Kyriacou, M. Moraldo, J. Mayet, A. D. Hughes and D. P. Francis, “When Is an Optimization Not an Optimiza- tion? Evaluation of Clinical Impli-Cations of Information Content (Signal-to-Noise Ratio) in Optimization of Car- diac Resynchronization Therapy, and How to Measure and Maximize It,” Heart Failure Reviews, Vol. 16, No. 3, 2011, pp. 277-290. doi:10.1007/s10741-010-9203-5 [9] W. T. Abraham, D. Gras, C. M. Yu, L. Guzzo and M. S. Gupta, “Rationale and Design of A Randomized Clinical Trial to Assess the Safety and Efficacy of Frequent Opti- mization of Cardiac Resynchronization Therapy: The Frequent Optimization Study Using the Quickopt Method (FREEDOM) Trial,” American Heart Journal, Vol. 159, No. 6, 2010, pp. 944-948. doi:10.1016/j.ahj.2010.02.034 [10] Z. I. Whinnett, J. E. R. Davies, K. Willson, C. H. Manisty, A. C. Chow, R. A. Foale, D. W. Davies, A. D. Hughes, J. Mayet and D. P. Francis, “Haemodynamic Effects of Changes in AV and VV Delay in Cardiac Resynchroniza- tion Therapy Show a Consistent Pattern: Analysis of Shape, Magnitude and Relative Importance of AV and VV Delay,” Heart, Vol. 92, 2006, pp. 1628-1634. doi:10.1136/hrt.2005.080721 [11] A. C. Doyle, “The Curious Incident of the Dog in the Night-Time,” Strand Magazine, Vol. 4, No. 4, 1892, pp. 645-660. http://www.archive.org/download/StrandMagazine24/ Strand24_text.pdf [12] C. H. Manisty, A. Al-Hussaini, B. Unsworth, R. Baruah, P. A. Pabari, J. Mayet, A. D. Hughes, Z. I. Whinnett and D. P. Francis, “The Acute Effects of Changes to AV De- lay on Blood Pressure and Stroke Volume: Potential Im- plications for Design of Pacemaker Optimization Proto- cols,” Circulation: Arrhythmia and Electrophysiology, Article in Press. Copyright © 2011 SciRes. AM
|