This paper presents a decoupling theory named as Vascular Loading Decoupling Technique (VLDT), extended from the beat-based Tissue Control Method (TCM), for noninvasive measurement of real-time-based intra-arterial blood pressure and dynamic compliance of blood vessel. Both VLDT and TCM are based upon the decoupling theory to cause the arterial pulsation without the influence from surrounding tissues and measure the variations of the vascular diameter at critical depth. Meanwhile, the AC part of blood pressure is lost, that is, the reference pressure for AC controller is absent as well. To remedy this problem, VLDT employs Step-Hold control rules and cubic spline curve fitting technique to estimate the reference pressure, identify the impedance of blood vessel, and calculate the AC control gain at Hold stage, then track the AC part of blood pressure and compute the real-time arterial blood pressure at Step stage by turns, which makes the real time measurement feasible; moreover, the dynamic compliance of blood vessel can be extracted from real-time impedances of blood vessel. Since the compliance of blood vessel is decoupled from surrounding tissues, it is valuable for diagnosing the severity of vascular sclerosis. The simulation results showed that the VLDT approach is superiority over TCM. This could provide new prospective to circulatory medicine research.
Since Marey [
Based upon the model for transmission of arterial blood pressure [
Therefore, Lin [
To remedy this problem, we proposed a new approach named as real-time based Vascular Unloading Decoupling Technique (VLDT). It employs the Step-Hold control rules and featured spline curve fitting technique to estimate the reference pressure and measure the impedance of blood vessel. It makes the measurements of real time intra-arterial blood pressure and the dynamic compliance of blood vessel feasible.
The models and theory of the Tissue Control Method is restated in this paragraph.
a, the surface of the radius is d and the points b and c are the blood vessel’s upper and lower points respectively. Assuming that the motion model of the arterial blood pressure is lumped, one may express each part, from the detection head to the radius, by a mass element (M), a damping element (D) and a stiffness element (K) to describe the motion behavior caused by arterial blood pressure transmitted from inside the vessel to the skin above. If the force transmissibility is equivalent to the pressure transmissibility from the blood vessel to the skin, the elements of M, D and K may be defined as
M = P M λ ¨ , D = P D λ ˙ , K = P K λ (1)
where P is the pressure and λ is the displacement. Then, by using the Laplace transform, one may define the impedance Z for each part as
Z ( s ) = P λ = M s 2 + D s + K (2)
If the subscripts h, t1, v and t2 denote the parts of detection head, upper tissue, blood vessel and lower tissue, respectively, then the symbols of Zh, Zt1, Zv and Zt2 indicate the impedances for the corresponding parts of the lumped model.
If the displacement λ is analogous to the voltage and the pressure P is analogous to the current, then the mechanical impedance will be analogous to the electrical admittance.
impedances of the detection head (Zh), upper tissue (Zt1) and the lower tissue (Zt2) as the impedance of tissue (Zt) to simplify the circuit, as shown in
1 Z t = 1 Z h + 1 Z t 1 + 1 Z t 2 (3)
The abovementioned detection head can be a cuff or an artery tonometry with actuators and sensors.
According to the superposition theorem, one may express the response of the blood pressure measured on the skin (Ps) due to excitations of the pressed-down depth (λs) and the arterial blood pressure (Pb) as
P s = Z t Z v Z t + Z v λ s + Z t Z t + Z v P b (4)
where the arterial blood pressure Pb can be summed by DC part of blood pressure ( P ¯ b ) and the AC part of blood pressure (ΔPb).
