Open Journal of Orthopedics, 2013, 3, 269-277
http://dx.doi.org/10.4236/ojo.2013.36050 Published Online October 2013 (http://www.scirp.org/journal/ojo)
269
Biodegradation of Polymethylmethacrylate Bone Cement
May Not Be a Serious Issue in Total Hip Arthroplasty—
Retrieval Study for Knoop Hardness and Young’s
Modulus*
Masaaki Maruyama1,2#, William N. Capello2
1Department of Orthopaedic Surgery, Indiana University Medical Center, Clinical Building Suite, Indianapolis, USA; 2Department of
Orthopaedic Surgery, Shinonoi General Hospital, Nagano, Japan.
Email: #sgh_iizu@grn.janis.or.jp
Received August 6th, 2013; revised September 7th, 2013; accepted September 22nd, 2013
Copyright © 2013 Masaaki Maruyama, William N. Capello. This is an open access article distributed under the Creative Commons
Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is
properly cited.
ABSTRACT
Introduction: To investigate a long-term in vivo deterioration of polymethylmethacrylate (PMMA) bone cement over
time, we evaluated retrieved PMMA cement in terms of chemical elements presenting in the cement using energy dis-
persive analysis of X-rays; Knoop hardness; and the Young’s modulus using scanning acoustic microscopy. Materials
and Methods: For mechanical evaluation, we could neglect the influences of entrapped air bubbles or blood by the use
of small specimens. The study was based on thirteen cement samples (six used in the acetabulum and seven in the femur)
derived from eight patients (age at revision surgery: mean 72.5, range 68 to 79). All of these samples were Simplex-P
cement. They were functioning well at least ten years after the previous surgery. Duration until revision surgery was
ranged 12 to 25 years (average, 17.4 years). The reason for revision was aseptic mechanical loosening. Twenty samples
of Simplex-P cement were served by manually mixing as a control. Results: The average of the hardness of the ce-
ment was 17.0 ± 1.2 (range, 13.4 - 20.6). In the control, the hardness was 17.8 ± 1.5 (range, 14.0 - 24.6). There was no
significant difference between these values. The mean of Young’s modulus of the cement was 5.61 ± 0.19 GPa (range,
5.09 - 6.10). In the control, the modulus was 6.04 ± 0.13 GPa (range, 5.68 - 6.45). Although the modulus was signifi-
cantly less than that of the control, there was only 7% decrease in average between twelve and twenty-five years in vivo.
Conclusions: Our results suggest that long-term implantation and functional loading in vivo may not be the limiting
factor in the mechanical integrity of the bone cement.
Keywords: Polymethylmethacrylate Bone Cement; Biodegradation; Total Hip Arthroplasty; Retrieval Study; Knoop
Hardness; Young’s Modulus
1. Introduction
Since Charnley [1] introduced polymethylmethacrylate
(PMMA) bone cement used as a grout to provide me-
chanical fixation of the hip prostheses to bone, and the ce-
ment has been widely used in total hip arthroplasty or
hemi-arthroplasty. The cement must endure considerable
stresses in vivo applications, because its main function is
transference of load from the prosthesis to the bone. In
addition, the cement has the Young’s modulus which was
ranked between that of the cancellous bone and the cor-
tical bone. As the mechanical properties of polymeric
materials are, in general, time-dependent [2], degradation
or deterioration of the cement may influence on the long
term results of total joint replacement surgery. Although
recent reports have confirmed relatively good long-term
results of the cemented hip arthroplasty, the rates of loo-
sening in the reports gradually increase over time [3-5].
Our hypothesis is that biodegradation of PMMA cement
*Comments: Recently, good longevity of cemented total hip arthro-
p
lasty was documented in many articles. One of the factor contributed
to the good results was mechanical properties of the polymethylmetha-
crylate bone cement with little biodegradation. Although there are some
limitations for the retrieval study, we would like to prove biodegrada-
tion of the cement.
#Corresponding author.
Copyright © 2013 SciRes. OJO
Biodegradation of Polymethylmethacrylate Bone Cement May Not Be a Serious Issue in Total Hip
Arthroplasty—Retrieval Study for Knoop Hardness and Young’s Modulus
270
may not contribute to mechanical loosening of the hip
prostheses in long time survival.
Concerning the effect of time on mechanical properties
of the PMMA cement, there were several reports dealing
with a possible deterioration of the cement. Jaffe et al. [6]
stored the cement specimens in bovine serum at 37˚C up
to two years and found no deterioration both in static
properties and in compression-fatigue behavior. During a
study of the causes of failed Judet prosthesis, Scales and
Zarek [7] found some evidence of alteration in the me-
chanical properties of acrylic resin in the course of time.
Lee et al. [8] stored bone cement specimens in isotonic
saline solution at 37˚C and found an increase in com-
pressive strength after seven days. However, they men-
tioned that the strength decreased by 7% and 8.5% after 6
and 12 months, respectively. Rostoker et al. [9] reported
that the PMMA cement removed from rabbits showed a
significant drop in fracture stress, which was determined
by a three-point bending test, during the period between
12 and 26 months after implantation. With regard to ac-
rylic cement retrieved from patients, Holz et al. [10] re-
ported that compressive strength was markedly decreased
between one and two years after implantation. But they
concluded that it was not the aging of the cement. The
follow-up period of these reports was 26 months or less,
which was too short to evaluate the long term effects of
implantation of PMMA. In the literature, little has been
written about the mechanical properties of the cement
which functioned over ten years in an in vivo environ-
ment.
