Vol.3, No.1, 56-65 (2011)
doi:10.4236/health.2011.31011
Copyright © 2011 SciRes. Openly accessible at http://www.scirp.org/journal/HEALTH/
Health
Performance comparison between three intraoral image
receptors of different technology at a variety of tube
potential, tube current and exposure time settings using
a stepwedge phantom
Elli Katsoni1*, Ioannis A. Tsalafoutas2, Panagiotis Gritzalis3, Evripidis Stefanou3,
Evangelos Georgiou1, Emmanuel Yakoumakis1
1Department of Medical Physics, Medical School, University of Athens, Athens, Greece;
2Medical Physics Department, Agios Savvas Hospital, Athens, Greece;
3Department of Oral Diagnosis and Radiology, Dental School, University of Athens, Athens, Greece;
*Corresponding Author: jzar@rocketmail.com
Received 22 November 2010; revise 30 November 2010; accepted 3 December 2010.
ABSTRACT
Purpose: To comparatively evaluate the per-
formance of three intraoral image receptors of
different technology when exposed to different
X-ray beam spectra, dose and dose rate levels
using a stepwedge phantom. Materials and
methods: The intraoral radiographic receptors
evaluated were: the Kodak Insight F speed class
film, the Kodak RVG 6000, and the Duerr Vis-
tascan Combi PSP system. A dental quality
control phantom made of Plexiglas, containing
an aluminium stepwedge with 12 steps and 7
holes drilled in each step was radiographed
using a dental X-ray unit offering a wide range
of tube potential, tube current and exposure
time settings. The visibility of the holes in the
images produced with each one of the three
receptors was assessed by three independent
observers. For each image the total image qual-
ity score (TS) was derived from the summation
of the number of visible holes in each step. The
numbers of perceptible holes in each experi-
mental condition (TSs) were statistically ana-
lyzed through use of analysis of variance. In-
traobserver and interobserver agreement was
also measured. Results: Vistascan exhibited the
most extended useful exposure range, follow ed
by RVG 6000 and Insight. RVG 6000 exhibited
the largest TS values in all tube potential set-
tings except 70 kV where the Vistascan per-
formed better. Insight performed better than
Vistascan only at 60 and 63 kV. Vistascan per-
formed better at 66 and 70 kV, Insight at 60 and
66 kV, whereas RVG performed equally well at
all tube potential settings, except than at 52 and
70 kV. For the Insight the largest TS values w ere
obtained with the smallest ESAK values whe-
reas with the Vistascan the largest TS were ob-
tained with ESAK values that where the largest
observed. Conclusions: The performance of all
receptors tested was greatly dependent on the
exposure parameters and mainly on the kV set-
tings. Overall, the RVG 6000 offered the best
image quality at doses somewhere in between
those required by the Insight and t he Vistascan.
Keywords: Radiography; Dental; Digital; Dosage
1. INTRODUCTION
According to the UNSCEAR 2000 Report, dental in-
traoral radiography is among the most frequently per-
formed radiological procedures [1]. Although the patient
exposure associated with dental radiography is relatively
low, intraoral radiography should be optimised in order
to keep the radiation risk “as low as reasonably achiev-
able”, something that is widely known as the ALARA
principle [2].
Over the past 20 years both the X-ray units and the
X-ray receptors used in dental radiology have been
evolved. Modern dental X-ray units incorporate high
frequency generators, operate at higher tube potentials
and produce X-ray spectra that have higher mean energy
and therefore are more penetrating compared to those
produced by older dental X-ray units. These improve-
ments have contributed in the reduction of the radiation
dose to the entrance skin surface of the patient and the
E. Katsoni et al. / Health 3 (2011) 56-65
Copyright © 2011 SciRes. Openly accessible at http://www.scirp.org/journal/HEALTH/
57
Table 1. Specifications of the receptors evaluated in this study according to the data provided by the manufacturers.
Model Manufacturer Pixel size (μm) TechnologySoftware Bit/pixelWidth Height File size (MB)
INSIGHT KODAK N/A SILVER
HALIDE N/A N/A 3.1 cm 4.1 cm N/A
RVG 6000 KODAK 18.5×18.5 CMOS
KODAK
WINDOWS
6.0.1
8 1200
pixels
1600
pixels 1.8
PSP VISTASCAN DUERR SCAN PITCH
12.5
IMAGE
PLATE
DBSWIN
V.3.3 16 2476
pixels
3195
pixels Up to 9.3
enhancement of image quality [3,4] Concerning the
X-ray receptors, new digital systems have been intro-
duced in to the clinical practice and nowadays digital
radiography is considered an accepted imaging tech-
nique in dentistry. Currently, solid-state detectors based
on CCD or CMOS technology, photostimulable storage
phosphor (PSP) systems, along with the old-fashioned
but still widely used silver halide based films, are com-
mercially available for intraoral radiography.
In the international literature many articles can be
found describing digital detector systems and presenting
comparisons among various systems with regard to their
diagnostic performance [5-9]. The characteristics of the
x-ray beam, especially the x-ray beam energy, can also
contribute to image quality. Earlier studies have found
that the x-ray beam energy can affect image contrast
[10-12] and the signal-to-noise ratio [13]. For those stu-
dies various x-ray energies were used maintaining the
same exposure. Most of the equipment for intraoral ra-
diography provides limited adjustments for tube currents
and x-ray energy values. On most machines, only the
exposure time is usually set based on experience and
according tooth type.