If one presses the skin to a critical depth ( λ ¯ s ) such that the DC part of sensor’s pressure ( P ¯ s ) measured on the skin equals the DC part of arterial blood pressure ( P ¯ b ), then (4) can be written as
P ¯ s = Z ¯ t Z ¯ v Z ¯ t + Z ¯ v λ ¯ s + Z ¯ t Z ¯ t + Z ¯ v P ¯ b , P ¯ s = P ¯ b (5)
Solving (5) gives
P ¯ s = Z ¯ t λ ¯ s = P ¯ b ⇒ Z ¯ t = P ¯ s λ ¯ s = Z ¯ s , (6)
Obviously, (6) states that at the critical depth ( λ ¯ s ) where, the impedance measured on the skin is equal to the equivalent impedance of detection head, upper tissue and lower tissue only, and the impedance of blood vessel is no longer involved in the analysis. It further illustrates that the impedance of arterial blood vessel ( Z ¯ v ) and the impedance of other surrounding tissues ( Z ¯ t ) are decoupled.
When the AC part of arterial blood pressure (ΔPb) is imposed on the DC part of the arterial blood pressure ( P ¯ b ) at critical depth ( λ ¯ s ), (4) can be rewritten as
P ¯ s + Δ P s = Z t Z v Z t + Z v ( λ ¯ s + Δ λ s ) + Z t Z t + Z v ( P ¯ b + Δ P b ) (7)
Substituting (6) into (7) yields
Δ P s = Z t Z v Z t + Z v ( 1 − Z ¯ t Z t ) λ ¯ s + Z t Z v Z t + Z v Δ λ s + Z t Z t + Z v Δ P b (8)
If one moves the detection head up and down ( Δ λ s ) at critical depth ( λ ¯ s ) to keep the AC part of the sensor’s pressure measured on the skin ( Δ P ¯ s ) is equal to zero, i.e. Δ P ¯ s = 0 and then the impedance of surrounding tissues remains unchanged, i.e. Z t = Z ¯ t , because it compressed by P ¯ s = P ¯ b only. Then (8) becomes
Δ λ s = − 1 Z v Δ P b = Δ λ v or Δ P b = − Z v Δ λ s (9)
where Δ λ v is the variation of the diameter of blood vessel. It also represents that the movements of blood vessel under this situation is varying with zero transmural pressure, because the artery’s inside and outside pressures are equal, P ¯ s = P ¯ b , that is truly vascular unloading. This derivation reveals that the variation of the displacement measured on the skin ( Δ λ s ) is equivalent to the variation of the diameter of blood vessel ( Δ λ v ) due to tracking AC part of the arterial blood pressure, but meanwhile, the AC part of blood pressure is lost, Δ P b ≈ 0 .
However, the significant implication of this result is that the pulsation of blood vessel is therefore decoupled from the surrounding tissues which enable one may identify the impedance of blood vessel (Zv) and then compute the intra-arterial blood pressure as shown in (10).
P b = P ¯ b + Δ P b = P ¯ s − Z v Δ λ s (10)
For obtaining the situation of (10), we present the real-time-based Vascular Loading Decoupling Technique (VLDT) instead of the beat-based Tissue Control Method (TCM) to detect the dynamic impedance of blood vessel and the instantaneous intra-arterial blood pressure.
Three key techniques are used for the VLDT. There are 1) separating the arterial blood pressure into two parts: DC part of blood pressure ( P ¯ b ) and the AC part of blood pressure ( Δ P b ) and using self-adaptive control rules to create the decoupled situation to provide the possibility of measuring the impedances of surrounding tissues and blood vessel respectively; 2) using Step-Hold control rules to estimate reference pressure, identify the impedance of blood vessel, track the AC part of blood pressure and compute the real-time intra-arterial blood pressure; 3) extracting the dynamic compliance of blood vessel (C3) from the impedance of blood vessel (Zv) by using parameter estimation technique.
Observing
P ¯ s s = G d H 1 + H 2 1 + G d H 1 P ¯ b (11)
where the dividend impedance H 2 = Z t / ( Z t + Z v ) is smaller than 1.
Therefore, set the DC open loop gain K d = G d H 1 = 499 , the error of DC control results in (11) is less than 0.2%. It means that setting the DC control gain (Gd) to have DC open loop gain (Kd) is constant will keep the detection head nearly to the critical depth, which is similar to the analysis of DC part of arterial blood pressure excitation.