The purpose of our study was to investigate both
Knoop hardness and Young’s modulus of the retrieved
PMMA cement which normally functioned at least ten
years after cemented hemi- or total hip arthroplasty. In
addition, we also aimed to detect chemical elements pre-
senting in the cement and to know some effects of in vivo
environment.
2. Materials and Methods
Between August and December, 1995, forty seven pa-
tients underwent revision total hip arthroplasty in Indiana
University Medical Center and the satellite hospitals. The
main cause for the revision was aseptic mechanical loo-
sening. Loosening due to infection was excluded from
the study, since it seemed to be changed normal physio-
logical condition around the cement. The samples re-
trieved by the ultrasonic tool, such as ULTRADRIVE
(Biomet Inc., Warsaw, Indiana, USA), also were exclud-
ed from the study, because the chemical fabrication of
the cement might be greatly altered due to extremely high
temperature induced by the tool. Nine of the patients (fif-
teen cement samples: seven used in the acetabulum and
eight in the femur) had functioned well as a part of artifi-
cial hip joint during at least ten years after the previous
surgery. PMMA samples from these patients were clear-
ed of bone and soft tissue, and washed out blood and de-
bris. To avoid virus infection in processing them into test
specimens and to prevent from contamination and breed-
ing of microorganisms during storage at room tempera-
ture (23˚C ± 1˚C), they were immersed in 10% formalin
for 72 hours. Subsequently, formalin was washed out by
using distilled water and each sample was cut into two
specimens. The first set of specimens was stored in air at
the room temperature for an average of 15 weeks (range,
10 to 20 weeks). They were subjected to the elemental
analysis. The second set of specimens was stored in phy-
siological saline solution (0.9% sodium chloride) at the
room temperature for an average of 7 weeks (range, 2 to
12 weeks).
Although two of the retrieved cement samples were
known as Simplex-P (Howmedica, Rutherford, New
Jersey, USA), the brands of the remaining thirteen sam-
ples were unknown. Simplex-P cement consisted of the
following components: a liquid containing methylmetha-
crylate monomer, N, N-dimethyl-p-toluidine for promo-
ter of polymerization, and a small amount of hydroqui-
none for inhibitor; and a solid containing methyl metha-
crylate-styrene copolymer, benzoyl peroxide for initiator,
and barium sulfate for radiopaque property. We used
Fourier transform infrared (FT-IR, Nicolet 20SXB FT-IR
Spectrophotometer, Nicolet, Fremont, California, USA)
to identify styrene, which is specific for Simplex-P ce-
ment. As a result, two (No. 5-a and 5-b) of the remaining
thirteen samples did not include styrene element and ele-
ven previously unknown samples were identified a Sim-
plex-P cement. After adding two previously known
Simplex-P samples, the subsequent studies were based
on thirteen cement samples (six used in the acetabulum
and seven in the femur) derived from eight patients (age
at revision surgery: mean 72.5, range 68 to 79) (Table 1 ).
They were functioned well during ten years or more after
the previous surgery. Duration until revision surgery was
ranged 12 to 25 years (average, 17.4 years).
Each sample was sectioned radially (cement used in
the acetabulum) or transversely (cement used in the fe-
mur) by using of thin sectioning machine (South Bay Te-
chnology Inc., California, USA) under water cooling.
The thickness of the test specimens was set for 4 mm to
measure Knoop hardness and 0.7 mm to assess Young’s
modulus. The sectioned surfaces were polished using po-
lishing machines (LECO DS-20, VARI/POL™, and VP-
150, LECO Corporation, Michigan, USA) with metallo-
graphic papers (grit #240, 360, 400, and 600) under wa-
er to achieve the smooth surface. t
Copyright © 2013 SciRes. OJO
Biodegradation of Polymethylmethacrylate Bone Cement May Not Be a Serious Issue in Total Hip
Arthroplasty—Retrieval Study for Knoop Hardness and Young’s Modulus
Copyright © 2013 SciRes. OJO
271
Table 1. Retrieved polymethylmethacrylate bone cement samples.