The purpose of this study was to comparatively evalu-
ate, in a systematic inter-equipment manner, the per-
formance of three different intraoral image receptors
when exposed to different X-ray beam spectra, dose and
dose rate levels in detecting subtle radiographic density
differences using an aluminium stepwedge phantom.
2. MATERIALS AND METHODS
2.1. Image Acquisition
The intraoral radiographic receptors evaluated in this
study were: the Kodak Insight F speed class film (East-
man Kodak Company, Rochester, NY) the Kodak RVG
6000 (Eastman Kodak Company, Rochester, NY) and
the Duerr Vistascan Combi PSP system (Duerr Dental,
Bietingsheim-Bissingen, Germany). The first receptor is
a conventional silver halide film that requires chemical
processing, the second is a digital solid state receptor
based in CMOS technology that offers a direct display of
the digital image within seconds after the end of the ex-
posure and the third is a photostimulable storage phos-
phor (PSP) imaging plate which produces a digital image
after it has been scanned using a dedicated laser scanner
system. The main technical characteristics of the systems
tested are summarized in Table 1.
A modern dental X-ray unit (Prostyle Intra DC,
Plan-meca Oy, Helsinki, Finland) was used, offering
eight tube potential settings (ranging from 50 to 70 kVp),
seven tube current settings (ranging from 2-8 mA) and
26 exposure time settings (ranging from 0.01 to 3.2 sec).
The nominal total filtration was 2 mm Al and the focus
to collimator end distance was 30 cm.
A calibrated ion chamber dosimeter (Dosimeter 9010,
ionization chamber type 90x6-6; Radcal Corporation
Monrovia, USA) positioned at 30 cm from the tube fo-
cus was used to measure the dose in free air and deter-
mine the tube output at that distance. These measure-
ments were carried out for all the available tube potential,
tube current and exposure time selections, in order to
identify possible variations in output with different tube
loading values. The tube potential accuracy and repro-
ducibility were checked using a calibrated kilovolt peak
meter (Gammex RMI 245, 802108-1272, calibrated
31.01.2007, Gammex Inc., Middleton, USA).
Image quality was assessed using a quality control
phantom especially designed for dental radiography
(standard phantom for dental radiography, version 2.0),
shown in Figure 1. This phantom (henceforth referred to
as stepwedge phantom), is a Plexiglas parallelepiped
with a 30 mm by 40 mm base and 23 mm height, con-
taining a 12 step aluminium stepwedge which increases
in height from 1 to 12 mm, in 1 mm steps. In each step,
7 holes of 0.5 mm diameter have been drilled in depths
varying from 0.05 mm to 0.35 mm, in 0.05 mm incre-
ments. The stepwedge phantom has been described and
validated by Yoshiura et al. [14,15] The stepwedge
phantom was attached to the collimator end and the fo-
cus to receptor distance was kept constant at 35 cm.
The stepwedge phantom was radiographed using for
each receptor all possible tube potential, tube current and
exposure time selections combinations. The specific
X-ray unit used offered 182 mAs selections at each one
o
f the 8 different tube potential settings, that is, 1456
E. Katsoni et al. / Health 3 (2011) 56-65
Copyright © 2011 SciRes. Openly accessible at http://www.scirp.org/journal/HEALTH/
58
26mm
20mm
1mm
12mm
10mm
1mm
2mm
0.5mm
1mm
5mm
Plexiglas
Aluminum steps
hole (depth:0.05, 0.1, 0.15,.....,0.3, 0.35mm
at each corresponding step)
10mm 30mm
40mm
Standard Phantom for Dental Radiography
(version 2.0)
0.5mm
[bottom view]
[lateral view]
5mm
5mm
Figure 1. Photographs (a,b) and schematic; (c) diagrams of the stepwedge phantom that was used for the
evaluation of image quality.
different exposure factor settings. The evaluation of im-
age quality was performed in two stages. In the first stage
the images for which all the 12 steps of the stepwedge
were visible were identified. The images satisfying this
criterion were included in the sample of images that were
further evaluated and will be henceforth referred to
as‘diagnosable images’ (see Figure 2). The images that
were not satisfying this criterion were rejected and will
be referred to as ‘non-diagnosable images’. The tube
current-exposure time combinations that produced diag-
nosable images determined the useful exposure range of
each receptor at each tube potential setting.
Concerning the exposures made with the Insight films,
all films were processed immediately after exposure,
using an automatic processor (Velopex Extra-X, Mediv-
ance Instuments, England) using the Readymatic dental
developer and Readymatic dental fixer solutions (East-
man Kodak, Rochester NY) at a temperature of 27℃.
This processor features an automatic replenishment sys-
Figure 2. Radiographic image of the
stepwedge phantom.
tem, however, in order to ensure that the processing
conditions remained fairly constant during the experi-
ments, the processing stability was repeatedly tested
every 50 films using sensitometry (Pehamed densitome-
ter Densinorm 21, PEHA med. Geräte GmbH-Sulzbach,
Germany). The acquired film radiographs were mounted
in opaque plastic holders and coded for later use. The
film radiographs were evaluated on a viewing box, with
all extraneous light masked. The observers were allowed
to use magnifying glasses (at multiple × 2 magnifica-
tion).
Concerning the RVG 6000 receptor, the original soft-
ware of the system was used for image capture. No im-
age processing was performed to enhance image quality
other than the system’s default pre-process. Finally,
concerning the Vistascan, the image plates were un-
packed in a dimly lit room and read out immediately
after exposure, using the Vistascan Combi system.[16,17]
The scanner’s resolution pitch settings were adjusted to
12.5 μm (corresponding to a theoretical resolution of 40
line pairs per mm).