When the DC-driven actuator is pressed to reach the critical depth ( λ ¯ s ), an AC controller is needed to move up and down the AC-driven actuator with displacement ( Δ λ s ) in order to have the AC part of sensor’s pressure be zero, Δ P s s = 0 , as shown in
The control theory gives the AC part of sensor’s pressure ( Δ P s s ) shown in (12).
Δ P s s = H 2 Δ P b 1 + G a H 1 (12)
Similarly, the self-adaptive control rule is used. Adjust the AC control gain (Ga) so as to maintain the AC open loop gain to be the fixed value, G a H 1 = K a . If set K a = 199 , AC part of sensor’s pressure ( Δ P s s ) is smaller than 0.5%, i.e. Δ P s s ≈ 0 . By now, the pulsation of arterial blood vessel at critical depth is similar to the analysis of AC part of arterial blood pressure ( Δ P b ).
From
Δ P ^ b ( n ) = Δ P s s ( n ) − Δ P ˜ b ( n ) = − Δ P ( n ) − Δ P e ( n − 1 ) G a ( n − 1 ) H 1 (n)
Then the parallel impedance (H1) can be measured by (13).
H 1 ( n ) = − Δ P e ( n ) − Δ P ^ b ( n ) Δ P e ( n − 1 ) G a ( n − 1 ) (13)
However,
sum of the reference pressure ( Δ P ^ b ) and the output pressure ( Δ P ˜ b ) of parallel impedance (H1), Δ P s s = − Δ P e = Δ P ^ b + Δ P ˜ b . The reference pressure ( Δ P ^ b ) cannot be measured unless the output pressure ( Δ P ˜ b ) of parallel impedance (H1) to be zero. Therefore, we provided Step-Hold control rules to cope with this problem. The period of either Step or Hold stage lasts 2 millisecond, which is at the frequency rate of 500 Hz. At Hold stage, let AC-driven actuator is idling, the AC part of displacement ( Δ λ s ) is zero and causes the output pressure ( Δ P ˜ b ( n − 1 ) ) of parallel impedance is zero, then the measurement of AC part of sensor’s pressure is equal to the reference pressure, Δ P s s ( n − 1 ) = Δ P ^ b ( n − 1 ) .
Use three Hold stages at beginning to have three instant reference pressures Δ P ^ b ( n − 3 ) , Δ P ^ b ( n − 2 ) , and Δ P ^ b ( n − 1 ) , and employ the featured cubic spline curve fitting technique to estimate the nth stage of reference pressure ( Δ P ^ b ( n ) ), then compute the impedance of blood vessel ( Z v ( n ) ) at Hold stage by (13) and (14).
1 Z v ( n ) = 1 H 1 ( n ) − 1 Z t ( n ) (14)
and the AC control gain ( G a ( n ) ) is obtained by (15) as well.
G a ( n ) = K a H 1 ( n ) (15)
After that, go to Step stage (nth stage) to actuate the AC-driven actuator with AC control gain ( G a ( n ) ) that maintains the AC part of sensor’s pressure is equal to the one of (1 + Ka) of reference pressure, i.e. Δ P s s ( n ) = − Δ P ^ b ( n ) / ( 1 + K a ) and earn variation of blood vessel diameter from AC part of displacement sensor, i.e. Δ λ s ( n ) ≅ Δ λ v ( n ) .
Finally, (16) gives the nth stage of intra-arterial blood pressure.
P b ( n ) = P ¯ s ( n ) − Δ λ s ( n ) Z v ( n ) (16)
Repeat the Hold and Step stages to obtain the impedance of blood vessel ( Z v ( n ) ) and the arterial blood pressure ( P b ( n ) ) at each cycle until to the end of the measurement.