Sample
No. Diagnosis Gender
Age at
Revision Used with: Duration until
Revision (years)
Causes of
Revision
Cement
fracture
1-a Osteoarthritis Male 74 Acetabular Socket 18 Aseptic Loosening-
1-b Osteoarthritis Male 74 Femoral Stem 18 Aseptic Loosening+
2 Osteoarthritis Female 70 Femoral Stem 15 Aseptic Loosening-
3-a Osteoarthritis Female 75 Femoral Stem 12 Aseptic Loosening+
3-b Osteoarthritis Female 75 Acetabular Socket 12 Aseptic Loosening-
4 Osteoarthritis Female 79 Acetabular Socket 14 Aseptic Loosening-
6-a Osteoarthritis Female 74 Femoral Stem 16 Aseptic Loosening-
6-b Osteoarthritis Female 74 Acetabular Socket 16 Aseptic Loosening-
7-a Osteoarthritis Male 68 Femoral Stem 20 Aseptic Loosening+
7-b Osteoarthritis Male 68 Acetabular Socket 20 Aseptic Loosening-
8-a Osteoarthritis Female 70 Femoral Stem 25 Osteolysis -
8-b Osteoarthritis Female 70 Acetabular Socket 25 Aseptic Loosening+
9 Rheumatoid Arthritis Female 72 Femoral Stem 15 Osteolysis -

0.789
ae
www

Twenty samples of Simplex-P cement (5 mm width
and 30 mm length) were served by manually mixing as a
control, because most of the retrieved cement were made
by manually mixing in air (no vacuum mixing) when the
patients underwent total hip arthroplasty 12 to 25 years
ago. They were stored in physiological saline solution for
three weeks after mixing at the room temperature to
eliminate any effect of after polymerization on the results.
Then, all of the samples were also immersed in 10% for-
malin for 72 hours. Subsequently, formalin was washed
out and the specimens were stored in physiological saline
solution for additional three weeks. Test specimens were
made by cutting and polishing as the same manner (the
same sizes and thickness) as the above mentioned.
where,
a: apparent density affected by the porosity of
the specimen, w: weight of the specimen, we: weight of
the specimen in 100% ethanol, density of 100% ethanol
= 0.789.
The apparent density was corrected by the porosity of
the specimen. As a result, true density of the specimen (
)
was calculated as follows:

s
1p
a


where, ps: porosity of the samples
2.1. Detection of Chemical Elements Presenting
in the Retrieved PMMA Cement
Porosity (ps) and density (
) of the samples were cal-
culated on the 0.7 mm-thick specimens. Contact micro-
radiography (CMR) was taken for each specimen. The
number (n) and radius (r) of the entrapped air bubbles
was counted by direct observation on the radiography us-
ing light microscopy. Five fields of view were randomly
selected for each specimen. Porosity was estimated through
the following formula [11]:

s
Sum of volumes of pores in the field of view S
pArea of the field of viewthickness of specimen
All of the retrieved bone cement samples were examined
using energy dispersive analysis of X-rays (EDAX)
(9900-EDS X-ray Fluorescence, Philips Electronic Co.,
Mahwah, New Jersey, USA). All specimens were steril-
ized with a double dose of gas plasma. Thin films were
pressed at 180˚C from all of the samples and were inves-
tigated by the EDAX analyzer at 5.00 KEV. Chemical
elements were identified from X-ray dispersive pattern.
Simplex-P cement was composed of carbon, hydrogen,
oxygen, nitrogen, barium and sulfur. These elements were
excluded from the detection because the EDAX was qua-
litative analysis.
where, 3
1
4π3r
n
i
i
S

n: the number of the entrapped air bubbles, r: radius of
the entrapped air bubbles 2.2. Knoop Hardness
Knoop hardness measurements were obtained from the
sectioned surface of each 4 mm-thick specimen using
Hardness Tester (LECO M-400, LECO Corporation, Mi-
The average value was regarded as the porosity of the
specimen.
Density was calculated by the following formula:
Biodegradation of Polymethylmethacrylate Bone Cement May Not Be a Serious Issue in Total Hip
Arthroplasty—Retrieval Study for Knoop Hardness and Young’s Modulus
272
chigan, USA). The diamond indenter point was kept on
the surface for 20 second with a 50 g load. The measure-
ments were performed at the area less than 0.5 mm dis-
tant from the surface to the bone (A), the middle area of
the specimen (B), and the area less than 0.5 mm distant
from the surface to the prosthesis (C) (Figure 1). Five
measurement sites were determined at each area (A, B,
and C) so that each site was more than 100 m from an
adjacent site, and was avoided entrapped air bubbles and
blood. The average value was obtained.
In the control group, the point A was identified with
the point C. All other measurements were same as describ-
ed.
2.3. Young’s Modulus
The measurements for Young’s modulus (E) were con-
ducted with the same location as the hardness on the “sa-
lami” specimen with a 0.7 mm-thickness. Specimens were
put in the bottom of a chamber in a scanning acoustic mi-
croscope (UH3, Olympus, Tokyo, Japan). The chamber
was filled with distilled water at a controlled room tem-
perature of 23˚C ± 1˚C. A 50-MHz lens was used to gen-
erate and receive acoustic waves in pulse echo mode.
The acoustic energy was reflected from the top surface of
the specimen and the bottom surface of the specimen.
The two reflections were seen on an oscilloscope and the
time delay between them was measured on oscilloscope.
Ultrasonic velocity (VU) was calculated as follows:
Figure 1. Bone cement specimen (Sliced). The measure-
ments were performed at the area less than 0.5 mm distant
from the surface to the bone (A), the middle area of the
specimen (B), and the area less than 0.5 mm distant from
the surface to the prosthesis (C). Five measurement sites
were plotted at each area for Young’s modulus as well as
Knoop hardness.