2.2. Image Evaluation
Three certified dentists, postgraduate students in the
department of Oral Diagnosis and Radiology of the
Dental School of the University of Athens, served as
observers. They have over 5 years of experience inter-
preting analog and digital images. Before the actual im-
age quality evaluation session, the observers were first
trained in order to get familiar with the radiographic
images of the phantom and the rating process. During
E. Katsoni et al. / Health 3 (2011) 56-65
Copyright © 2011 SciRes. Openly accessible at http://www.scirp.org/journal/HEALTH/
59
the image quality evaluation sessions, the observers had
to state the number of holes that they could perceive at
each step with 100% confidence level. The observers
rated the images independently (to account for in-
ter-observer variability), one image at a time and in a
random order, and dictated the ratings orally. Each image
was evaluated twice by each observer to account for
intra-observer variability. Due to the large number of
images, several sessions were required for each ob-
server to evaluate all images. For this reason, at the
beginning of each session and before the actual evalua-
tion, each observer was asked to observe 10 test control
images and rate them in order to get familiar with the
images again, before proceeding to the actual evalua-
tion session. For each image the average value of the
number of visible holes perceived by the three observ-
ers in each step was calculated and the summation of
these averages in all 12 steps was accounted as the total
image quality score (TS) of that image.
All digital images were viewed in fit to screen mode
on a 19-inch TFT monitor (Sony SDMHS95PR), with
1280 × 1024 resolution under subdued lighting condi-
tions. The monitor’s brightness and contrast were ad-
justed using the SMPTE (Society of Motion Picture and
Television Engineers) test pattern [18,19].
2.3. Statistical Analysis
The numbers of perceptible holes in each experimen-
tal condition (TSs) were statistically analyzed through
use of analysis of variance. Repeated measures ANOVA
were calculated for validity in relation to kV selection,
mA selection, exposure time and entrance surface air
kerma. Post hoc analysis was carried out using Tukey’s
test. P values less than 0.05 were considered to be statis-
tically significant. Intraobserver agreement was meas-
ured by Cohen’s Kappa and interobserver agreement was
measured by Fleiss kappa coefficient.
3. RESULTS
The output measurements revealed that the linearity of
output was within accepted limits at all tube potentials,
even though for small tube loadings (0.5 mAs) a reduc-
tion in output of up to 20% was observed with respect to
the mean value of output over the whole mAs range. The
reproducibility of output using the same exposure condi-
tions was better than 1%. The tube potential accuracy
and reproducibility were better than 3% and 1%, respec-
tively.
For all the images acquired, the receptor type, the ex-
posure parameters (kV, mA, s) were noted and the re-
spective entrance surface air kerma (ESAK) values were
then assigned. The number of diagnosable images was in
total 1230; 275 with Insight, 339 with RVG 6000 and
616 with Vistascan. From these values and Figure 3,
where the useful exposure range of each receptor is
shown, it is obvious that the VistaScan exhibited the
most extended useful exposure range for all tube poten-
tial settings. The useful exposure range of RVG 6000
was smaller compared to Vistascan but clearly larger
compared to that of the Inshight. It must be stressed
however, that the useful exposure ranges shown in Fig-
ure 3 are strictly valid only for the specific phantom and
geometric conditions used.
The useful exposure range of each receptor was de-
termined taking into account only the diagnosable im-
ages, that is, those images where all 12 steps were visi-
ble. However, within the sample of diagnosable images
great variations in the TS values were observed, which
are graphically represented in Figure 4. In this figure,
the lower bars extend from the minimum TS value ob-
served up to the 1st quartile value, the white boxes from
the 1st quartile value to the median value, the black
boxes from the median to the 3rd quartile value and the
upper bars from the 3rd quartile value to the maximum
TS value observed at the specific tube potential setting.
The black and white boxes represent the 50% of the ob-
served TS values at each tube potential setting. This fig-
ure clearly indicates that image quality was strongly de-
pended on the exposure factors used, not only for the
Insight, but for the digital receptors also. Even if the
exposure latitude of digital receptors is extended via
software manipulations of the recorded signals, still im-
age quality cannot be maintained constant within the
useful exposure range.
While Figure 4 is indicative of the large variations in
image quality with exposure factor settings, it is not so
useful for straightforward comparisons among receptors,
where the main interest is which receptor performs best
and at which tube potential setting and ESAK levels. To
facilitate comparisons among the receptors in terms of
the image quality offered and the respective radiation
dose required, the TS values of the best image obtained
with each receptor at each one of the tube potential set-
tings used and the respective exposure factors and ESAK
values required to obtain these images are given in Table
2. It can be seen that with the exception of 70 kV where
the largest TS value has been obtained with the Vistas-
can, for all the other tube potential settings, the largest
TS value was obtained with the RVG 6000 and it was
close to or larger than 60. The doses required were
roughly double compared to the respective doses re-
quired to obtain the maximum TS with the Insight. The
largest doses to obtain the maximum TS were observed
for the Vistascan and ranged from 3.2 to 4.8 mGy. It can
be seen that for the Insight the largest TS was obtained at
60 kV and at the same tube potential the largest TS
overall was obtained with the RVG 6000. For the Vis-
E. Katsoni et al. / Health 3 (2011) 56-65
Copyright © 2011 SciRes. http://www.scirp.org/journal/HEALTH/
60
Figure 3. Graphical representation of the exposure ranges of the three receptors tested in terms of mAs and in terms of
ESAK used to obtain the stepwedge phantom images at all tube potential settings.