In the light of the definition of impedance Zv, the equivalent mechanical lumped elements, such as mass (Mv), damping (Dv) and stiffness (Kv) constitute (17) as follows:
− Z v ( s ) Δ λ ′ s = Δ P ′ b ⇒ − ( M v s 2 + D v s + K v ) Δ λ ′ s = Δ P ′ b (17)
Through bilinear transform, matrix parameters of mass (Mv), damping (Dv) and stiffness (Kv) can be identified and shown in (18):
[ M D K ] = [ Δ λ s ( n ) Δ λ s ( n − 1 ) Δ λ s ( n − 2 ) Δ λ s ( n − 1 ) Δ λ s ( n − 2 ) Δ λ s ( n − 3 ) Δ λ s ( n − 2 ) Δ λ s ( n − 3 ) Δ λ s ( n − 4 ) ] − 1 [ Z v ( n ) + 2 Z v ( n − 1 ) + Z v ( n − 2 ) Z v ( n − 1 ) + 2 Z v ( n − 2 ) + Z v ( n − 3 ) Z v ( n − 2 ) + 2 Z v ( n − 3 ) + Z v ( n − 4 ) ] (18)
The mass (Mv), damping (Dv) and stiffness (Kv) of blood vessel can be further extracted by (19) as follows:
M v = T 2 16 ( M − D + K ) ; D v = T 4 ( M − K ) ; K v = 1 4 ( M + D + K ) (19)
where T is the sampling period.
The reciprocal of the stiffness (Kv) is the dynamic compliance of blood vessel, which is defined as C3. It is different from the Artery Elasticity Indexes of C1 and C2 provided by HDI CvProfilor [
To verify the feasibility of VLDT and the superiority over TCM, the framework of the simulation program run by Simulink software shown in
Thirty-five previous TCM experimental data sets give each set of arterial blood pressure are used as the real intra-arterial blood pressures and each set of mean impedances of surrounding tissues and blood vessel are used and assumed as parabolic-like curve for simulation.
VLDT proposed the featured cubic spline curve fitting technique to estimate the reference pressure. It segments the estimated spline curves as a combination of monotonic functions by featured points. The featured points are selected as the points where the function value is zero or the nth derivative of function equals to zero, n = 1, 2, and 3. This effectively avoids the occurrence of data scattering or abnormal peaks and robust for estimation. A typical simulation result is shown in
Once the estimated reference pressure is accurate enough, then the results of the impedance of blood vessel and the intra-arterial blood pressure will be satisfied as well.
According to (18) and (19), the lumped parameters of mass (Mv), damping (Dv) and stiffness (Kv) are extracted as shown in
In order to reveal the superiority of VLDT over TCM,
To summarize the superiority of VLDT over TCM, they are shown in
Measurements | TCM | VLDT |
---|---|---|
Intra-arterial blood pressure | 1. Beat based measurement: based upon the previous beat of blood pressure to calculate the current blood pressure. | 1. Real time based measurement: using Step-Hold algorithm and featured spline curve fitting technology to identify the impedance of blood vessel and estimate the blood pressure |
2. Less accuracy: The estimated error about ±2 mmHg. | 2. More accuracy: The estimated error within ±1 mmHg. | |
Dynamic compliance of blood vessel | 1. Only the average of blood vessel’s impedance is measured. 2. Can’t calculate dynamic compliance of blood vessel. | 1. The dynamic compliance of blood vessel (C3) is measured. 2. C3 is directly computed by blood vessel’s displacement and pressure under decoupling condition without the effect of surrounding tissues. 3. it more valuable than C1 and C2 to diagnosing the severity of vascular sclerosis |
This paper primarily extended the beat-based TCM into the real-time-based VLDT for noninvasively measuring the dynamic impedance of blood vessel and the intra-arterial blood pressure. The advantages of VLDT are decoupling the pulsation of blood vessel from surrounding tissues and using Step-Hold control rules and featured cubic spline curve fitting technique to make real-time measurement feasible. The simulation results show that both estimated errors of impedance of blood vessel and intra-arterial continuous blood pressure are less than 1%; and the extracted dynamic compliance C3 is more reliable in diagnosing the severity of vascular sclerosis. This could create a promising instrument for clinical researches of circulatory medicine. We require more efforts for constructing the instrument based upon VLDT, and further clinical experiments are planned.