2
U
Vat
where, a: thickness of the specimen, t: time delay
Young’s modulus was calculated for each specimen
according to the methods described by Briggs [12] and
Turner et al. [13,14]

2
US L
VV V

where,
: shear modulus,
: density of the specimen, VS:
shear velocity, VL: longitudinal velocity

21E

where,
: Poisson ratio
In case of PMMA,
= 0.34, SL
VV = 13302700
2
0.65 U
EV
Five measurement sites on each area (A, B, and C)
were determined and the average was calculated.
In the control group, measurement sites for Young’s
modulus were also determined in the same manner as for
the hardness measurement.
All procedures were performed at the room tempera-
ture of 23˚C ± 1˚C. The results were analyzed by unpair-
ed t-tests, considering p values of less than 0.05 to be sig-
nificant.
3. Results
3.1. Elements Presenting in the Retrieved
PMMA Cement
The following elements were detected: sodium, silicon,
sulfur, chlorine, calcium, barium, and phosphorus (Table
2). Of these elements, barium and sulfur (barium sulfate)
are found as a radiopaque material in the Simplex-P ce-
ment ingredient lists.
3.2. Knoop Hardness
Knoop hardness of the retrieved PMMA cement was 17.6
± 2.1 (range, 13.4 - 20.6), 17.0 ± 1.7 (range, 14.5 - 19.8),
and 16.4 ± 1.4 (range, 14.6 - 18.4) at the area of A, B,
and C, respectively. In the control group, the hardness
was 18.5 ± 3.7 (range, 15.8 - 24.6) and 17.0 ± 1.4 (range,
14.0 - 18.5) at the area of A (C) and B, respectively.
There was no significant difference between the retrieved
cement and the control in each area (Table 3).
3.3. Young’s Modulus
Young’s modulus of the retrieved PMMA cement was
5.66 ± 0.27 GPa (range, 5.09 - 6.10), 5.56 ± 0.22 GPa
(range, 5.18 - 5.94), and 5.60 ± 0.23 GPa (range, 5.20 -
5.96) at the area of A, B, and C, respectively. In the con-
trol group, the modulus was 5.88 ± 0.37 GPa (range, 5.63
- 7.01) and 5.88 ± 0.35 GPa (range, 5.66 - 6.90) at the
Copyright © 2013 SciRes. OJO
Biodegradation of Polymethylmethacrylate Bone Cement May Not Be a Serious Issue in Total Hip
Arthroplasty—Retrieval Study for Knoop Hardness and Young’s Modulus
Copyright © 2013 SciRes. OJO
273
Table 2. Elements presenting in the retrieved polymethylmethacrylate bone cement samples.
Sample No. Sodium Chlorine Calcium Phosphorus Silicon Barium Sulfur
1-a + + + + +
1-b + + + +
2 + +
3-a + + + + +
3-b + + + +
4 + + + +
6-a + + +
6-b + + + + + +
7-a + + + + + +
7-b + + + +
8-a + + + + + +
8-b + + + +
9 + + + + + +
“+” indicates element present in sample.
Table 3. Hardness and young’s modulus of the r e tri eved polymethylmethacrylate bone cement samples.
Knoop Hardness Young’s Modulus
Sample
No.
Width
(mm)
Porosity
(%) Density (A) <0.5 mm
from the
bone
(B) Middle
of the
sample
(C) <0.5 mm
from the
prosthesis
Average
(A) <0.5
mm from
the bone
(B) Middle
of the
sample
(C) <0.5 mm
from the
prosthesis
Average
1-a 4.3 1.4 1.13 18.1 16.2 14.7 16.3 5.62 5.37 5.56 5.52
1-b 5.3 0.9 1.18 17.6 16.3 14.9 16.3 5.34 5.33 5.33 5.33
2 4.3 2.3 1.27 17.4 14.5 16 16.0 5.87 5.69 5.74 5.77
3-a 3.3 2.1 1.25 17.9 19.8 18.2 18.6 5.75 5.77 5.80 5.77
3-b 3.5 1.9 1.23 16.9 16.5 15.1 16.2 5.70 5.70 5.86 5.75
4 5.3 1.6 1.21 19.2 19.6 16.7 18.5 5.38 5.27 5.20 5.28
6-a 5.5 0.8 1.15 15.3 16.8 16.6 16.2 5.09 5.18 5.69 5.32
6-b 4.3 1.4 1.19 20.6 15.5 17.4 17.8 5.94 5.55 5.59 5.69
7-a 4 3.1 1.20 20.6 16.5 18.4 18.5 5.61 5.69 5.38 5.56
7-b 4.3 1.6 1.19 18.2 18.7 18.4 18.4 5.67 5.61 5.50 5.59
8-a 3.5 1.7 1.19 15.1 15 16.1 15.4 5.61 5.66 5.96 5.74
8-b 6.4 2.1 1.16 13.4 18.3 14.6 15.4 6.10 5.49 5.39 5.66
9 2.8
2.8 1.23 18.1 17.2 16.3 17.2 5.84 5.94 5.84 5.87
Average 4.4 1.8 1.20 17.6 17.0 16.4 17.0 5.66 5.56 5.60 5.61
S.D. 1.0 0.7 0.04 2.1 1.7 1.4 1.2 0.27 0.22 0.23 0.19
Control (n = 20)
Average 5 0.1 1.25 18.5 17 18.5* 17.8 6.19 5.88 6.19** 6.04
S.D. 0 0 0.03 3.7 1.4 3.7* 1.5 0.21 0.12 0.21** 0.13
t-test N.S. N.S. N.S. N.S. p = 0.001p < 0.001 p < 0.001 p < 0.001
*: same as the data (A) (Hardness); **: same as the data (A) (Young’s modulus).