Figure 4. Graphical description of the statistical variations observed in the number of visible holes (TS)
at different tube potential settings for each receptor. The lower bars extend from the minimum TS value
observed up to the 1st quartile value, the white boxes from the 1st quartile value to the median value, the
black boxes from the median to the 3rd quartile value and the upper bars from the 3rd quartile value to the
maximum TS value observed at the specific tube potential setting.
Openly accessible at
E. Katsoni et al. / Health 3 (2011) 56-65
Copyright © 2011 SciRes. Openly accessible at http://www.scirp.org/journal/HEALTH/
61
Table 2. The total score (TS), the tube loading (mAs) and the respective ESAK (in mGy) of the image with the largest TS are given.
Receptor¥ kV 50 52 55 57 60 63 66 70
Insight
46.7 @ 3.0 mAs
(6 mA × 0.5 s):
45.3 @ 2.4 mAs
(3 mA × 0.8 s):
1.0 mGy 1.2 mGy
46.7 @ 2.0
mAs (5 mA ×
0.4 s): 1.0 mGy
45.3 @ 1.6
mAs (8 mA ×
0.2 s): 0.9 mGy
63.0 @ 1.4 mAs
(7 mA × 0.2 s):
0.8 mGy
56.7 @ 1.4 mAs
(7 mA × 0.2 s):
0.9 mGy
57.0 @ 1.3 mAs
(2 mA × 0.64 s):
0.9 mGy
50.3 @ 1.0
mAs (6 mA ×
0.16 s): 0.8
mGy
RVG 6000
60.3 @ 6.3 mAs
(5 mA × 1.25
s): 2.5 mGy
59.7 @ 5.0 mAs
(4 mA × 1.25
s): 2.2 mGy
66.3 @ 4.8
mAs (3 mA ×
1.6 s): 2.3 mGy
64.3 @ 4.8
mAs (3 mA ×
1.6 s): 2.5 mGy
66.7 @ 2.5 mAs
(2 mA × 1.25
s): 1.4 mGy
63.3 @ 2.5 mAs
(2 mA × 1.25
s): 1.6 mGy
66.3 @ 3.2 mAs
(2 mA × 1.6 s):
2.2 mGy
55.3 @ 2.5
mAs (2 mA ×
1.25 s): 2.0
mGy
Vistascan
48.3 @ 8.0 mAs
(5mA × 1.6s):
3.2 mGy
48.7 @10 mAs
(4mA × 2.5s):
4.4 mGy
49.0 @ 6.4
mAs (8mA ×
0.8s): 3.2 mGy
53.0 @ 8.0
mAs (8mA ×
1s):
4.4 mGy
54.0 @ 8.0 mAs
(4mA × 2s):
50.3 @ 6.0 mAs
(6mA × 1s):
4.8 mGy 4.1 mGy
61.7 @ 5.1 mAs
(8mA × 0.64s):
3.9 mGy
58.3 @ 3.8
mAs (6 mA ×
0.64 s): 3.3
mGy
Table 3. Influence of the kV selection on the performance of the evaluated receptors. Bold marked digits indicate statistical
significance (p < 0.05).
INSIGHT
kV 50 52 55 57 60 63 66 70
50 - 0.001 0.963 <0.001 <0.001 <0.001 <0.001 0.797
52 0.001 - <0.001 0.968 <0.001 <0.001 <0.001 <0.001
55 0.963 <0.001 - <0.001 <0.001 <0.001 <0.001 1.000
57 <0.001 0.968 <0.001 - <0.001 <0.001 <0.001 <0.001
60 <0.001 <0.001 <0.001 <0.001 - 0.001 0.011 <0.001
63 <0.001 <0.001 <0.001 <0.001 0.001 - 0.999 <0.001
66 <0.001 <0.001 <0.001 <0.001 0.011 0.999 - <0.001
70 0.797 <0.001 1.000 <0.001 <0.001 <0.001 <0.001 -
RVG 6000
kV 50 52 55 57 60 63 66 70
50 - 0.079 <0.001 <0.001 <0.001 0.012 0.025 0.968
52 0.079 - 0.070 0.773 0.148 0.996 1.000 0.001
55 <0.001 0.070 - 0.900 1.000 0.413 0.149 <0.001
57 <0.001 0.773 0.900 - 0.977 0.993 0.917 <0.001
60 <0.001 0.148 1.000 0.977 - 0.618 0.284 <0.001
63 0.012 0.996 0.413 0.993 0.618 - 1.000 <0.001
66 0.025 1.000 0.149 0.917 0.284 1.000 - <0.001
70 0.968 0.001 <0.001 <0.001 <0.001 <0.001 <0.001 -
VISTASCAN
kV 50 52 55 57 60 63 66 70
50 - 0.031 1.000 0.006 0.007 0.889 <0.001 <0.001
52 0.031 - 0.089 <0.001 <0.001 0.488 <0.001 <0.001
55 1.000 0.089 - 0.001 0.001 0.985 <0.001 <0.001
57 0.006 <0.001 0.001 - 1.000 <0.001 <0.001 <0.001
60 0.007 <0.001 0.001 1.000 - <0.001 <0.001 <0.001
63 0.889 0.488 0.985 <0.001 <0.001 - <0.001 <0.001
66 <0.001 <0.001 <0.001 <0.001 <0.001 <0.001 - 0.041
70 <0.001 <0.001 <0.001 <0.001 <0.001 <0.001 0.041 -
E. Katsoni et al. / Health 3 (2011) 56-65
Copyright © 2011 SciRes. Openly accessible at http://www.scirp.org/journal/HEALTH/
62
Figure 5. The distribution of images with TS larger or
equal with the 90% of the largest total score observed
for each receptor overall, with respect to the mA setting
and the ESAK at all tube potential settings is shown.
tascan the largest TS value was observed at 66 kV.