area of A (C) and B, respectively. The moduli were sig-
nificantly less for the areas B and C than those of the
control (p < 0.05) However, the differences of the moduli
between the retrieved PMMA cement and the control
were approximately 7% of the values in average (Table
3).
Biodegradation of Polymethylmethacrylate Bone Cement May Not Be a Serious Issue in Total Hip
Arthroplasty—Retrieval Study for Knoop Hardness and Young’s Modulus
274
4. Discussion
All of the retrieved PMMA bone cement samples had
yellow-brown discolored band which was observed in the
area contacted with bone tissue (Figure 2). There was a
scent of urea or ammonium at the time of cutting the
samples. These physical findings suggested the presence
of bilirubin (C33H36N4O6) by color or urea (CO(NH2)2)
and by smell or both in the cement. Unfortunately, our ma-
chine was unequipped to detect these chemical com-
pounds. However, sodium, chlorine, calcium, and phos-
phorus were detected in the retrieved PMMA bone ce-
ment. These chemical elements were usually present in
human plasma and they seemed to infiltrate into the ce-
ment. All of these elements were not always detected in
the samples. This might be due to threshold of the ma-
chine. There did not seem to be a good explanation for
the presence of silicon. It is unknown whether the pres-
ence of the elements deteriorates the cement. In dentistry,
composite resins, such as Bis-GMA with silica filler,
have been often used for reconstruction of teeth. The re-
sins included the filler so that the mechanical properties
and resistance to wear could be increased. Degradation,
however, occurred by infiltration of water and variable
chemical substances into the resins [15]. This might cau-
se degradation of the cement.
Concerning biodegradation of the cement, Hughes et
al. [16] reported that scission-based structural degrada-
tion occurred in the acrylic bone cement with FTIR ana-
lysis. Ries et al. [17] mentioned that fracture toughness
of Simplex cement did not correlate with time in vivo
between 1 month and 27 years of the samples retrieved
from 43 patients undergoing revision total hip arthroplasty.
Figure 2. Retrieved bone cement sample used in the ace-
tabulum (No. 1-a). The sample had yellow-brown discolored
band (allow) which was observed in the area contacted with
bone tissue.
Oonishi et al. [18] reported that the bending strength in
CMW1 cement after implantation decreased with in-
creasing time in vivo and depended on the density of the
bone cement, which we assume to be determined by the
porosity.
To investigate and evaluate all the mechanical proper-
ties of the retrieved cement, such as tensile, compressive,
impact, flexural, and shear strengths, hardness, and Young’s
modulus were not difficult; however, some difficulties
did occur in evaluating the mechanical strengths of the
samples. There were extra inclusions, such as air or blood
entrapments or both, and laminations in the samples.
These inclusions affected the mechanical strengths. Most
of the retrieved cement was made by manually mixing in
air (no vacuum mixing) when the patients had total hip
arthroplasty. Porosity of the cement caused by entrap-
ment of air bubbles was inevitable in clinical practice. De
Wijn et al. [19] reported that the overall effect of poros-
ity caused by entrapment of air bubbles was approxi-
mately 50% reduction in the impact and flexural strength,
by using specimens with a 20-mm square cross section.
Gruen et al. [20] confirmed that the presence of blood at
the interface further weakened the bond to approximately
25% (tension) and 36% (shear) of the virgin strengths of
the cement, by using specimens 13 mm in diameter and
114 mm long. Holm [21] also tested CMW, Simplex and
Palacos R bone cement by using specimens a size of 4 ×
10 × 110 mm and observed that a 3 ml blood inclusion
decreased the flexural strength significantly in all of the
brands. The ratios of the entrapped air bubbles or blood
were not constant among the retrieved samples, however,
it was impossible to measure the amount and distribution
of these inclusions in the cement. In addition, it was nec-
essary to make several specimens, such as cylinders 12
mm high and 6 mm in diameter for evaluating compres-
sive strength or beams with 1.5 × 10 × 25 mm size for
evaluating flexural strength, as suggested by ASTM stan-
dards [22,23]. Thus, these strengths seemed to be unreli-
able in the retrieved study for evaluating the effect of
time. Unfortunately, since specimens for these strength
tests were too large to exclude the inclusions, such as air
bubbles or blood, it was impossible to neglect its effects.
These sizes were too large to gain enough number of the
specimens with adequate shape and size, because the
retrieved samples were limited in their shape and size.
Porosity of the cement caused by entrapment of air
bubbles was observed in all of the retrieved specimens
which were made by mixing in air at the time of the pre-
vious surgeries. Laminations and blood entrapment were
also recognized in our specimens. To eliminate the influ-
ence of the inclusions, we focused on the measurement
of Knoop hardness and the evaluation of Young’s modu-
lus by scanning acoustic microscopy. Because these pa-
Copyright © 2013 SciRes. OJO
Biodegradation of Polymethylmethacrylate Bone Cement May Not Be a Serious Issue in Total Hip
Arthroplasty—Retrieval Study for Knoop Hardness and Young’s Modulus
275
rameters were proven to be evaluated by 0.1 mm distance
at random on the surface of sectioned specimens, it was
possible to exclude the areas with entrapped air bubbles
or blood, and laminations from the measurement [13,24].
As a result, the size of the specimens was 10 cubic mm
enough to determine these parameters.
Knoop hardness measurements were used as the indi-
cators of the relative degree of cure in light-activated
composite resins [24]. Polymerization of the resins de-
pended upon the formation of free radicals. The presence
of oxygen could have caused retardation of polymeriza-
tion if oxygen reacted with the free radicals. As a result,
the hardness gradually decreased with increasing depth in
the light-activated composite resins. In our study, the
hardness was unrelated with depth. The reason for this
was probably to be explained that our PMMA samples
were polymerized by hand mixing.
To our knowledge, little has been reported on in vivo
changes of the hardness of the PMMA cement for long
term. In Bis-GMA-based dental composite resins, soften-
ing of the matrix was recognized when the resins were
exposed 75% ethanol solution [15]. Our study demon-
strated that the hardness of the retrieved PMMA cement
was not changed more than twelve years in vivo. The
hardness was decreased at twenty five years postopera-
tively (sample No.8), however, its ratio to normal hard-
ness of PMMA cement was more than 80%.
Young’s moduli of solid matter were generally deter-
mined by one of the following three methods: 1. to mea-
sure the strain by applying a known compressive, tensile,
flexural or torsional force to a block of the material; 2. to
measure the natural frequency of vibration of a rod of the
material simply supported at its ends and heavily loaded
by a mass in the middle; and 3. to measure the velocity of
sound in the material, as the velocity depending on
Young’s modulus and the density [12]. The conventional
methods (No. 1 and 2) usually needed relatively large
specimens, which might reflect the influence of the in-
clusions, such as air or blood entrapment. The best of all
methods of measuring Young’s modulus of small speci-
mens are acoustic methods (No. 3) [25]. Acoustic testing
using a scanning acoustic microscope allowed measure-
ment of ultrasonic velocities in PMMA cement with a re-
solution of about 60 m. This was also useful to exclude
the influences of the entrapped air bubbles or blood from
the measurement of Young’s modulus. In addition, the
results were confirmed to reduce experimental error,
achieving much greater precision than the conventional
methods [14].
In the literature, Young’s moduli of Simplex cement
determined by using the conventional method No. 1 were
varied of 0.85 (by torsion) [26,27], 2.14 (by compression)
[26,27], 2.35 - 2.84 (by flexure) [19], 4 and 2.55 GPa (by
tension) [28]. Holm [21] measured Young’s modulus by
using the conventional method No. 2 and indicated that a
3 ml blood additive decreased the modulus significantly
in all brands of CMW, Simplex and Palacos R bone ce-
ment, as well as the flexural strength. However, the mo-
duli determined by the conventional methods (0.85 - 2.84
GPa) were relatively lower than our results (5.09 - 6.45
GPa). The reason for this discrepancy was probably to be
explained that the former methods give bulk properties of
PMMA cement including the effects of pores and the lat-
ter was calculated by measuring ultrasonic velocities in
small areas of the cement reflecting solid PMMA without
pores.
Although Young’s moduli of the retrieved polymethy-
lmethacrylate cement were significantly decreased during
twelve or more years postoperatively in our study, the
differences of the moduli between the retrieved cement
and the control were only 7% of the values in average
(Table 3). In dentistry, Oshida et al. reported that mo-
dulus of elasticity of dental composite resins reduced by
increasing the amount of absorbed water up to three
weeks [29]. In the present study, all of the samples in-
cluding the control were also immersed in physiological
saline solution for six weeks or more. If the amount of
absorbed water in the PMMA cement increases over a
long time, this is an explanation for reduction of the mo-
dulus. In the regression model of the retrieved bone ce-
ment, it was supposed that combination among mole-
cules was cut off in several areas compared with the ce-
ment just after polymerization (Figure 3).
The components of Simplex-P cement were not chan-
ged between 1970s and 1990s. The retrieved cement spe-
cimens were manually mixtured in air until the first half
of 1980s. Therefore, the control samples also were made
by manually mixing in air. Biomechanical properties of
the cured PMMA bone cement, such as Young’s modu-
lus, might be affected by room temperature and humidity
at the time of preparing, as well as porosity, inclusions,
or laminations of the cement. In fact, there was differ-
ence in the modulus between the sample No. 6-a and -b,
as well as No. 1-a and -b (Table 3). The condition at the
Figure 3. Microstructure of the polymethylmethacrylate
bone cement. In the regression model, combination among
molecules was cut off in sever al areas, so called scission-ba-
sed structural degradation.