Within each receptor the statistical significance of the
tube potential settings in regard to the overall perform-
ance is presented in Tab le 3 . In this table it can be seen
that for the Insight film and the Vistascan system the TS
was greatly varying with the tube potential selection.
This dependence of TS with tube potential selection can
be also appreciated from Figure 4, when looking at the
median and the maximum TS values obtained at each
tube potential. The Insight showed an increased per-
formance at 60 kV, Vistascan at 66 kV whereas the RVG
6000 could perform equally well at almost all tube po-
tentials except 52 and 70 kV. It must be noted finally,
that no straightforward correlation was observed be-
tween TS and ESAK in any of the receptors compared,
not even for the Vistascan where the image quality was
up to a point increasing with increasing dose.
An interesting observation was made during the
analysis of the results, concerning the dependence of TS
on the tube current selection, which for a given tube po-
tential selection is what determines the incident dose rate
to the receptor. For the RVG 6000 it was observed that
for a given kV selection, the same mAs that produced a
diagnosable image when a low mA selection was used
produced a non-diagnosable image when a high mA se-
lection was used. Table 2 is rather suggestive of this
behaviour since for all tube potential settings above 52
kV the image with the largest TS was obtained with ei-
ther the 3 or the 2 mA selection. Furthermore, the statis-
tical analysis also verified the dependence of the RVG
6000 performance on the mA selection, as the TS values
obtained with the 2 mA selection differed statistically
significant (p < 0.05) from the other selections available.
Unlike the RVG 6000, for the Insight and the Vistascan
not any dependence of their performance on dose rate
was documented.
The dependence of image quality on tube potential,
tube current and ESAK can be better appreciated from
Figure 5. In this figure the distribution of images with
TS of at least 90% of the largest TS value observed for
each receptor within the whole useful exposure range is
given with respect to the tube current settings and ESAK
at the various tube potential settings used. The threshold
of 90% was chosen in order to sort out the images with
the best quality overall. For the Insight almost all except
one best quality images were obtained at 60 kV, with
medium to high mA settings and with ESAK values
ranging from 0.52 to 1.36 mGy. For the RVG 6000 the
best quality images were obtained over a wide range of
tube potentials (50-66 kV) and with ESAK values rang-
ing from 0.55 to 2.82 mGy. However, it is characteristic
that 8 out of the 10 images were obtained using either
the 2 or the 3 mA tube current selections. For the Vis-
tascan, the best quality images were all obtained at 66
kV (using the 3 largest mA selections) and at 70 kV with
all mA selections, with the ESAK values ranging from
0.96 to 4.2 mGy.
Finally, in Figure 6 the variation of the average num-
ber of visible holes in each one of the 12 steps is given,
considering only the images with the highest TS at each
one of the tube current selection available, for all the
tube potential settings used. For the images obtained
with Vistascan the maximum number of visible holes
was observed in the first step and then the number of
visible holes gradually reduced when moving to the
thicker Aluminium steps, similarly with the images ac-
quired with the Insight, with the only exception that in
the Insight images the maximum number of visible holes
was most often observed in the second step rather than in
the first. In the images acquired with the RVG 6000 the
visibility of the holes was severely reduced in the first
two steps, reached a maximum in steps 5 and 6 and then
gradually reduced again. Evidently, the visibility of
holes in the thicker steps is poorer in the images ac-
quired with the Insight and the Vistascan than in the im-
ages acquired with the RVG 6000 system.
Due to the subjective nature of the image quality
evaluation procedure, in order to accept the validity of
the above results it was necessary to also validate statis-
tically the reliability of the observers’ evaluations. A
high level of agreement between the first and the second
evaluation and consequently high intra-observer reliabil-
ity scores (kappa value) were obtained, as can be seen in
Table 4. Concerning the inter-observer variability, kappa
values of 0.92, 0.88 and 0.96 were observed respectively
for the Insight, Kodak RVG 6000 and Vistascan, indi-
ating a very good inter-observer agreement. c
E. Katsoni et al. / Health 3 (2011) 56-65
Copyright © 2011 SciRes. Openly accessible at http://www.scirp.org/journal/HEALTH/
63
Figure 6. The variation of the average number of visible holes in each one of the 12 steps (considering only the
images with the highest TS for each tube current setting) is given for the different tube potential settings used.
Table 4. Intra observer agreement measured by Cohen’s Kap-
pa.
overall Observer 1 Observer 2 Observer 3
Insight 0.95 0.94 0.95 0.96
RVG 6000 0.90 0.89 0.91 0.90
Vistascan 0.93 0.93 0.92 0.94
4. DISCUSSION
All comparisons presented above are based on the step-
wedge phantom and therefore their validity relies on the
assumption that this phantom is appropriate for simulat-
ing adequately clinical conditions. Yoshiura et al. [14,15]
who evaluated this phantom, concluded that the X-ray
attenuation range produced with this phantom is similar
to that produced in clinical practice. Furthermore, they
concluded that image quality could be quantitatively
evaluated by means of the number of visible holes, since
their detection is equivalent with the detection of small
lesions in an actual clinical situation. Although this
phantom does not represent all diagnostic tasks encoun-
tered in actual clinical situations, it is by all means suffi-
cient for the relative comparison of performance among
different systems.
No image enhancement was carried out in this study.
The software provided by the manufactures displayed
postprocessed images immediately after acquisition.
Several studies on the effects of image quality and ob-
server performance have been published. Further image
processing could improve the quality of the displayed
images according to some studies [20,21] whereas ac-
cording to other does not make any significant difference
[22,23]. The effect of image processing varies per sys-
tem but its impact on image quality was outside the
scope of the current study.
Concerning the Vistascan, the extended useful expo-
sure range observed in this study is in agreement with
what is already known from the applications of the PSP
technology receptors in general and dental radiology
[24-28]. Furthermore, the high ESAK values required to
produce images with the largest TS are in agreement
with the study of Berkhout et al. [24] who demonstrated
that PSP systems produce the best quality images at
doses up to ten times the minimum dose that produces a
diagnosable image. However, this characteristic of Vis-
tascan can lead to the systematic use of higher doses than
those actually needed for diagnosis, something that from
the aspect of radiation protection is considered disadvan-
tageous and in the general radiology applications it has
been reported as ‘the exposure factor creep’ [29].
In this study, the dependence of the performance on
the tube potential setting was observed mainly for the
Inshight and the Vistascan. Likewise, Svenson et al. [30]
demonstrated the differences in diagnostic accuracy for
different tube potential settings for two films differing in
speed. Concerning the Insight, Kaeppler et al. [31] also
concluded that the Insight film achieved the best results
at 60 kV.
In the present study it was seen that in the overall rat-
ing, the Kodak RVG6000 performed better in the detec-
tion of holes compared to the other two receptors tested.
This result is in agreement with the results of other re-
searchers who also demonstrated the superiority of
RVG6000 receptor [33].
It is well known that the majority of the dental X-ray
units currently in use worldwide offer a single tube po-
tential and tube current selection. Since specific recep-
tors perform better at certain tube potential and tube
current settings, it is important for the potential buyer to
take this into account when a new receptor or dental
X-ray unit is to be combined with existing equipment or
when both a new receptor and a new dental X-ray unit
are to be bought. Although, an image of adequate diag-
nostic quality can be achieved by all receptors tested in
many tube potential and tube current settings, the best
image quality was obtained with the least radiation ex-
posure only when specific exposure parameters were
used. For example, even the Vistascan which exhibited
the most extended useful exposure range at all tube po-
tential settings from 50 to 70 kV, performed better at 66
E. Katsoni et al. / Health 3 (2011) 56-65
Copyright © 2011 SciRes. Openly accessible at http://www.scirp.org/journal/HEALTH/
64
and at 70 kV. Therefore, Vistascan should be combined
with a dental X-ray unit operating at 70 kV rather with
one operating at 60 kV. Unlike to Vistascan, RVG 6000
and Inshight should be combined preferably with an
X-ray unit operating at 60 kV.
The ALARA statement endorses the principle that (in-
dividual) doses should be as low as reasonably achiev-
able. In this context European guidelines [30] recom-
mended the establishment of diagnostic reference levels
(DRLs). The European guidelines [30] recommended a
DRL of 4 mGy absorbed dose in air measured at the end
of the spacer cone for a standard maxillary molar projec-
tion. In this study it was shown that for the specific
phantom used, images of good diagnostic quality could
be obtained with all receptors tested using just a fraction
of the proposed DRL (see Figure 5).
Similar surveys conducted in dental radiographic fa-
cilities [34-36] over the last 10 years have demonstrated
a trend for reduction of the entrance surface dose, with
the use of faster films and digital receptors, as well as
with modern x-ray units and rectangular collimation.
Compared with the previous guidelines, the 2007 ICRP
[37] recommendations for estimating risk associated
with exposure to radiation, increased the emphasis
given on the structures located in the oral region, par-
ticularly the salivary glands. According to the revised
recommendations for calculating effective dose, dental
radiography involves 32% to 422% more risk than that
previously thought [38]. Therefore, efforts should be
made to reduce dose as much as possible but not at ex-
pense of image quality and diagnostic accuracy. In this
context it is important to determine for each receptor the
exposure factors settings which can produce a good
quality image with the least radiation dose possible. For
this reason it would be ideal if receptor manufacturers
could determine, using a phantom like the one used in
this study, the optimal exposure factor settings and the
respective TS and ESAK values for each receptor that
becomes commercially available and include this in-
formation in their technical data sheets. In this way the
potential users would be able to determine if a given
receptor is well suited for the dental X-ray unit that may
already have or they intend to buy and furthermore they
would be able to compare receptors using performance
indices that relate to the clinical practice and therefore
are easy to comprehend.
The main results of this study in terms of the relative
performance of the three receptors studied can be sum-
marized as follows: Vistascan exhibited the most ex-
tended useful exposure range, followed by RVG 6000
and Insight. RVG 6000 exhibited the largest TS values in
all tube potential settings except 70 kV where the Vis-
tascan performed better. Insight performed better than
Vistascan only at 60 and 63 kV. Vistascan exhibited its
largest TS values at 66 and 70 kV, Insight at 60 and 66
kV, whereas RVG exhibited large TS values at all tube
potential settings, except than at 52 and 70 kV. For the
Insight the largest TS values were obtained with the
smallest ESAK values whereas with the Vistascan the
largest TS were obtained with ESAK values that where
the largest observed. RVG 6000 was the only system that
exhibited a dependence on the dose rate since most of
the large TS values were obtained at low mA selections.