Copyright © 2013 SciRes. OJO
Biodegradation of Polymethylmethacrylate Bone Cement May Not Be a Serious Issue in Total Hip
Arthroplasty—Retrieval Study for Knoop Hardness and Young’s Modulus
276
manually mixing might influence on the modulus as
compared with the time duration in vivo.
It is difficult to evaluate objectively in vivo deteriora-
tion or aging of the PMMA cement for a long time. As re-
gards the hardness and Young’s modulus, however, the
measurements were easily performed even if the samples
were restricted in size and shape. The decreased Young’s
modulus of the cement was only 7% in average and still
ranked between the moduli of the cancellous bone and
the cortical bone. We concluded that biodegradation eva-
luated by Knoop hardness and Young’s was not a serious
issue for mechanical loosening of the hip prostheses, be-
cause the retrieved PMMA bone cement had no change
in hardness and a slight decrease in Young’s modulus
between twelve and twenty-five years in vivo.
5. Acknowledgements
The authors wish to thank Charles H. Turner, Ph.D., Di-
rector of Orthopaedic Research, Department of Ortho-
paedic Surgery, Indiana University School of Medicine,
Yuichi Takano, M.D., Ph.D., Department of Orthopaedic
Surgery, Akita Red Cross Hospital, Masashi Miyazaki,
D.D.S., Ph.D., Department of Operative Dentistry, Nihon
University School of Dentistry, and Yoshiki Oshida, B.S.,
M.S., Ph.D., Department of Restorative Dentistry, Dental
Materials Research Laboratory, Indiana University School
of Dentistry, for their advice and excellent technical as-
sistance. We acknowledge Keith B. Moore, Ph.D., Pro-
fessor, Department of Dental Materials, Indiana Univer-
sity School of Dentistry, Thomas W. Bauer, M.D., Ph.D.,
Departments of Pathology and Orthopaedic Surgery, The
Cleveland Clinic Foundation for their advice. We also
wish to thank Mr. Bob Hastings P.E., Group Manager,
Biomechanical Technologies, DePuy Inc., for his excel-
lent technical assistance.
REFERENCES
[1] J. Charnley, “Anchorage of the Femoral Head Prosthesis
to the Shaft of the Femur,” The Journal of Bone & Joint
Surgery, Vol. 42-B, 1960, pp. 28-30.
[2] S. Saha and S. Pal, “Mechanical Properties of Bone Ce-
ment: A Review,” Journal of Biomedical Materials Re-
search, Vol. 18, No. 4, 1984, pp. 435-462.
http://dx.doi.org/10.1002/jbm.820180411
[3] L. Neumann, K. G. Freund and K. H. Sørenson, “Long-
Term Results of Charnley Total Hip Replacement. Re-
view of 92 Patients at 15 to 20 Years,” The Journal of
Bone & Joint Surgery, Vol. 76-B, 1994, pp. 245-251.
[4] K. R. Schulte, J. J. Callaghan, S. S. Kelley and R. C. John-
ston, “The Outcome of Charnley Total Hip Arthroplasty
with Cement after a Minimum Twenty-Year Follow-Up.
The Results of One Surgeon,” The Journal of Bone &
Joint Surgery, Vol. 75-A, 1993, pp. 961-975.
[5] B. M. Wroblewski and P. D. Siney, “Charnley Low-Fric-
tion Arthroplasty of the Hip: Long-Term Results,” Clini-
cal Orthopology, Vol. 292, 1993, pp. 191-201.
[6] W. L. Jaffe, R. M. Rose and E. L. Radin, “On the Stabili-
ty of the Mechanical Properties of Self-Curing Acrylic
Bone Cement,” The Journal of Bone & Joint Surgery, Vol.
56-A, 1974, pp. 1711-1714.
[7] J. T. Scales and J. M. Zarek, “Biomechanical Problems of
the Original Judet Prosthesis,” British Medical Journal,
Vol. 1, 1954, p. 1007.
http://dx.doi.org/10.1136/bmj.1.4869.1007
[8] A. J. C. Lee, R. S. M. Ling and S. S. Vangala, “The Mecha-
nical Properties of Bone Cements,” Journal of Medical
Engineering & Technology, Vol. 2, 1977, pp. 137-140.
http://dx.doi.org/10.3109/03091907709160626
[9] W. Rostoker, P. Lereim and J. O. Galante, “Effect of an in
Vivo Environment on the Strength of Bone Cement,” Jour-
nal of Biomedical Materials Research, Vol. 13, 1979, pp.
365-370. http://dx.doi.org/10.1002/jbm.820130303
[10] U. Holz, W. Hemminger and H. Gasse, “Mechanishe Un-
tersuchungen an Explantierten und Frischen Knochenze-
menten,” Archives of Orthopaedic and Trauma Surgery,
Vol. 91, 1979, pp. 365-370.
[11] V. Fano, I. Ortalli and K. Pozela, “Porosity in Composite
Resins,” Biomaterials, Vol. 16, 1995, pp. 1291-1295.
http://dx.doi.org/10.1016/0142-9612(95)91043-X
[12] A. Briggs, “Acoustic Microscopy,” Clarendon Press, Ox-
ford, 1992.