REFERENCES
[1] United Nations Scientific Committee on the Effects of
Atomic Radiation (2000) Sources and effect of ionizing
radiation. UNSCEAR publications, Report, 1.
[2] European Union. Council Directive 97/43 Euratom (1997)
On health protection of individuals against the dangers of
ionizing radiation in relation to medical exposures, and re-
pealing directive 84/466 Euratom. Official Journal of the
European Communities, 22, L180.
[3] Kitagawa, H. and Farman, A.G. (2004) Effect of beam
energy and filtration on the signal-to-noise ratio of the
Dexis intraoral X-ray detector. Dentomaxillofacial Radi-
ology, 33, 21-24. doi:10.1259/dmfr/26493631
[4] Wenzel, A. (2006) A review of dentists' use of digital
radiography and caries diagnosis with digital systems.
Dentomaxillofacial Radiology, 5, 307-314.
doi:10.1259/dmfr/64693712
[5] Hellen-Halme, K. (2007) Quality aspects of digital radi-
ography in general dental practice. Swedish Dental
Journal. Supplement, 184, 9-60.
[6] Hellen-Halme, K., Nilsson, M. and Petersson, A. (2007)
Digital radiography in general dental practice: A field
study. Dentomaxillofacial Radiology, 36, 249-255.
doi:10.1259/dmfr/95125494
[7] Cowen, A.R., Kengyelics, S.M. and Davies, A.G. (2008)
Solid-state, flat-panel, digital radiography detectors and
their physical imaging characteristics. Clinical Radiology,
63, 487-498. doi:10.1016/j.crad.2007.10.014
[8] Hintze, H. (2006) Diagnostic accuracy of two software
modalities for detection of caries lesions in digital radio-
graphs from four dental systems. Dentomaxillofacial Ra-
diology, 35, 78-82. doi:10.1259/dmfr/50356588
[9] Lanca, L., Silva, A., Alves, E., Serranheira, F. and Cor-
reia, M. (2008) Evaluation of exposure parameters in
plain radiography: A comparative study with European
guidelines. Radiation Protection Dosimetry, 129, 316-320.
doi:10.1093/rpd/ncn144
[10] Harada, T., Nishikawa, K., Shibuya, H., Hayakawa, Y.
and Kuroyanagi, K. (1995) Sens-A-Ray characteristics
with variation in beam quality. Oral Surgery, Oral Medi-
cine, Oral Pathology, Oral Radiology & Endodontics, 80,
120-123. doi:10.1016/S1079-2104(95)80028-X
[11] Goshima, T., Goshima, Y., Scarfe, W.C. and Farman, A.G.
(1996) Sensitometric response of the Sens-A-Ray, a charge
coupled imaging device, to changes in beam energy. Den-
tomaxillofacial Rad io lo gy, 25, 17-18.
[12] McDonnell, D. and Price, C. (1993) An evaluation of the
Sens-A-Ray digital dental imaging system. Dentomaxil-
lofacial Radiology, 22, 121-126.
[13] Kitagawa, H. and Farman, A.G. (2004) Effect of beam
energy and filtration on the signal to noise ratio of the
E. Katsoni et al. / Health 3 (2011) 56-65
Copyright © 2011 SciRes. http://www.scirp.org/journal/HEALTH/Openly accessible at
65
Dexis intraoral x-ray detector. Dentomaxillofacial Radi-
ology, 33, 21-24. doi:10.1259/dmfr/26493631
[14] Yoshiura, K., Kawazu, T., Chikui, T., Tatsumi, M., To-
kumori, K., Tanaka, T., et al. (1999) Assessment of image
quality in dental radiography, part 1: Phantom validity.
Oral Surgery, Oral Medicine, Oral Pathology, Oral Ra-
diology, and Endodontics, 87, 115-122.
doi:10.1016/S1079-2104(99)70304-5
[15] Yoshiura, K., Kawazu, T., Chikui, T., Tatsumi, M., To-
kumori, K., Tanaka, T., et al. (1999) Assessment of image
quality in dental radiography, part 2: Optimum exposure
conditions for detection of small mass changes in 6 in-
traoral radiography systems. Oral Surgery, Oral Medi-
cine, Oral Pathology, Oral Radiology, and Endodontics,
87, 123-129. doi:10.1016/S1079-2104(99)70305-7
[16] Ang, D.B., Angelopoulos, C. and Katz, J.O. (2006) How
does signal fade on photo-stimulable storage phosphor
imaging plates when scanned with a delay and what is
the effect on image quality? Oral Surgery, Oral Medicine,
Oral Pathology, Oral Radiology, and Endodontics, 102,
673-679. doi:10.1016/j.tripleo.2005.11.002
[17] Ramamurthy, R., Canning, C.F., Scheetz, J.P. and Farman,
A.G. (2004) Impact of ambient lighting intensity and du-
ration on the signal-to-noise ratio of images from photo-
stimulable phosphor plates processed using DenOptix
and ScanX systems. Dentomaxillofacial Radiology, 33,
307-311. doi:10.1259/dmfr/91373164
[18] Parsons, D.M., Kim, Y. and Haynor, D.R. (1995) Quality
control of cathode-ray tube monitors for medical imaging
using a simple photometer. Journal of Digital Imaging, 8,
10-20. doi:10.1007/BF03168051
[19] Jervis, S.E. and Brettle, D.S. (2003) A practical approach
to soft-copy display consistency for PC-based review
workstations. The British Journal of Radiology, 76,
648-652. doi:10.1259/bjr/25693100
[20] Alpoz, E., Sogur, E. and Baksi Akdeniz, B.G. (2007)
Perceptibility curve test for digital radiographs before
and after application of various image processing algo-
rithms. Dentomaxil lofacial Radiology, 36, 490-494.