[13] K. Hasegawa, C. H. Turner, R. R. Recker, E. Wu and D.
B. Burr, “Elastic Properties of Osteoporotic Bone Measur-
ed by Scanning Acoustic Microscopy,” Bone, Vol. 16,
1995, pp. 85-90.
[14] C. H. Turner and D. B. Burr, “Basic Biomechanical Mea-
surements of Bone: A Tutorial,” Bone, Vol. 14, 1993, pp.
595-608.
http://dx.doi.org/10.1016/8756-3282(93)90081-K
[15] M. Kawaguchi, T. Fukushima and K. Miyazaki, “Variables
Affecting the Degradation of Dental Composite Resins. 2.
SEM Investigation of Aged Composite Resin,” Dental
Materials, Vol. 13, 1994, pp. 116-121.
[16] K. F. Hughes, M. D. Ries and L. A. Pruitt, “Structural De-
gradation of Acrylic Bone Cements Due to in Vivo and
Simulated Aging,” Journal of Biomedical Materials Re-
search A, Vol. 65, 2003, pp. 126-135.
http://dx.doi.org/10.1002/jbm.a.10373
[17] M. D. Ries, E. Young, L. Al-Marashi, P. Goldstein, A. He-
therington, T. Petrie and L. Pruitt, “In Vivo Behavior of
Acrylic Bone Cement in Total Hip Arthroplasty,” Bioma-
terials, Vol. 27, 2006, pp. 256-261.
http://dx.doi.org/10.1016/j.biomaterials.2005.05.103
[18] H. Oonishi, H. Akiyama, M. Takemoto, T. Kawai, K. Ya-
mamoto, T. Yamamuro, H. Oonishi and T. Nakamura, “The
Long-Term in Vivo Behavior of Polymethyl Methacrylate
Bone Cement in Total Hip Arthroplasty,” Acta Orthopae-
dica, Vol. 82, 2011, pp. 553-558.
http://dx.doi.org/10.3109/17453674.2011.625538
[19] J. R. de Wijn, T. J. J. H. Sloff and F. C. M. Driessens, “Cha-
Copyright © 2013 SciRes. OJO
Biodegradation of Polymethylmethacrylate Bone Cement May Not Be a Serious Issue in Total Hip
Arthroplasty—Retrieval Study for Knoop Hardness and Young’s Modulus
Copyright © 2013 SciRes. OJO
277
racterization of Bone Cement,” Acta Orthopaedica Scan-
dinavica, Vol. 46, 1975, pp. 38-51.
http://dx.doi.org/10.3109/17453677508989190
[20] T. A. Gruen, K. L. Markolf and H. C. Amstutz, “Effects
of Laminations and Blood Entrapment on the Strength of
Acrylic Bone Cement,” Clinical Orthopaedics, Vol. 119,
1976, pp. 250-255.
[21] N. J. Holm, “The Modulus of Elasticity and Flexural Strength
of Some Acrylic Bone Cements,” Acta Orthopaedica
Scandinavica, Vol. 48, 1977, pp. 436-442.
http://dx.doi.org/10.3109/17453677708989727
[22] “Standard Specification for Acrylic Bone Cement,” An-
nual Book of ASTM Standards, F451-95, Philadelphia.
[23] “Standard Test Methods for Flexural Properties of Unre-
inforced and Reinforced Plastics and Electrical Insulating
Materials [Metric],” Annual Book of ASTM Standards,
D790M-86, Philadelphia.
[24] H. Onose, H. Sano, H. Kanto, S. Ando and T. Hasuike, “Se-
lected Curing Characteristics of Light-Activated Compos-
ite Resins,” Dental Materials, Vol. 1, 1985, pp. 48-54.
http://dx.doi.org/10.1016/S0109-5641(85)80024-5
[25] M. F. Ashby and D. R. H. Jones, “Engineering Materials
1: An Introduction to Their Properties and Applications,”
Pergamon Press, Oxford, New York, Seoul, Tokyo, 1993.
[26] A. J. C. Lee, R. S. M. Ling and J. D. Wrighton, “Some
Properties of Polymethylmethacrylate with Reference to
Its Use in Orthopaedic Surgery,” Clinical Orthopaedics,
Vol. 95, 1973, pp. 281-287.
[27] A. J. C. Lee and R. S. M. Ling, “Further Studies of Mono-
mer Loss by Evaporation during the Preparation of Ac-
rylic Cement for Use in Orthopaedic Surgery,” Clinical
Orthopaedics, Vol. 106, 1975, pp. 122-125.
http://dx.doi.org/10.1097/00003086-197501000-00017
[28] R. P. Kusy, “Characterization of Self-Curing Acrylic Bone
Cements,” Journal of Biomedical Materials Research,
Vol. 12, 1978, pp. 271-305.
http://dx.doi.org/10.1002/jbm.820120304
[29] Y. Oshida, A. Hashem and R. Elsalawy, “Some Mechani-
stic Observation on Water-Deteriorated Dental Composite
Resins,” Biomedical Materials and Engineering, Vol. 5,
1995, pp. 93-115.