doi:10.1259/dmfr/20897311
[21] Li, G., Sanderink, G.C., Berkhout, W.E., Syriopoulos, K.
and van der Stelt, P.F. (2007) Detection of proximal car-
ies in vitro using standard and task-specific enhanced
images from a storage phosphor plate system. Caries
Research, 41, 231-234. doi:10.1159/000099324
[22] Koob, A., Sanden, E., Hassfeld, S., Staehle, H.J. and
Eickholz, P. (2004) Effect of digital filtering on the
measurement of the depth of proximal caries under dif-
ferent exposure conditions. American Journal of Den-
tistry, 17, 388-393.
[23] Li, G., Engström, P.E. and Welander, U. (2007) Meas-
urement accuracy of marginal bone level in digital radio-
graphs with and without color coding. Acta Odontologica
Scandinavica, 65, 254-258.
doi:10.1080/00016350701452089
[24] Berkhout, W.E., Beuger, D.A., Sanderink, G.C. and van der
Stelt, P.F. (2004) The exposure range of digital radio-
graphic systems: Dose reduction or risk of overexposure?
Dentomaxillofacial Radiology, 33, 1-5.
doi:10.1259/dmfr/40677472
[25] Borg, E., and Grondahl, H.G. (1996) On the exposure
range of different X-ray photon detectors in intra-oral ra-
diography. A comparison of image quality in film,
charge-coupled device and storage phosphor systems.
Dentomaxillofacial Radiology, 25, 82-88.
[26] Borg, E., Attaelmanan, A. and Grondahl, H.G. (2000)
Image plate systems differ in physical performance. Oral
Surgery, Oral Medicine, Oral Pathology, Oral Radiology,
and Endodontics, 89, 118-124.
doi:10.1016/S1079-2104(00)80026-8
[27] Farman, A.G. and Farman, T.T. (2005) A comparison of
18 different x-ray detectors currently used in dentistry.
Oral Surgery, Oral Medicine, Oral Pathology, Oral Ra-
diology, and Endodontics, 99, 485-489.
doi:10.1016/j.tripleo.2004.04.002
[28] Kitagawa, H., Farman, A.G., Scheetz, J.P., Brown, W.P.,
Lewis, J., Benefiel, M., et al. (2000) Comparison of three
intra-oral storage phosphor systems using subjective im-
age quality. Dentomaxillofacial Radiology, 29, 272-276.
doi:10.1038/sj.dmfr.4600532
[29] Willis, C.E. (2004) Strategies for dose reduction in ordinary
radiographic examinations using CR and DR. Pediat ric Ra-
diolog y, 34, 196-200. doi:10.1007/s00247-004-1269-6
[30] Svenson, B., Grondahl, H.G., Petersson, A. and Olving, A.
(1985) Accuracy of radiographic caries diagnosis at dif-
ferent kilovoltages and two film speeds. Swedish Dental
Journal, 9, 37-43.
[31] Kaeppler, G., Dietz, K. and Reinert, S. (2007) Influence
of tube potential setting and dose on the visibility of le-
sions in intraoral radiography. Dentomaxillofacial Radi-
ology, 36, 75-79. doi:10.1259/dmfr/75743437
[32] Ramamurthy, R., Scheetz, J.P., Clark, S.J. and Farman,
A.G. (2006) Effects of imaging system and exposure on
accurate detection of the second mesio-buccal canal in
maxillary molar teeth. Oral Surgery, Oral Medicine, Oral
Pathology, Oral Radiology, and Endodontics, 102, 796-802.
doi:10.1016/j.tripleo.2006.02.009
[33] Radiation Protection (2004) European guidelines on ra-
diation protection in dental radiology. The safe use of ra-
diographs in dental practice. Issue No. 136.
[34] Yakoumakis, E.N., Tierris, C.E., Stefanou, E.P., Pha-
nourakis, I.G. and Proukakis, C.C. (2001) Image quality
assessment and radiation doses in intraoral radiography.
Oral Surgery, Oral Medicine, Oral Pathology, Oral Ra-
diology, and Endodontics, 91, 362-368.
doi:10.1067/moe.2001.111940
[35] Hatziioannou, K., Psarouli, E., Papanastassiou, E., Bous-
bouras, P., Kodona, H., Kimoundri, O., et al. (2005)
Quality control and diagnostic reference levels in in-
traoral dental radiographic facilities. Dentomaxillofacial
Radiology, 34, 304-307. doi:10.1259/dmfr/38802780
[36] Gonzalez, L. and Moro, J. (2007) Patient radiation dose
management in dental facilities according to the X-ray
focal distance and the image receptor type. Dentomaxil-
lofacial Radiology, 36, 282-284.
[37] Valentin, J. (2007) The 2007 recommendations of the
international commission on radiological protection. El-
sevier, Oxford. doi:10.1259/dmfr/67494525
[38] Ludlow, J.B., Davies-Ludlow, L.E. and White, S.C.
(2008) Patient risk related to common dental radio-
graphic examinations: The impact of 2007 International
Commission on Radiological Protection recommenda-
tions regarding dose calculation. Journal of American
Dental Association, 139, 1237-1243.